Literature DB >> 23528556

Pseudotumour formation due to tribocorrosion at the taper interface of large diameter metal on polymer modular total hip replacements.

Richard B Cook1, Benjamin J R F Bolland, Julian A Wharton, Simon Tilley, Jeremy M Latham, Robert J K Wood.   

Abstract

We present an in-depth failure analysis of two large diameter bearing metal-on-polymer (MoP) modular total hip replacements, which have required revision surgery due to pseudotumour formation. The failure analysis showed a discrete pattern of material loss from the distal end of the head taper/stem trunnion interface. We postulate that the use of a proximal contacting taper design had provided insufficient mechanical locking between the head and the stem, enabling the head to toggle on the trunnion. In addition, the difference in angle between the taper and the trunnion formed a crevice between the two components. Through a combination of crevice environment, mechanically assisted corrosion, mechanical wear and erosion; debris and metal-ions have been released resulting in the adverse local tissue reactions (ALTR).
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  corrosion; fretting; hip; pseudotumour; taper

Mesh:

Substances:

Year:  2013        PMID: 23528556     DOI: 10.1016/j.arth.2013.02.009

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  27 in total

1.  Development and first validation of a simplified CT-based classification system of soft tissue changes in large-head metal-on-metal total hip replacement: intra- and interrater reliability and association with revision rates in a uniform cohort of 664 arthroplasties.

Authors:  Martijn F Boomsma; Mireille A Edens; Christiaan P Van Lingen; Niek Warringa; Harmen B Ettema; Cees C P M Verheyen; Mario Maas
Journal:  Skeletal Radiol       Date:  2015-05-06       Impact factor: 2.199

Review 2.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

3.  Modern trunnions are more flexible: a mechanical analysis of THA taper designs.

Authors:  David A Porter; Robert M Urban; Joshua J Jacobs; Jeremy L Gilbert; José A Rodriguez; H John Cooper
Journal:  Clin Orthop Relat Res       Date:  2014-09-30       Impact factor: 4.176

4.  CORR Insights(®): Modular to Monoblock: Difficulties of Detaching the M(2)a-Magnum(™) Head Are Common in Metal-on-metal Revisions.

Authors:  Rodrigo M Mardones
Journal:  Clin Orthop Relat Res       Date:  2016-05-18       Impact factor: 4.176

5.  Mechanical, chemical and biological damage modes within head-neck tapers of CoCrMo and Ti6Al4V contemporary hip replacements.

Authors:  Deborah J Hall; Robin Pourzal; Hannah J Lundberg; Mathew T Mathew; Joshua J Jacobs; Robert M Urban
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2017-08-26       Impact factor: 3.368

Review 6.  Wear of dual-mobility cups: a review article.

Authors:  Anna Di Laura; Harry Hothi; Clement Battisti; Arianna Cerquiglini; Johann Henckel; John Skinner; Alister Hart
Journal:  Int Orthop       Date:  2016-12-09       Impact factor: 3.075

7.  Does Taper Angle Clearance Influence Fretting and Corrosion Damage at the Head-Stem Interface? A Matched Cohort Retrieval Study.

Authors:  Sevi B Kocagöz; Richard J Underwood; Shiril Sivan; Jeremy L Gilbert; Daniel W Macdonald; Judd S Day; Steven M Kurtz
Journal:  Semin Arthroplasty       Date:  2013-12-01

8.  Direct in vivo inflammatory cell-induced corrosion of CoCrMo alloy orthopedic implant surfaces.

Authors:  Jeremy L Gilbert; Shiril Sivan; Yangping Liu; Sevi B Kocagöz; Christina M Arnholt; Steven M Kurtz
Journal:  J Biomed Mater Res A       Date:  2014-03-26       Impact factor: 4.396

9.  Diagnosis and Management of Adverse Reactions to Metal Debris.

Authors:  Richard A Wawrose; Kenneth L Urish
Journal:  Oper Tech Orthop       Date:  2019-10-01

Review 10.  [Endoprostheses in the elderly : Biomaterials, implant selection and fixation technique].

Authors:  M M Morlock; M Jäger
Journal:  Orthopade       Date:  2017-01       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.