Literature DB >> 23811979

Spontaneous delayed migration/shortening of the pipeline embolization device: report of 5 cases.

N Chalouhi1, S I Tjoumakaris, L F Gonzalez, D Hasan, P J Pema, G Gould, R H Rosenwasser, P M Jabbour.   

Abstract

Five patients were found to have spontaneous delayed migration/shortening of their Pipeline Embolization Devices on follow-up angiography. The device migrated proximally in 4 patients and distally in 1 patient. One patient had a subarachnoid hemorrhage and died as a result of migration of the Pipeline Embolization Device, and another patient presented with complete MCA occlusion and was left severely disabled. Mismatch in arterial diameter between inflow and outflow vessels was a constant finding. Migration of the Pipeline Embolization Device was managed conservatively, with additional placement of the device, or with parent vessel occlusion. Obtaining complete expansion of the embolization device by using a longer device, increasing vessel coverage, using adjunctive aneurysm coiling, and avoiding dragging and stretching of the device are important preventive measures. Neurointerventionalists should be aware of this potentially fatal complication and take all necessary preventive measures.

Entities:  

Mesh:

Year:  2013        PMID: 23811979      PMCID: PMC7965201          DOI: 10.3174/ajnr.A3632

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  17 in total

1.  Stent-assisted coiling for aneurysms and the phenomenon of stent migration: is it worth it?

Authors:  Charles J Prestigiacomo
Journal:  J Neurointerv Surg       Date:  2010-12       Impact factor: 5.836

2.  Proximal migration and compaction of an Enterprise stent into a coiled basilar apex aneurysm: a posterior circulation phenomenon?

Authors:  Shervin R Dashti; David Fiorella; Maria M Toledo; Yin Hu; Cameron G McDougall; Felipe C Albuquerque
Journal:  J Neurointerv Surg       Date:  2010-08-23       Impact factor: 5.836

3.  Possible mechanisms for delayed migration of the closed cell--designed enterprise stent when used in the adjunctive treatment of a basilar artery aneurysm.

Authors:  B Gao; A M Malek
Journal:  AJNR Am J Neuroradiol       Date:  2010-09-30       Impact factor: 3.825

4.  Delayed migration of a self-expanding intracranial microstent.

Authors:  M E Kelly; R D Turner; S I Moskowitz; V Gonugunta; M S Hussain; D Fiorella
Journal:  AJNR Am J Neuroradiol       Date:  2008-08-21       Impact factor: 3.825

5.  Early migration of a self-expanding intracranial stent after the treatment of a basilar trunk aneurysm: report of a second case.

Authors:  Rakesh Khatri; Gustavo J Rodriguez; Farhan Siddiq; Ramachandra P Tummala
Journal:  Neurosurgery       Date:  2011-08       Impact factor: 4.654

6.  Treatment of intracranial aneurysms using the pipeline flow-diverter embolization device: a single-center experience with long-term follow-up results.

Authors:  I Saatci; K Yavuz; C Ozer; S Geyik; H S Cekirge
Journal:  AJNR Am J Neuroradiol       Date:  2012-07-19       Impact factor: 3.825

7.  Intracranial aneurysms: midterm outcome of pipeline embolization device--a prospective study in 143 patients with 178 aneurysms.

Authors:  Simon Chun-Ho Yu; Ching-Kwong Kwok; Pui-Wai Cheng; Kwong-Yau Chan; Samuel Shun Lau; Wai-Man Lui; Ka-Ming Leung; Raymand Lee; Harold Kin-Ming Cheng; Yuk-Ling Cheung; Chi-Ming Chan; George Kwok-Chu Wong; Joyce Wai-Yi Hui; Yiu-Chung Wong; Chong-Boon Tan; Wai-Lun Poon; Kai-Yuen Pang; Alain Kai-Sing Wong; Kai-Hung Fung
Journal:  Radiology       Date:  2012-09-20       Impact factor: 11.105

8.  Immediate and midterm results following treatment of unruptured intracranial aneurysms with the pipeline embolization device.

Authors:  W McAuliffe; V Wycoco; H Rice; C Phatouros; T J Singh; J Wenderoth
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-06       Impact factor: 3.825

9.  Spontaneous delayed proximal migration of enterprise stent after staged treatment of wide-necked basilar aneurysm: technical case report.

Authors:  Sean D Lavine; Philip M Meyers; E Sander Connolly; Robert S Solomon
Journal:  Neurosurgery       Date:  2009-05       Impact factor: 4.654

10.  Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience.

Authors:  Pedro Lylyk; Carlos Miranda; Rosana Ceratto; Angel Ferrario; Esteban Scrivano; Hugh Ramirez Luna; Aaron L Berez; Quang Tran; Peter K Nelson; David Fiorella
Journal:  Neurosurgery       Date:  2009-04       Impact factor: 4.654

View more
  34 in total

1.  Spontaneous regression of intracranial aneurysm following remote ruptured aneurysm treatment with pipeline stent assisted coiling.

Authors:  Asterios Tsimpas; William W Ashley; Anand V Germanwala
Journal:  BMJ Case Rep       Date:  2015-08-13

2.  Flow Diversion versus Standard Endovascular Techniques for the Treatment of Unruptured Carotid-Ophthalmic Aneurysms.

Authors:  F Di Maria; S Pistocchi; F Clarençon; B Bartolini; R Blanc; A Biondi; H Redjem; J Chiras; N Sourour; M Piotin
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

3.  Initial Experience with p64: A Novel Mechanically Detachable Flow Diverter for the Treatment of Intracranial Saccular Sidewall Aneurysms.

Authors:  S Fischer; M Aguilar-Pérez; E Henkes; W Kurre; O Ganslandt; H Bäzner; H Henkes
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

4.  Focal, transient mechanical narrowing of a pipeline embolization device following treatment of an internal carotid artery aneurysm.

Authors:  Timothy Ryan Miller; Gaurav Jindal; Dheeraj Gandhi
Journal:  BMJ Case Rep       Date:  2014-09-29

5.  Computation of the change in length of a braided device when deployed in realistic vessel models.

Authors:  Hector Fernandez; Juan M Macho; Jordi Blasco; Luis San Roman; Werner Mailaender; Luis Serra; Ignacio Larrabide
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-06-11       Impact factor: 2.924

6.  Endovascular techniques for achievement of better flow diverter wall apposition.

Authors:  Anna L Kühn; Katyucia de Macedo Rodrigues; Ajay K Wakhloo; Ajit S Puri
Journal:  Interv Neuroradiol       Date:  2018-11-28       Impact factor: 1.610

7.  Deployment of flow diverter devices: prediction of foreshortening and validation of the simulation in 18 clinical cases.

Authors:  Robert Kellermann; Steffen Serowy; Oliver Beuing; Martin Skalej
Journal:  Neuroradiology       Date:  2019-08-31       Impact factor: 2.804

8.  A single pipeline embolization device is sufficient for treatment of intracranial aneurysms.

Authors:  N Chalouhi; S Tjoumakaris; J L H Phillips; R M Starke; D Hasan; C Wu; M Zanaty; D Kung; L F Gonzalez; R Rosenwasser; P Jabbour
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-30       Impact factor: 3.825

9.  Persistent aneurysm growth following pipeline embolization device assisted coiling of a fusiform vertebral artery aneurysm: a word of caution!

Authors:  Mena Kerolus; Manish K Kasliwal; Demetrius K Lopes
Journal:  Neurointervention       Date:  2015-02-28

10.  Delayed rupture of intracranial aneurysms after placement of intra-luminal flow diverter.

Authors:  Kun Hou; Guichen Li; Xianli Lv; Baofeng Xu; Kan Xu; Jinlu Yu
Journal:  Neuroradiol J       Date:  2020-08-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.