Anna Di Laura1,2, Harry Hothi3,4, Clement Battisti3,4, Arianna Cerquiglini3,4, Johann Henckel3,4, John Skinner3,4, Alister Hart3,4.
Abstract
Dual-mobility (DM) cups have been clinically used in hip surgery in Europe for more than 35 years and continue to gain popularity worldwide due to promising results at reducing instability. Concerns related to polyethylene wear apply as in conventional standard bearings but are accentuated by the larger-diameter articulations with multiple surfaces. We critically reviewed the reported literature regarding the in vivo and in vitro wear occurring on all surfaces involved. We looked for patterns to create a rational classification of sites of wear and to identify areas for future research. Wear was a significant problem for first-generation designs and appeared to be design related. Improved polyethylene, thinner and smoother trunnions, chamfered rims and eccentric configuration of insert and shell seem to enhance outcome performance; however, long-term clinical evidence and retrieval studies are needed to better understand the balance of benefit and risk when opting for DM bearings.Entities:
Keywords: Bearings; Dual mobility cups; Hip arthroplasty; Modularity; Wear
Mesh:
Year: 2016 PMID: 27942888 DOI: 10.1007/s00264-016-3326-9
Source DB: PubMed Journal: Int Orthop ISSN: 0341-2695 Impact factor: 3.075