Validation of minimally invasive, image-guided cochlear implantation using Advanced Bionics, Cochlear, and Medel electrodes in a cadaver model.

Theodore R McRackan1, Ramya Balachandran, Grégoire S Blachon, Jason E Mitchell, Jack H Noble, Charles G Wright, J Michael Fitzpatrick, Benoit M Dawant, Robert F Labadie.   

Abstract

PURPOSE: Validation of a novel minimally invasive, image-guided approach to implant electrodes from three FDA-approved manufacturers-Medel, Cochlear, and Advanced Bionics-in the cochlea via a linear tunnel from the lateral cranium through the facial recess to the cochlea.
METHODS: Custom microstereotactic frames that mount on bone-implanted fiducial markers and constrain the drill along the desired path were utilized on seven cadaver specimens. A linear tunnel was drilled from the lateral skull to the cochlea followed by a marginal, round window cochleostomy and insertion of the electrode array into the cochlea through the drilled tunnel. Post-insertion CT scan and histological analysis were used to analyze the results.
RESULTS: All specimens ([Formula: see text]) were successfully implanted without visible injury to the facial nerve. The Medel electrodes ([Formula: see text]) had minimal intracochlear trauma with 8, 8, and 10 (out of 12) electrodes intracochlear. The Cochlear lateral wall electrodes (straight research arrays) ([Formula: see text]) had minimal trauma with 20 and 21 of 22 electrodes intracochlear. The Advanced Bionics electrodes ([Formula: see text]) were inserted using their insertion tool; one had minimal insertion trauma and 14 of 16 electrodes intracochlear, while the other had violation of the basilar membrane just deep to the cochleostomy following which it remained in scala vestibuli with 13 of 16 electrodes intracochlear.
CONCLUSIONS: Minimally invasive, image-guided cochlear implantation is possible using electrodes from the three FDA-approved manufacturers. Lateral wall electrodes were associated with less intracochlear trauma suggesting that they may be better suited for this surgical technique.

Entities:  

Mesh:

Year:  2013        PMID: 23633113      PMCID: PMC4431760          DOI: 10.1007/s11548-013-0842-6

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  30 in total

1.  Speech recognition with altered spectral distribution of envelope cues.

Authors:  R V Shannon; F G Zeng; J Wygonski
Journal:  J Acoust Soc Am       Date:  1998-10       Impact factor: 1.840

2.  Optimizing the perception of soft speech and speech in noise with the Advanced Bionics cochlear implant system.

Authors:  Laura K Holden; Ruth M Reeder; Jill B Firszt; Charles C Finley
Journal:  Int J Audiol       Date:  2011-01-28       Impact factor: 2.117

3.  The suprameatal approach: an alternative surgical approach to cochlear implantation.

Authors:  Jona Kronenberg; Wolfgang Baumgartner; Lela Migirov; Tal Dagan; Minka Hildesheimer
Journal:  Otol Neurotol       Date:  2004-01       Impact factor: 2.311

4.  The suprameatal approach in cochlear implant surgery: our experience with 80 patients.

Authors:  Jona Kronenberg; Lela Migirov; Wolf Dieter Baumgartner
Journal:  ORL J Otorhinolaryngol Relat Spec       Date:  2002 Nov-Dec       Impact factor: 1.538

5.  Percutaneous cochlear implant drilling via customized frames: an in vitro study.

Authors:  Ramya Balachandran; Jason E Mitchell; Grégoire Blachon; Jack H Noble; Benoit M Dawant; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2010-03       Impact factor: 3.497

6.  Complication rate of minimally invasive cochlear implantation.

Authors:  Elias D Stratigouleas; Brian P Perry; Susan Marenda King; Charles A Syms
Journal:  Otolaryngol Head Neck Surg       Date:  2006-09       Impact factor: 3.497

7.  Percutaneous cochlear access using bone-mounted, customized drill guides: demonstration of concept in vitro.

Authors:  Frank M Warren; Ramya Balachandran; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otol Neurotol       Date:  2007-04       Impact factor: 2.311

8.  Clinical validation of percutaneous cochlear implant surgery: initial report.

Authors:  Robert Frederick Labadie; Jack H Noble; Benoit M Dawant; Ramya Balachandran; Omid Majdani; J Michael Fitzpatrick
Journal:  Laryngoscope       Date:  2008-06       Impact factor: 3.325

9.  Clinical validation study of percutaneous cochlear access using patient-customized microstereotactic frames.

Authors:  Robert F Labadie; Ramya Balachandran; Jason E Mitchell; Jack H Noble; Omid Majdani; David S Haynes; Marc L Bennett; Benoit M Dawant; J Michael Fitzpatrick
Journal:  Otol Neurotol       Date:  2010-01       Impact factor: 2.311

10.  A manually operated, advance off-stylet insertion tool for minimally invasive cochlear implantation surgery.

Authors:  Louis B Kratchman; Daniel Schurzig; Theodore R McRackan; Ramya Balachandran; Jack H Noble; Robert J Webster; Robert F Labadie
Journal:  IEEE Trans Biomed Eng       Date:  2012-07-25       Impact factor: 4.538

View more
  8 in total

1.  Configuration optimization and experimental accuracy evaluation of a bone-attached, parallel robot for skull surgery.

Authors:  Jan-Philipp Kobler; Kathrin Nuelle; G Jakob Lexow; Thomas S Rau; Omid Majdani; Lueder A Kahrs; Jens Kotlarski; Tobias Ortmaier
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-09-26       Impact factor: 2.924

2.  An experimental evaluation of loads occurring during guided drilling for cochlear implantation.

Authors:  Jan-Philipp Kobler; Sergej Wall; G Jakob Lexow; Carl Philipp Lang; Omid Majdani; Lüder A Kahrs; Tobias Ortmaier
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-02-12       Impact factor: 2.924

Review 3.  Outlook and future of inner ear therapy.

Authors:  Jenna Devare; Samuel Gubbels; Yehoash Raphael
Journal:  Hear Res       Date:  2018-05-17       Impact factor: 3.208

4.  Temporal bone borehole accuracy for cochlear implantation influenced by drilling strategy: an in vitro study.

Authors:  Jan-Philipp Kobler; Michael Schoppe; G Jakob Lexow; Thomas S Rau; Omid Majdani; Lüder A Kahrs; Tobias Ortmaier
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-04-12       Impact factor: 2.924

5.  An oral and maxillofacial navigation system for implant placement with automatic identification of fiducial points.

Authors:  Chunxia Qin; Zhenggang Cao; Shengchi Fan; Yiqun Wu; Yi Sun; Constantinus Politis; Chunliang Wang; Xiaojun Chen
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-10-13       Impact factor: 2.924

6.  Forces and trauma associated with minimally invasive image-guided cochlear implantation.

Authors:  Pooyan Rohani; Jason Pile; Lueder A Kahrs; Ramya Balachandran; Grégoire S Blachon; Nabil Simaan; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2014-01-27       Impact factor: 3.497

7.  Individual Optimization of the Insertion of a Preformed Cochlear Implant Electrode Array.

Authors:  Thomas S Rau; Thomas Lenarz; Omid Majdani
Journal:  Int J Otolaryngol       Date:  2015-09-10

8.  High-resolution Imaging of the Human Cochlea through the Round Window by means of Optical Coherence Tomography.

Authors:  Anastasiya Starovoyt; Tristan Putzeys; Jan Wouters; Nicolas Verhaert
Journal:  Sci Rep       Date:  2019-10-03       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.