Literature DB >> 30811355

Are Serum Metal Ion Levels a Concern at Mid-term Followup of Revision Knee Arthroplasty With a Metal-on-metal Hinge Design?

Antonio Klasan1, Esko Meine, Susanne Fuchs-Winkelmann, Turgay Efe, Friedrich Boettner, Thomas Jan Heyse.   

Abstract

BACKGROUND: Elevated serum levels of chromium and cobalt ions in metal-on-metal (MoM) bearing surfaces is a well-known phenomenon in THA. However, few studies have addressed this issue in complex primary and revision knee arthroplasty using a MoM hinged mechanism, and no study, to our knowledge, has investigated knees with MoM hinges in patients without megaprostheses (tumor prostheses). QUESTIONS/PURPOSES: We analyzed a series of patients who received MoM hinged revision knee prostheses and asked: (1) What are the serum metal ion levels at short-term followup? (2) Is there any correlation between metal ion levels and the Knee Society Score (KSS) at this followup?
METHODS: Between 2013 and 2017, we performed 198 revision knee arthroplasties, of which 32 (17.7%) were performed with a latest-generation MoM hinge knee design. In addition, three complex primary TKAs utilizing the same design were included in this study. The device features a metal-on-polyethylene bearing with a MoM hinge. During that period, our general indications for using a hinge were single-stage and two-stage revision surgeries, revisions with large bone defects, and primary TKA with > 20° mechanical malalignment or collateral ligament insufficiency. Of the 35 patients who received this device, 23 patients (65% of the overall group who received this implant; 11 males, 12 females) were available for followup at a median of 28 months (range, 13-61 months), and the remaining 12 (35%) patients were lost to followup. Our rationale for reporting before the more typical 2-year minimum was the finding of elevated serum ion levels with unclear clinical significance. Median age at the time of surgery was 68 years (range, 52-84 years). None of the patients included in the study had other implants with MoM bearings. Serum ion levels of chromium (III) and cobalt were assessed using mass spectrometry. Ion levels > 5 ppb were considered elevated. Clinical outcome was assessed using the original KSS.
RESULTS: Median chromium serum level was 6.3 ppb (range, 0.6-31.9 ppb) and median cobalt serum level was 10.5 (range, 1.0-47.5 ppb). Of the 23 patients, 16 had elevated serum ion levels. There was a moderate correlation between KSS and both chromium (p = 0.029, r = 0.445) and cobalt (p = 0.012, r = 0.502) levels.
CONCLUSIONS: Elevated metal ion levels and radiolucent lines are common after surgery with this MoM hinge design at short-term followup, and we believe this finding is of great concern. Although no patient has yet been revised, these patients will be closely monitored. We recommend that serum ion analysis become a routine part of followup after any hinge TKA in an attempt to better understand the potential consequences of this phenomenon. LEVEL OF EVIDENCE: Level IV, therapeutic study.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30811355      PMCID: PMC7000071          DOI: 10.1097/CORR.0000000000000638

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  21 in total

1.  A comparison of inductively coupled plasma mass spectrometry with electrothermal atomic absorption spectrophotometry for the determination of trace elements in blood and urine from non occupationally exposed populations.

Authors:  M A White
Journal:  J Trace Elem Med Biol       Date:  1999-07       Impact factor: 3.849

2.  Loss of sight and sound. Could it be the hip?

Authors:  Maria Cristina Rizzetti; Paolo Liberini; Guido Zarattini; Simona Catalani; Ugo Pazzaglia; Pietro Apostoli; Alessandro Padovani
Journal:  Lancet       Date:  2009-03-21       Impact factor: 79.321

3.  Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study.

Authors:  Bilal Farouk El-Zayat; Thomas J Heyse; Nelson Fanciullacci; Luc Labey; Susanne Fuchs-Winkelmann; Bernardo Innocenti
Journal:  Arch Orthop Trauma Surg       Date:  2016-10-04       Impact factor: 3.067

4.  Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.

Authors:  A J Hart; S A Sabah; A S Bandi; P Maggiore; P Tarassoli; B Sampson; J A Skinner
Journal:  J Bone Joint Surg Br       Date:  2011-10

5.  Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results.

Authors:  L D Dorr; Z Wan; D B Longjohn; B Dubois; R Murken
Journal:  J Bone Joint Surg Am       Date:  2000-06       Impact factor: 5.284

6.  Primary total knee arthroplasty using rotating-hinge prosthesis in severely affected knees.

Authors:  Jae-Hyuk Yang; Jung-Ro Yoon; Chi-Hun Oh; Taik-Sun Kim
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-07-20       Impact factor: 4.342

7.  Wear in total knee arthroplasty--just a question of polyethylene?: Metal ion release in total knee arthroplasty.

Authors:  Jan Philippe Kretzer; Joern Reinders; Robert Sonntag; Sebastien Hagmann; Marcus Streit; Sebastian Jeager; Babak Moradi
Journal:  Int Orthop       Date:  2013-11-12       Impact factor: 3.075

8.  Knee salvage in revision arthroplasty after massive bone loss of the femur condyles (≥Engh III) with a single-modular-hinged knee revision implant.

Authors:  Arne Streitbuerger; Jendrik Hardes; Georg Gosheger; Ralf Dieckmann; Steffen Hoell
Journal:  Arch Orthop Trauma Surg       Date:  2016-07-01       Impact factor: 3.067

9.  Serum metal ion concentrations in paediatric patients following total knee arthroplasty using megaprostheses.

Authors:  Jörg Friesenbichler; Patrick Sadoghi; Werner Maurer-Ertl; Joanna Szkandera; Mathias Glehr; Kathrin Ogris; Matthias Wolf; Christian Weger; Andreas Leithner
Journal:  Biomed Res Int       Date:  2014-09-08       Impact factor: 3.411

10.  Current Pathologic Scoring Systems for Metal-on-metal THA Revisions are not Reproducible.

Authors:  Christiaan Smeekes; Arjen H G Cleven; Bart C H van der Wal; Stefan V Dubois; Remigio W Rouse; Bastiaan F Ongkiehong; Ron Wolterbeek; Rob G H H Nelissen
Journal:  Clin Orthop Relat Res       Date:  2017-07-07       Impact factor: 4.176

View more
  4 in total

1.  CORR Insights®: Are Serum Metal Ion Levels a Concern at Mid-term Followup of Revision Knee Arthroplasty With a Metal-on-metal Hinge Design?

Authors:  Antti Eskelinen
Journal:  Clin Orthop Relat Res       Date:  2019-09       Impact factor: 4.176

Review 2.  Assistive technologies in knee arthroplasty: fashion or evolution? Rate of publications and national registries prove the Scott Parabola wrong.

Authors:  Cécile Batailler; Sébastien Parratte
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-14       Impact factor: 3.067

3.  Retrieval Analysis of Modern Knee Tumor Megaendoprosthesis Shows Considerable Volumetric Metal Wear Generated at the Rotating Hinge.

Authors:  Therese Bormann; Sebastian Jäger; J Philippe Kretzer; Laura Nebel; Lucas Clarius; Georg Omlor; Rudi Bitsch; Burkhard Lehner
Journal:  Materials (Basel)       Date:  2020-03-26       Impact factor: 3.623

Review 4.  Implant Survival, Clinical Outcome and Complications of Megaprosthetic Reconstructions Following Sarcoma Resection.

Authors:  Christoph Theil; Jan Schwarze; Georg Gosheger; Burkhard Moellenbeck; Kristian Nikolaus Schneider; Niklas Deventer; Sebastian Klingebiel; George Grammatopoulos; Friedrich Boettner; Tom Schmidt-Braekling
Journal:  Cancers (Basel)       Date:  2022-01-11       Impact factor: 6.639

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.