Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner.

Mitchell M Goodsitt1, Emmanuel G Christodoulou, Sandra C Larson.   

Abstract

PURPOSE: This study was performed to investigate the accuracies of the synthesized monochromatic images and effective atomic number maps obtained with the new GE Discovery CT750 HD CT scanner.
METHODS: A Gammex-RMI model 467 tissue characterization phantom and the CT number linearity section of a Phantom Laboratory Catphan 600 phantom were scanned using the dual energy (DE) feature on the GE CT750 HD scanner. Synthesized monochromatic images at various energies between 40 and 120 keV and effective atomic number (Z(eff)) maps were generated. Regions of interest were placed within these images/maps to measure the average monochromatic CT numbers and average Z(eff) of the materials within these phantoms. The true Z(eff) values were either supplied by the phantom manufacturer or computed using Mayneord's equation. The linear attenuation coefficients for the true CT numbers were computed using the NIST XCOM program with the input of manufacturer supplied elemental compositions and densities. The effects of small variations in the assumed true densities of the materials were also investigated. Finally, the effect of body size on the accuracies of the synthesized monochromatic CT numbers was investigated using a custom lumbar section phantom with and without an external fat-mimicking ring.
RESULTS: Other than the Z(eff) of the simulated lung inserts in the tissue characterization phantom, which could not be measured by DECT, the Z(eff) values of all of the other materials in the tissue characterization and Catphan phantoms were accurate to 15%. The accuracies of the synthesized monochromatic CT numbers of the materials in both phantoms varied with energy and material. For the 40-120 keV range, RMS errors between the measured and true CT numbers in the Catphan are 8-25 HU when the true CT numbers were computed using the nominal plastic densities. These RMS errors improve to 3-12 HU for assumed true densities within the nominal density +/- 0.02 g/cc range. The RMS errors between the measured and true CT numbers of the tissue mimicking materials in the tissue characterization phantom over the 40-120 keV range varied from about 6 HU-248 HU and did not improve as dramatically with small changes in assumed true density.
CONCLUSIONS: Initial tests indicate that the Z(eff) values computed with DECT on this scanner are reasonably accurate; however, the synthesized monochromatic CT numbers can be very inaccurate, especially for dense tissue mimicking materials at low energies. Furthermore, the synthesized monochromatic CT numbers of materials still depend on the amount of the surrounding tissues especially at low keV, demonstrating that the numbers are not truly monochromatic. Further research is needed to develop DE methods that produce more accurate synthesized monochromatic CT numbers.

Entities:  

Mesh:

Year:  2011        PMID: 21626956     DOI: 10.1118/1.3567509

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  57 in total

1.  Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality.

Authors:  Lifeng Yu; Jodie A Christner; Shuai Leng; Jia Wang; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Diagnostic performance of dual-energy CT and subtraction CT for renal lesion detection and characterization.

Authors:  Ali Pourvaziri; Anushri Parakh; Amirkasra Mojtahed; Avinash Kambadakone; Dushyant Vasudeo Sahani
Journal:  Eur Radiol       Date:  2019-05-27       Impact factor: 5.315

3.  Prospects for in vivo estimation of photon linear attenuation coefficients using postprocessing dual-energy CT imaging on a commercial scanner: comparison of analytic and polyenergetic statistical reconstruction algorithms.

Authors:  Joshua D Evans; Bruce R Whiting; Joseph A O'Sullivan; David G Politte; Paul H Klahr; Yaduo Yu; Jeffrey F Williamson
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

4.  Assessment of carotid plaque composition using fast-kV switching dual-energy CT with gemstone detector: comparison with extracorporeal and virtual histology-intravascular ultrasound.

Authors:  Yuki Shinohara; Makoto Sakamoto; Keita Kuya; Junichi Kishimoto; Naoki Iwata; Yasutoshi Ohta; Shinya Fujii; Takashi Watanabe; Toshihide Ogawa
Journal:  Neuroradiology       Date:  2015-05-16       Impact factor: 2.804

5.  Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review.

Authors:  Anne T Davis; Antony L Palmer; Andrew Nisbet
Journal:  Br J Radiol       Date:  2017-05-23       Impact factor: 3.039

6.  Preliminary study on the differentiation between parapelvic cyst and hydronephrosis with non-calculous using only pre-contrast dual-energy spectral CT scans.

Authors:  Dong Han; Guangming Ma; Lequn Wei; Chenglong Ren; Jieli Zhou; Chen Shen; Taiping He
Journal:  Br J Radiol       Date:  2017-03-10       Impact factor: 3.039

7.  Spectral detector CT for cardiovascular applications.

Authors:  Prabhakar Rajiah; Suhny Abbara; Sandra Simon Halliburton
Journal:  Diagn Interv Radiol       Date:  2017 May-Jun       Impact factor: 2.630

8.  Spectral performance of a whole-body research photon counting detector CT: quantitative accuracy in derived image sets.

Authors:  Shuai Leng; Wei Zhou; Zhicong Yu; Ahmed Halaweish; Bernhard Krauss; Bernhard Schmidt; Lifeng Yu; Steffen Kappler; Cynthia McCollough
Journal:  Phys Med Biol       Date:  2017-08-21       Impact factor: 3.609

9.  Single-energy metal artifact reduction technique for reducing metallic coil artifacts on post-interventional cerebral CT and CT angiography.

Authors:  Masaki Katsura; Jiro Sato; Masaaki Akahane; Taku Tajima; Toshihiro Furuta; Harushi Mori; Osamu Abe
Journal:  Neuroradiology       Date:  2018-08-24       Impact factor: 2.804

10.  Changes in measured size of atherosclerotic plaque calcifications in dual-energy CT of ex vivo carotid endarterectomy specimens: effect of monochromatic keV image reconstructions.

Authors:  Lorenzo Mannelli; Lee M Mitsumori; Marina Ferguson; Dongxiang Xu; Baocheng Chu; Kelley R Branch; William P Shuman; Chun Yuan
Journal:  Eur Radiol       Date:  2012-08-21       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.