Forces and trauma associated with minimally invasive image-guided cochlear implantation.

Pooyan Rohani1, Jason Pile, Lueder A Kahrs, Ramya Balachandran, Grégoire S Blachon, Nabil Simaan, Robert F Labadie.   

Abstract

OBJECTIVE: Minimally invasive image-guided cochlear implantation (CI) utilizes a patient-customized microstereotactic frame to access the cochlea via a single drill-pass. We investigate the average force and trauma associated with the insertion of lateral wall CI electrodes using this technique. STUDY
DESIGN: Assessment using cadaveric temporal bones.
SETTING: Laboratory setup. SUBJECTS AND METHODS: Microstereotactic frames for 6 fresh cadaveric temporal bones were built using CT scans to determine an optimal drill path following which drilling was performed. CI electrodes were inserted using surgical forceps to manually advance the CI electrode array, via the drilled tunnel, into the cochlea. Forces were recorded using a 6-axis load sensor placed under the temporal bone during the insertion of lateral wall electrode arrays (2 each of Nucleus CI422, MED-EL standard, and modified MED-EL electrodes with stiffeners). Tissue histology was performed by microdissection of the otic capsule and apical photo documentation of electrode position and intracochlear tissue.
RESULTS: After drilling, CT scanning demonstrated successful access to cochlea in all 6 bones. Average insertion forces ranged from 0.009 to 0.078 N. Peak forces were in the range of 0.056 to 0.469 N. Tissue histology showed complete scala tympani insertion in 5 specimens and scala vestibuli insertion in the remaining specimen with depth of insertion ranging from 360° to 600°. No intracochlear trauma was identified.
CONCLUSION: The use of lateral wall electrodes with the minimally invasive image-guided CI approach was associated with insertion forces comparable to traditional CI surgery. Deep insertions were obtained without identifiable trauma.

Entities:  

Keywords:  cochlear electrode; cochlear histology; cochlear implantation; insertion forces; minimally invasive surgery

Mesh:

Year:  2014        PMID: 24468898      PMCID: PMC4465091          DOI: 10.1177/0194599813519747

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  19 in total

1.  Suprameatal approach: new surgical approach for cochlear implantation.

Authors:  J Kronenberg; L Migirov; T Dagan
Journal:  J Laryngol Otol       Date:  2001-04       Impact factor: 1.469

2.  The suprameatal approach: an alternative surgical approach to cochlear implantation.

Authors:  Jona Kronenberg; Wolfgang Baumgartner; Lela Migirov; Tal Dagan; Minka Hildesheimer
Journal:  Otol Neurotol       Date:  2004-01       Impact factor: 2.311

3.  Friction force measurement during cochlear implant insertion: application to a force-controlled insertion tool design.

Authors:  Mathieu Miroir; Yann Nguyen; Guillaume Kazmitcheff; Evelyne Ferrary; Olivier Sterkers; Alexis Bozorg Grayeli
Journal:  Otol Neurotol       Date:  2012-08       Impact factor: 2.311

4.  Percutaneous cochlear implant drilling via customized frames: an in vitro study.

Authors:  Ramya Balachandran; Jason E Mitchell; Grégoire Blachon; Jack H Noble; Benoit M Dawant; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2010-03       Impact factor: 3.497

5.  Percutaneous cochlear access using bone-mounted, customized drill guides: demonstration of concept in vitro.

Authors:  Frank M Warren; Ramya Balachandran; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otol Neurotol       Date:  2007-04       Impact factor: 2.311

6.  Clinical validation of percutaneous cochlear implant surgery: initial report.

Authors:  Robert Frederick Labadie; Jack H Noble; Benoit M Dawant; Ramya Balachandran; Omid Majdani; J Michael Fitzpatrick
Journal:  Laryngoscope       Date:  2008-06       Impact factor: 3.325

7.  Evaluation of portable CT scanners for otologic image-guided surgery.

Authors:  Ramya Balachandran; Daniel Schurzig; J Michael Fitzpatrick; Robert F Labadie
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-07-21       Impact factor: 2.924

8.  Clinical validation study of percutaneous cochlear access using patient-customized microstereotactic frames.

Authors:  Robert F Labadie; Ramya Balachandran; Jason E Mitchell; Jack H Noble; Omid Majdani; David S Haynes; Marc L Bennett; Benoit M Dawant; J Michael Fitzpatrick
Journal:  Otol Neurotol       Date:  2010-01       Impact factor: 2.311

9.  Force measurement of insertion of cochlear implant electrode arrays in vitro: comparison of surgeon to automated insertion tool.

Authors:  Omid Majdani; Daniel Schurzig; Andreas Hussong; Thomas Rau; Justin Wittkopf; Thomas Lenarz; Robert F Labadie
Journal:  Acta Otolaryngol       Date:  2010       Impact factor: 1.494

10.  A manually operated, advance off-stylet insertion tool for minimally invasive cochlear implantation surgery.

Authors:  Louis B Kratchman; Daniel Schurzig; Theodore R McRackan; Ramya Balachandran; Jack H Noble; Robert J Webster; Robert F Labadie
Journal:  IEEE Trans Biomed Eng       Date:  2012-07-25       Impact factor: 4.538

View more
  3 in total

1.  Accuracy of linear drilling in temporal bone using drill press system for minimally invasive cochlear implantation.

Authors:  Neal P Dillon; Ramya Balachandran; Robert F Labadie
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-07-17       Impact factor: 2.924

2.  Insertion forces and intracochlear trauma in temporal bone specimens implanted with a straight atraumatic electrode array.

Authors:  Marjan Mirsalehi; Thomas S Rau; Lenka Harbach; Silke Hügl; Saleh Mohebbi; Thomas Lenarz; Omid Majdani
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-02-25       Impact factor: 2.503

3.  The impact of round window vs cochleostomy surgical approaches on interscalar excursions in the cochlea: Preliminary results from a flat-panel computed tomography study.

Authors:  Nicole T Jiam; Charles J Limb
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2016-11-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.