Analysis of "Laboratory and Benchtop Performance of a Mealtime Insulin Delivery System".

Eric Zijlstra1.   

Abstract

In the current issue of Journal of Diabetes Science and Technology, Dreon et al give a comprehensive overview of the technical performance of a wearable patch for bolus insulin delivery. The test results generated by the manufacturer of the bolus-patch provide the technical prerequisites for clinical application. As the device received FDA clearance already in 2010, positive results from the nonclinical performance testing were to be expected, but present nevertheless interesting insights into the device development. The single-dose accuracy verification results seem especially promising, but tighter accuracy criteria could have been specified and information on outliers is missing from the analysis. The clinical application of the bolus-patch is currently under investigation in a large-scale 44-week intervention trial.

Entities:  

Keywords:  diabetes; dose accuracy; insulin bolus

Mesh:

Substances:

Year:  2018        PMID: 29619894      PMCID: PMC6134312          DOI: 10.1177/1932296818767940

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  17 in total

1.  Comparison of a novel insulin bolus-patch with pen/syringe injection to deliver mealtime insulin for efficacy, preference, and quality of life in adults with diabetes: a randomized, crossover, multicenter study.

Authors:  Nancy Bohannon; Richard Bergenstal; Robert Cuddihy; Davida Kruger; Susan List; Elaine Massaro; Mark Molitch; Philip Raskin; Heather Remtema; Suzanne Strowig; Fred Whitehouse; Rocco L Brunelle; Darlene Dreon; Meng Tan
Journal:  Diabetes Technol Ther       Date:  2011-07-06       Impact factor: 6.118

2.  Smart Pens Will Improve Insulin Therapy.

Authors:  David C Klonoff; David Kerr
Journal:  J Diabetes Sci Technol       Date:  2018-02-07

3.  Patient Perceptions and Preferences for a Mealtime Insulin Delivery Patch.

Authors:  Mark Peyrot; Darlene Dreon; Vivien Zraick; Brett Cross; Meng H Tan
Journal:  Diabetes Ther       Date:  2018-01-11       Impact factor: 2.945

4.  Insulin pump use in type 2 diabetes.

Authors:  Bruce W Bode
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

5.  Effect of glargine insulin delivery method (pen device versus vial/syringe) on glycemic control and patient preferences in patients with type 1 and type 2 diabetes.

Authors:  Stacey A Seggelke; R Matthew Hawkins; Joanna Gibbs; Neda Rasouli; Cecilia C Low Wang; Boris Draznin
Journal:  Endocr Pract       Date:  2014-06       Impact factor: 3.443

Review 6.  A systematic review of adherence with medications for diabetes.

Authors:  Joyce A Cramer
Journal:  Diabetes Care       Date:  2004-05       Impact factor: 19.112

7.  Comparative dose accuracy of durable and patch insulin infusion pumps.

Authors:  Luis G Jahn; Jorge J Capurro; Brian L Levy
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

8.  Injecting without pressing a button: An exploratory study of a shield-triggered injection mechanism.

Authors:  Eric Zijlstra; Hans-Veit Coester; Tim Heise; Leona Plum-Mörschel; Ole Rasmussen; Tord Rikte; Line Kynemund Pedersen; Marianne Qvist; Thomas Sparre
Journal:  Diabetes Obes Metab       Date:  2018-01-25       Impact factor: 6.577

9.  Characteristics Relating to Adherence and Persistence to Basal Insulin Regimens Among Elderly Insulin-Naïve Patients with Type 2 Diabetes: Pre-Filled Pens versus Vials/Syringes.

Authors:  S Lane Slabaugh; Jonathan R Bouchard; Yong Li; Jean C Baltz; Yunus A Meah; D Chad Moretz
Journal:  Adv Ther       Date:  2015-11-13       Impact factor: 3.845

10.  Complex Assessment of Metabolic Effectiveness of Insulin Pump Therapy in Patients with Type 2 Diabetes Beyond HbA1c Reduction.

Authors:  Rudolf Chlup; Sarah Runzis; Javier Castaneda; Scott W Lee; Xuan Nguyen; Ohad Cohen
Journal:  Diabetes Technol Ther       Date:  2017-12-07       Impact factor: 6.118

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.