Factors interfering with the accuracy of five blood glucose meters used in Chinese hospitals.

Hong Lv1, Guo-jun Zhang, Xi-xiong Kang, Hui Yuan, Yan-wei Lv, Wen-wen Wang, Rollins Randall.   

Abstract

BACKGROUND: The prevalence of diabetes is increasing in China. Glucose control is very important in diabetic patients. The aim of this study was to compare the accuracy of five glucose meters used in Chinese hospitals with a reference method, in the absence and presence of various factors that may interfere with the meters.
METHODS: Within-run precision of the meters was evaluated include Roche Accu-Chek Inform®, Abbott Precision PCx FreeStyle®, Bayer Contour®, J&J LifeScan SureStep Flexx®, and Nova Biomedical StatStrip®. The interference of hematocrit level, maltose, ascorbic acid, acetaminophen, galactose, dopamine, and uric acid were tested in three levels of blood glucose, namely low, medium, and high concentrations. Accuracy (bias) of the meters and analytical interference by various factors were evaluated by comparing results obtained in whole blood specimens with those in plasma samples of the whole blood specimens run on the reference method. Impact of oxygen tension on above five blood glucose meters was detected.
RESULTS: Precision was acceptable and slightly different between meters. There were no significant differences in the measurements between the meters and the reference method. The hematocrit level significantly interfered with all meters, except StatStrip. Measurements were affected to varying degrees by different substances at different glucose levels, e.g. acetaminophen and ascorbic acid (Freestyle), maltose and galactose (FreeStyle, Accu-Chek), uric acid (FreeStyle, Bayer Contour), and dopamine (Bayer Contour).
CONCLUSIONS: The measurements with the five meters showed a good correlation with the plasma hexokinase reference method, but most were affected by the hematocrit level. Some meters also showed marked interference by other substances.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  blood glucose; correlation; hematocrit; hexokinase

Mesh:

Substances:

Year:  2013        PMID: 24038220      PMCID: PMC6807623          DOI: 10.1002/jcla.21611

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  10 in total

1.  Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

Authors:  Z Tang; X Du; R F Louie; G J Kost
Journal:  Am J Clin Pathol       Date:  2000-01       Impact factor: 2.493

2.  Technical evaluation of five glucose meters with data management capabilities.

Authors:  J J Chance; D J Li; K A Jones; K L Dyer; J H Nichols
Journal:  Am J Clin Pathol       Date:  1999-04       Impact factor: 2.493

3.  Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.

Authors:  Z Tang; J H Lee; R F Louie; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-08       Impact factor: 5.534

4.  Intensive insulin therapy in critically ill patients.

Authors:  G van den Berghe; P Wouters; F Weekers; C Verwaest; F Bruyninckx; M Schetz; D Vlasselaers; P Ferdinande; P Lauwers; R Bouillon
Journal:  N Engl J Med       Date:  2001-11-08       Impact factor: 91.245

5.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

6.  Prevalence of diabetes among men and women in China.

Authors:  Wenying Yang; Juming Lu; Jianping Weng; Weiping Jia; Linong Ji; Jianzhong Xiao; Zhongyan Shan; Jie Liu; Haoming Tian; Qiuhe Ji; Dalong Zhu; Jiapu Ge; Lixiang Lin; Li Chen; Xiaohui Guo; Zhigang Zhao; Qiang Li; Zhiguang Zhou; Guangliang Shan; Jiang He
Journal:  N Engl J Med       Date:  2010-03-25       Impact factor: 91.245

7.  Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.

Authors:  Brad S Karon; Laurie Griesmann; Renee Scott; Sandra C Bryant; Jeffrey A Dubois; Terry L Shirey; Steven Presti; Paula J Santrach
Journal:  Diabetes Technol Ther       Date:  2008-04       Impact factor: 6.118

8.  Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose.

Authors:  J C Boyd; D E Bruns
Journal:  Clin Chem       Date:  2001-02       Impact factor: 8.327

9.  Outcomes and perioperative hyperglycemia in patients with or without diabetes mellitus undergoing coronary artery bypass grafting.

Authors:  Carlos A Estrada; James A Young; L Wiley Nifong; W Randolph Chitwood
Journal:  Ann Thorac Surg       Date:  2003-05       Impact factor: 4.330

10.  Effect of hyperglycemia and continuous intravenous insulin infusions on outcomes of cardiac surgical procedures: the Portland Diabetic Project.

Authors:  Anthony P Furnary; YingXing Wu; Stephen O Bookin
Journal:  Endocr Pract       Date:  2004 Mar-Apr       Impact factor: 3.443

  10 in total
  3 in total

1.  A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood.

Authors:  Huiping Wei; Fang Lan; Qitian He; Haiwei Li; Fuyong Zhang; Xue Qin; Shan Li
Journal:  J Clin Lab Anal       Date:  2016-08-25       Impact factor: 2.352

Review 2.  Approach to the interpretation of unexpected laboratory results arising in the care of patients with inborn errors of metabolism (IEM).

Authors:  Andre Mattman; Murray Potter
Journal:  Rev Endocr Metab Disord       Date:  2018-03       Impact factor: 6.514

Review 3.  Assessment of glycemia in chronic kidney disease.

Authors:  Mohamed Hassanein; Tariq Shafi
Journal:  BMC Med       Date:  2022-04-13       Impact factor: 8.775

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.