| Literature DB >> 24036870 |
Masashi Yagi1, Takashi Ueguchi, Masahiko Koizumi, Toshiyuki Ogata, Sachiko Yamada, Yutaka Takahashi, Iori Sumida, Yuichi Akino, Koji Konishi, Fumiaki Isohashi, Noriyuki Tomiyama, Yasuo Yoshioka, Kazuhiko Ogawa.
Abstract
The monochromatic images acquired by Gemstone spectral imaging (GSI) mode on the GE CT750 HD theoretically determines the computed tomography (CT) number more accurately than that of conventional scanner. Using the former, the CT number is calculated from (synthesized) monoenergetic X-ray data. We reasoned that the monochromatic image might be applied to radiotherapy treatment planning (RTP) to calculate dose distribution more accurately. Our goal here was to provide CT to electron density (ED) conversion curves with monochromatic images for RTP. Therefore, we assessed the reproducibility of CT numbers, an important factor on quality assurance, over short and long time periods for different substances at varying energy. CT number difference between measured and theoretical value was investigated. The scanner provided sufficient reproducibility of CT numbers for dose calculation over short and long time periods. The CT numbers of monochromatic images produced by this scanner had reasonable values for dose calculation. The CT to ED conversion curve becomes linear with respect to the relationship between CT numbers and EDs as the energy increases. We conclude that monochromatic imaging from a fast switching system can be applied for the dose calculation, keeping Hounsfield units (HU) stability.Entities:
Mesh:
Year: 2013 PMID: 24036870 PMCID: PMC5714581 DOI: 10.1120/jacmp.v14i5.4335
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Physical characteristics of rods. The data are listed in ascending order by electron density
|
| Electron Density Relative to Water | Physical Density |
|---|---|---|
| LN‐300 Lung | 0.284 | 0.290 |
| LN‐450 Lung | 0.445 | 0.460 |
| AP6 Adipose | 0.924 | 0.941 |
| BR‐12 Breast | 0.957 | 0.980 |
| CT Solid Water | 0.988 | 1.017 |
| Water Insert | 1.000 | 1.000 |
| BRN‐SR2 Brain | 1.049 | 1.053 |
| LV1 Liver | 1.062 | 1.094 |
| IB Inner Bone | 1.097 | 1.144 |
| B200 Bone Mineral | 1.096 | 1.143 |
|
| 1.279 | 1.334 |
|
| 1.470 | 1.560 |
| SB3 Cortical Bone | 1.696 | 1.824 |
Figure 1Phantom rod alignment. The high‐density materials were positioned to minimize artifacts.
Scan parameters
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| Regular | Helical | 120 | 630 | 0.5 | Large body | 2.5 | 40 |
| GSI | Helical | 80/140 | 600 | 1.0 | Large body | 2.5 | 40 |
SFOV = scan field of view.
Reconstruction parameters of the two imaging modes
|
|
|
|
|
|---|---|---|---|
| Regular | 50 | Standard | N/A |
| GSI | 50 | Standard | Mono 60–140 keV |
DFOV = display field of view, N/A = not applicable.
Figure 2Overview of polychromatic and monochromatic images: a) 120 kVp, b) 60 keV, c) 77 keV, d) 100keV, e) 140 keV The arrow indicates an artifact appearing as a dark band around the cortical bone rod.
Figure 3Distribution of CT number among materials as a function of monochromatic energy.
Figure 4Standard deviations of CT numbers over a short time period.
Figure 5Standard deviations of CT numbers over a long time period.
Figure 6CT number difference between monochromatic image (measured) and theoretical (true) value in soft tissues.
Figure 7CT number difference between monochromatic image (measured) and theoretical (true) value in bony materials.
Figure 8CT to ED conversion curves obtained by monochromatic images. The cutout shows the plots around 0 HU.