Scatter-to-primary ratio in dentomaxillofacial cone-beam CT: effect of field of view and beam energy.

Ruben Pauwels1,2, Pisha Pittayapat2, Phonkit Sinpitaksakul2, Soontra Panmekiate2.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the effect of field of view (FOV) and beam energy on the scatter-to-primary ratio (SPR) in dental cone-beam CT (CBCT).
METHODS: An anthropomorphic phantom representing an adult male (ATOM Max 711-HN, Norfolk, VA, USA) was scanned using the 3D Accuitomo 170 CBCT (J. Morita, Kyoto, Japan) using 11 FOVs. During each scan, half of the X-ray beam was blocked. Each scan was performed at three exposure settings with varying beam energy and equal radiation dose: 90 kV 5 mA, 77 kV 7.5 mA and 69 kV 10 mA. The SPR was estimated by measuring the grey values in the blocked and non-blocked regions of the RAW data. The effect of FOV on SPR was evaluated using Dunn's multiple comparison test, and the effect of the exposure settings was compared using a Wilcoxon signed rank test.
RESULTS: Larger FOVs showed increased scatter. FOVs with a shorter isocenter-detector distance showed a particularly high SPR. Most intercomparisons between FOVs were statistically significant. The largest difference was found between 17 × 12 cm and 6 × 6 cm (lower jaw), with the former showing a 4.9-fold higher SPR. The effect of beam energy was relatively small and varied between FOV sizes and positions.
CONCLUSION: While the choice of FOV size and position is determined by the diagnostic region of interest, the image quality deterioration for large FOVs due to scatter provides another incentive to limit the FOV size as much as possible.

Entities:  

Keywords:  computer-assisted image analysis; cone-beam computed tomography; dentistry; image quality enhancement; radiation protection

Mesh:

Year:  2021        PMID: 33882256      PMCID: PMC8611284          DOI: 10.1259/dmfr.20200597

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  25 in total

1.  Characteristics of x-ray beams in two commercial multidetector computed tomography simulators: Monte Carlo simulations.

Authors:  Sangroh Kim; Haijun Song; Benjamin Movsas; Indrin J Chetty
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Incidental findings based on anatomical location and clinical significance in CBCT scans of dental implant patients.

Authors:  Sunil Mutalik; Kandasamy Rengasamy; Aditya Tadinada
Journal:  Quintessence Int       Date:  2018       Impact factor: 1.677

3.  Technical note: comparing coherent and incoherent scatter effects for cone-beam CT.

Authors:  Yiannis Kyriakou; Michael Meyer; Willi A Kalender
Journal:  Phys Med Biol       Date:  2008-04-25       Impact factor: 3.609

4.  Incidental findings on cone-beam computed tomographic images: paranasal sinus findings and nasal septum variations.

Authors:  Hakan Avsever; Kaan Gunduz; Omer Karakoç; Mesut Akyol; Kaan Orhan
Journal:  Oral Radiol       Date:  2017-03-27       Impact factor: 1.852

5.  Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation.

Authors:  R Pauwels; G Zhang; C Theodorakou; A Walker; H Bosmans; R Jacobs; R Bogaerts; K Horner
Journal:  Br J Radiol       Date:  2014-10       Impact factor: 3.039

6.  Optimal combination of anti-scatter grids and software correction for CBCT imaging.

Authors:  Uros Stankovic; Lennert S Ploeger; Marcel van Herk; Jan-Jakob Sonke
Journal:  Med Phys       Date:  2017-08-02       Impact factor: 4.071

7.  A Monte Carlo based scatter removal method for non-isocentric cone-beam CT acquisitions using a deep convolutional autoencoder.

Authors:  Brent van der Heyden; Martin Uray; Gabriel Paiva Fonseca; Philipp Huber; Defne Us; Ivan Messner; Adam Law; Anastasiia Parii; Niklas Reisz; Ilaria Rinaldi; Gloria Vilches-Freixas; Heinz Deutschmann; Frank Verhaegen; Phil Steininger
Journal:  Phys Med Biol       Date:  2020-04-15       Impact factor: 3.609

8.  A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation dose.

Authors:  R Pauwels; O Silkosessak; R Jacobs; R Bogaerts; H Bosmans; S Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2014-04-08       Impact factor: 2.419

9.  Accuracy of detecting vertical root fractures in non-root filled teeth using cone beam computed tomography: effect of voxel size and fracture width.

Authors:  X L Guo; G Li; J Q Zheng; R H Ma; F C Liu; F S Yuan; P J Lyu; Y J Guo; S Yin
Journal:  Int Endod J       Date:  2019-02-12       Impact factor: 5.264

10.  Guest editorial--Self-referral: an ethical concern with respect to multidimensional imaging in dentistry?

Authors:  Allan G Farman
Journal:  J Appl Oral Sci       Date:  2009 Sep-Oct       Impact factor: 2.698

View more
  1 in total

1.  Influence of head positioning during cone-beam CT imaging on the accuracy of virtual 3D models.

Authors:  Maureen van Eijnatten; Jan Wolff; Ruben Pauwels; Kalle Karhu; Ari Hietanen; Henry der Sarkissian; Juha H Koivisto
Journal:  Dentomaxillofac Radiol       Date:  2022-07-13       Impact factor: 3.525

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.