| Literature DB >> 23857654 |
Panagiota N Eleni1, Magdalini K Krokida, Gregory L Polyzois, Lawrence Gettleman.
Abstract
OBJECTIVE: Disinfection procedures often cause deterioration in a maxillofacial prosthesis. Color and hardness alterations could lead to a replacement of the prosthesis.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23857654 PMCID: PMC3881904 DOI: 10.1590/1679-775720130112
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
Maxillofacial elastomers studied
|
|
|
|
|
| PDMS | A | Multisil Epithetik /Commercial addition curing | Bredent GmbH, Senden, Germany |
| CPE | B | Experimental Chlorinated Polyethylene | Department of Oral Health & Rehabilitation, School of Dentistry, University of Louisville (501 S. Preston St, Louisville, KY 40202-1701 USA) |
Disinfection procedures applied
|
|
|
|
|
| Microwave exposure (Panasonic, Model: NN 5256B, 2450 MHz, Matsushita Electric Industrial Co. Ltd., Osaka, Japan). | 365 times*3min | 1 year | DP1 |
| Sodium hypochlorite solution (1% w/w) | 30 hours | 1 year | DP2 |
| Neutral soap (Ns -Johnson & Johnson, GmbH, Italy) | 30 hours | 1 year | DP3 |
| Commercial disinfecting solution (Daro B-200-09, Technovent Ltd., South Wales, UK) | 30 hours | 1 year | DP4 |
Two-way analysis of variance results for ΔH and ΔE measurements
|
|
|
|
|
|
| |
|
| ||||||
| Disinfection procedures | ΔH | 0.691 | 3 | 0.230 | 1.789 | 0.169 |
| ΔE | 37.67 | 3 | 12.559 | 40.143 | 0.000 | |
| Material | ΔH | 1.757 | 1 | 1.757 | 13.651 | 0.001 |
| ΔE | 4.536 | 1 | 4.536 | 14.498 | 0.001 | |
| Interaction (Disinfection Procedure*Material) | ΔH | 1.428 | 3 | 0.476 | 3.697 | 0.022 |
| ΔE | 38.255 | 3 | 12.752 | 40.758 | 0.000 | |
| Error | ΔH | 4.119 | 32 | 0.129 | ||
| ΔE | 10.012 | 32 | 0.313 |
ΔE mean values and standard deviations for material A and B after the different disinfection procedures and Tukey's multiple comparisons results, (n=40)
|
|
|
|
|
|
| |
| DP1 | 1.51ª±0.65 | 3.54c±0.71 |
| DP2 | 4.15ª±0.26 | 1.54d±0.15 |
| DP3 | 3.64b±0.47 | 5.92e±0.72 |
| DP4 | 3.91b±0.56 | 4.90c,e±0.66 |
*a, b, c, d, e: Means with the same letter in columns for respective type of disinfection procedure are not significantly different (p>0.05). Values reported are an average of 5 specimens
Figure 4Mean values of hardness measurements and standard deviation before (grey columns) and after (black columns) each disinfection procedure for material a. A and b. B