| Literature DB >> 24083615 |
Donato Koyalta1, Mohammad-Ali Jenabian, Barou Djouater, Noël Djemadji-Oudjeil, Francois-Xavier Mbopi-Keou, Angélique Ndjoyi-Mbiguino, Laurent Bélec.
Abstract
BACKGROUND: Field and community evaluation of the routine usage of CD4 T counting platforms is essential in resource-poor countries for efficient and cost-effective monitoring of HIV-infected adults and children attending health care centers.Entities:
Mesh:
Year: 2013 PMID: 24083615 PMCID: PMC3849920 DOI: 10.1186/1472-6963-13-373
Source DB: PubMed Journal: BMC Health Serv Res ISSN: 1472-6963 Impact factor: 2.655
Figure 1Gating strategy for the volumetric Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes in one K-EDTA-anticoagulated blood sample, incubated with phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd): (A) CD45-PE-Dyomics649 axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; (B) CD4-PE axis-light scatter dot plot scattergram; double secondary CD4 gating allows counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; the CD4 T lymphocytes count in absolute number is given by the CD4-positive gating; the CD4 T lymphocytes count in percentage corresponds to the ratio of CD4-positive lymphocytes count in absolute number to the total CD4-postive and CD4-negative lymphocytes counts, corresponding to the sum of both secondary gating.
Figure 2Map showing the location of the 22 officially approved health care centers devoted to follow up adult and children patients infected with HIV-1 thorough Chad in 2013 (Am Timane, Abeche, Adre, Ati, Biltine, Bol, Bongor, Doba, Fada, Faya-Largeau, Gos Beïda, Kelo, Koumra, Lere, Maïssala, Mao, Mongo, Moundou, N’Djamena, Ngouri, Pala and Sahr), as well as the 8 centers equipped by the Auto40 flow cytometers (Apogee Flow Systems Ltd) (2 in N’Djamena at the hôpital Militaire D’Instruction, and the Centre D’Appui Psycho-Médico-Social Al-Nadjma; 1 in Mongo, Bongor, Doba, Kelo, Koumra and Maïssala), and the other 9 centers equipped by FACSCount (Becton Dickinson) (2 in N’Djamena; 1 in Abeche, Adre, Gos Beïda, Moundou, Pala and Sahr), or by CyFlow (Partec) (Adre, Biltine, Maïssala, Mao and Ngouri). Note that 5 health care centers for HIV-1-infected patients (Ati; Bol; Fada; Faya-Largeau; Lere) are not yet equipped for CD4 T cell enumeration. The reference laboratory of the hôpital de la Liberté, N’Djamena, is equipped by FACSCalibur (Becton Dickinson). Note that only the health care centers involved in caring HIV-infected patients are shown in the map.
Main characteristics describing the implementation and the 5 years usage (2008–2013) of 8 single-platform, volumetric, Auto40 flow cytometers for community-based point-of-care CD4 T lymphocytes enumeration installed in Chad
| | | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| | Level of laboratory facility* | 1 | 3 | 2 | 2 | 3 | 2 | 3 | 3 |
| Laboratory characteristics | Situation | Urban | Urban | Rural | Rural | Rural | Urban | rural | Rural |
| | Distance from the capital city NDJ (km) | 0 | 0 | 500 | 230 | 383 | 570 | 750 | 670 |
| | Location in closed dust-free room (yes/no) | Yes | Yes | No | No | No | Yes | No | No |
| | Mean temperature gradient (minimum/maximum in °C) | 22/45 | 22/45 | 22/45 | 20/40 | 20/40 | 20/40 | 20/40 | 22/40 |
| | Air conditioned (yes/no) | Yes | No | No | No | No | Yes | No | |
| Electricity & water | Unreliable electricity power supplies | No | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| | Unreliable water supplies | No | No | No | No | No | No | No | No |
| Human resources | Biologist (number) | No | No | No | No | No | No | No | No |
| | Trained technician (number) | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
| Reagents & consumable supplies | CD45, CD4 lyophilisated reagents | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| | Reagents delivery | Regular | Regular | Regular | Regular | Regular | Regular | Regular | Regular |
| | Reagents price (€) | 10 | 10 | 12 | 11 | 11 | 12 | 12 | 12 |
| | Sheath fluid recycling cassette | No | No | Yes | Yes | Yes | Yes | Yes | Yes |
| | Number of followed up patients (in 2012) | 102 | 458 | 57 | 53 | 198 | 87 | 41 | 157 |
| Medical activity in each site | Number of antiretroviral drugs-treated adult patients (in 2012) | 95 | 421 | 55 | 49 | 137 | 82 | 33 | 145 |
| | Number of antiretroviral drugs-treated children (< 5 years) patients (in 2012) | 7 | 37 | 2 | 4 | 61 | 5 | 4 | 12 |
| | Number of CD4 T cells enumeration** (in 2011) | 341 | 1156 | 55 | 123 | 559 | 230 | 62 | 358 |
| | Final cost of CD4 T cell count (in absolute number and percentage)*** | 11.0 | 11.0 | 13.0 | 12.0 | 12.1 | 13.1 | 13.4 | 13.2 |
| Quality control | Assurance quality procedures | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| | Internal quality control | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| | International external quality control | No | No | No | No | No | No | No | No |
| Instrument Support | Routine maintenance on site | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| | Manufacturer’s maintenance (agreement) | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Troubleshooting | No | No | Yes (electronic memory card) | No | No | No | No | No | |
*Level 1: Reference laboratory; Level 2: District laboratory; Level 3: Health care center laboratory.
**Results of CD4 T cell enumeration with Auto40 flow cytometer are systematically given in absolute count as well as in percentage.
***The estimated final cost of 1 CD4 T cell enumeration does not take into account the cost of the use of sheath fluid recycling cassette in province laboratories, as well as the contractual maintenance of the Auto40 flow cytometers by the distributor (Inodia, France).