BACKGROUND: The echocardiographic assessment of circulatory function in sick newborn infants has the potential to improve patient care. However, measurements are prone to error and have not been sufficiently validated. Phase-contrast magnetic resonance imaging (MRI) provides highly validated measures of blood flow and has recently been applied to the newborn population. The aim of this study was to validate measures of left ventricular output and superior vena caval flow volume in newborn infants.Entities:
Keywords: Echocardiography; LOA; LVO; Left ventricular output; Limits of agreement; MRI; Magnetic resonance imaging; PC; Phase-contrast; Phase-contrast MRI; Preterm infants; RI; Repeatability index; SVC; Superior vena caval; VTI; Velocity-time integral
Mesh:
Year: 2013 PMID: 24075229 PMCID: PMC3852205 DOI: 10.1016/j.echo.2013.08.019
Source DB: PubMed Journal: J Am Soc Echocardiogr ISSN: 0894-7317 Impact factor: 5.251
Figure 1Echocardiographic quantification of LVO, with (A) diameter measured from the parasternal long-axis view and (B) velocity assessed from the apical view. Approximate location of PC MRI acquisition plane is also demonstrated (red box). Ao, Aorta; LV, left ventricle.
Figure 2Echocardiographic quantification of superior vena caval flow, with (A) diameter measured from a modified parasternal long-axis view, (B) area measured from an axial view, and (C) velocity assessed from a low subcostal view. Ao, Aorta; RA, right atrium; SVC, superior vena cava.
Figure 3PC MRI quantification of (A) LVO at the level of the aortic valve and (B) superior vena caval flow at the level of the pulmonary trunk. Ao, Aorta; SVC, superior vena cava.
Mean ± SD of aortic diameter, left ventricular outflow VTI, heart rate, and LVO assessed by echocardiography and PC MRI in 47 newborn infants
| Parameter | Aortic diameter (mm) | Left ventricular stroke distance (cm) | Heart rate for LVO assessment (beats/min) | LVO (mL/kg/min) |
|---|---|---|---|---|
| Echocardiography | 6.0 ± 0.8 | 10.3 ± 2.0 | 143.8 ± 18.7 | 242.1 ± 99.3 |
| PC MRI | 8.7 ± 1.3 | 5.6 ± 2.3 | 141.0 ± 17.4 | 251.8 ± 92.2 |
Figure 4Comparison of LVO quantification by echocardiography and PC MRI in 48 newborn infants. CI, Confidence interval.
Mean ± SD of SVC diameter, SVC VTI, heart rate, and SVC flow assessed by echocardiography and PC MRI in 23 newborn infants
| Parameter | SVC diameter (mm) | SVC stroke distance (cm) | Heart rate for SVC assessment (beats/min) | SVC flow (mL/kg/min) |
|---|---|---|---|---|
| Echocardiography | 3.9 ± 0.9 | 14.5 ± 3.1 | 153.5 ± 17.7 | 118.4 ± 42.5 |
| PC MRI | 4.8 ± 0.8 | 7.3 ± 1.7 | 140.2 ± 19.1 | 101.8 ± 23.8 |
Figure 5Comparison of SVC flow quantification by echocardiography (sagittal view) and PC MRI in 23 newborn infants. CI, Confidence interval.