Accuracy of Bolus and Basal Rate Delivery of Different Insulin Pump Systems.

Guido Freckmann1, Ulrike Kamecke1, Delia Waldenmaier1, Cornelia Haug1, Ralph Ziegler2.   

Abstract

BACKGROUND: Insulin pumps are used for basal rate and bolus insulin delivery in patients with diabetes. In this in vitro study, accuracy of delivery of different commercial insulin pumps was evaluated.
MATERIALS AND METHODS: Accuracy of 10 different insulin pump systems (5 durable pumps with different insulin infusion sets and 1 patch pump) was tested with a microgravimetric method. Mean bolus accuracy of 25 successive 1 U boluses and of 12 successive 10 U boluses was assessed, and delivery time for 10 U boluses was measured. Basal rate accuracy at 1.0 U/h was evaluated for 72 h and for individual 1-h windows.
RESULTS: Mean bolus delivery was within ±5% of target for both tested bolus sizes, but precision of individual boluses was higher with the larger boluses. Delivery times varied between the different pump models but agreed with the specifications of the respective manufacturers. Regarding basal rate accuracy, the total deviation for 72 h was very small in all pumps; however, larger deviations were observed during the first 12 h. For the patch pump, large variations between individual 1-h windows were observed.
CONCLUSIONS: In general, all compared insulin pump systems showed a similar level of accuracy. Differences, especially between durable pumps and the patch pump, were observed when considering each hour of basal rate delivery separately.

Entities:  

Keywords:  Accuracy; Basal rate; Bolus; Continuous subcutaneous insulin infusion; Insulin pump

Mesh:

Substances:

Year:  2019        PMID: 30901232      PMCID: PMC6477586          DOI: 10.1089/dia.2018.0376

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  17 in total

1.  Establishing Methods to Determine Clinically Relevant Bolus and Basal Rate Delivery Accuracy of Insulin Pumps.

Authors:  Ulrike Kamecke; Delia Waldenmaier; Cornelia Haug; Ralph Ziegler; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-07-13

2.  In children using intensive insulin therapy, a 20-g variation in carbohydrate amount significantly impacts on postprandial glycaemia.

Authors:  C E Smart; B R King; P McElduff; C E Collins
Journal:  Diabet Med       Date:  2012-07       Impact factor: 4.359

3.  Consensus Statement by the American Association of Clinical Endocrinologists/American College of Endocrinology insulin pump management task force.

Authors:  George Grunberger; Jill M Abelseth; Timothy S Bailey; Bruce W Bode; Yehuda Handelsman; Richard Hellman; Lois Jovanovič; Wendy S Lane; Philip Raskin; William V Tamborlane; Caitlin Rothermel
Journal:  Endocr Pract       Date:  2014-05       Impact factor: 3.443

4.  Accurate Carbohydrate Counting Is an Important Determinant of Postprandial Glycemia in Children and Adolescents With Type 1 Diabetes on Insulin Pump Therapy.

Authors:  Asma Deeb; Ahlam Al Hajeri; Iman Alhmoudi; Nico Nagelkerke
Journal:  J Diabetes Sci Technol       Date:  2016-11-21

5.  Novel methodology to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system.

Authors:  Howard Zisser; Marc Breton; Eyal Dassau; Kateryna Markova; Wendy Bevier; Dale Seborg; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

Review 6.  Continuous subcutaneous insulin infusion therapy and multiple daily insulin injections in type 1 diabetes mellitus: a comparative overview and future horizons.

Authors:  Hood Thabit; Roman Hovorka
Journal:  Expert Opin Drug Deliv       Date:  2015-11-30       Impact factor: 6.648

7.  Comparative dose accuracy of durable and patch insulin infusion pumps.

Authors:  Luis G Jahn; Jorge J Capurro; Brian L Levy
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

8.  Accuracy and precision of low-dose insulin administration using syringes, pen injectors, and a pump.

Authors:  Katherine Keith; David Nicholson; Douglas Rogers
Journal:  Clin Pediatr (Phila)       Date:  2004 Jan-Feb       Impact factor: 1.168

9.  Quantifying the impact of a short-interval interruption of insulin-pump infusion sets on glycemic excursions.

Authors:  Howard Zisser
Journal:  Diabetes Care       Date:  2007-12-04       Impact factor: 19.112

10.  A Comparative Pulse Accuracy Study of Two Commercially Available Patch Insulin Infusion Pumps.

Authors:  Jenna L Bowen; Chris J Allender
Journal:  Eur Endocrinol       Date:  2016-08-28
View more
  3 in total

Review 1.  Insulin Pumps.

Authors:  Rayhan Lal; Lalantha Leelarathna
Journal:  Diabetes Technol Ther       Date:  2020-02       Impact factor: 6.118

2.  Insulin Delivery Hardware: Pumps and Pens.

Authors:  Rayhan Lal; Lalantha Leelarathna
Journal:  Diabetes Technol Ther       Date:  2022-03       Impact factor: 7.337

3.  Description of a Novel Patch Pump for Insulin Delivery and Comparative Accuracy Evaluation.

Authors:  Annette Baumstark; Jochen Mende; Joji Uchiyama; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2021-03-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.