Surgical knot tightening: how much pull is necessary?

Stefan Neuhofer1, Karl Wieser, Georg Lajtai, Daniel Müller, Christian Gerber, Dominik C Meyer.   

Abstract

PURPOSE: High-strength sutures allow tightening of a suture knot beyond the strength of the surgeon, possibly inflicting skin damage through the gloves. This study was undertaken to evaluate whether such effort is useful and how much tensioning on a surgical knot is necessary.
METHODS: Three different suture materials were tested: No. 2 Vicryl, FibreWire, and PDS. First, the force spontaneously applied on sutures during experimental knot tightening ("tying load") was measured in fifteen experienced surgeons. Second, with each suture material, surgical square knots were tied with increasing, standardized loads (range 0.5-50 N) using a custom-made apparatus. Thereby, knot seating after tying was evaluated, and by loading the knots to failure, evaluation for failure mode and failure load was performed.
RESULTS: FibreWire 5-throw square knots always failed by complete slipping of all knots (resolving), independent on the tying load. A nonlinear decrease of knot slippage and increased failure load were seen with increasing tying loads for all sutures. Major differences were seen between 0.5 and 10 N for FibreWire (slippage: 25 mm) and PDS (99.6 mm), whereas Vicryl showed major differences (22.7 mm) between 0.5 and 2 N. Increasing the tying load from 10 to 50 N decreased the mean knot slippage from 12 (FibreWire, ±2.6 SD), 9 (PDS, ±1.8 SD) and 8 (Vicryl, ±1.3 SD) mm to 6 (±2.9 SD), 3 (±1.5 SD) and 4 mm (±0.9 SD), respectively.
CONCLUSION: Slippage and self-seating of the knots under load is unavoidable even with highest tying loads. Relatively minor but possibly important differences can be seen for tying loads exceeding 2 N (Vicryl) and 10 N (PDS and FibreWire) for failure load and knot slippage. But also with a tying load of 50 N, a minimal slippage of approximately 3 mm seems unavoidable for all suture types. However, it is important to state that intense tightening does not prevent knot resolution and is only necessary in clinical situations that demand very tight sutures. Numbers and proper appliance of throws are more relevant than tying strength to reach the maximum failure load.

Entities:  

Mesh:

Year:  2013        PMID: 23494026     DOI: 10.1007/s00167-013-2452-9

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  9 in total

Review 1.  Arthroscopic knot typing techniques.

Authors:  W M Nottage; R K Lieurance
Journal:  Arthroscopy       Date:  1999 Jul-Aug       Impact factor: 4.772

2.  Failure characteristics of various arthroscopically tied knots.

Authors:  Robert K Lieurance; Daniel S Pflaster; Douglas Abbott; Wesley M Nottage
Journal:  Clin Orthop Relat Res       Date:  2003-03       Impact factor: 4.176

3.  Arthroscopic knots: determining the optimal balance of loop security and knot security.

Authors:  Ian K Y Lo; Stephen S Burkhart; K Casey Chan; Kyriacos Athanasiou
Journal:  Arthroscopy       Date:  2004-05       Impact factor: 4.772

4.  The effect of number of throws on knot security with nonidentical sliding knots.

Authors:  Joseph J Ivy; James B Unger; Jason Hurt; Debi Mukherjee
Journal:  Am J Obstet Gynecol       Date:  2004-11       Impact factor: 8.661

5.  Security of knots tied with ethibond, fiberwire, orthocord, or ultrabraid.

Authors:  Omer A Ilahi; Shiraz A Younas; David M Ho; Phillip C Noble
Journal:  Am J Sports Med       Date:  2008-09-30       Impact factor: 6.202

6.  The effect of gap formation at the repair site on the strength and excursion of intrasynovial flexor tendons. An experimental study on the early stages of tendon-healing in dogs.

Authors:  R H Gelberman; M I Boyer; M D Brodt; S C Winters; M J Silva
Journal:  J Bone Joint Surg Am       Date:  1999-07       Impact factor: 5.284

Review 7.  Technical and biological modifications for enhanced flexor tendon repair.

Authors:  H Mike Kim; Gregory Nelson; Stavros Thomopoulos; Matthew J Silva; Rosalina Das; Richard H Gelberman
Journal:  J Hand Surg Am       Date:  2010-06       Impact factor: 2.230

8.  Mechanical and handling properties of braided polyblend polyethylene sutures in comparison to braided polyester and monofilament polydioxanone sutures.

Authors:  Daniel M Wüst; Dominik C Meyer; Philippe Favre; Christian Gerber
Journal:  Arthroscopy       Date:  2006-11       Impact factor: 4.772

9.  An evaluation of knot integrity when tied robotically and conventionally.

Authors:  Tyler Muffly; T Chad McCormick; Julianne Dean; Aaron Bonham; Richard F C Hill
Journal:  Am J Obstet Gynecol       Date:  2008-12-27       Impact factor: 8.661

  9 in total
  4 in total

1.  The Sliding, Interlocking "8-2-Alpha Knot" for Arthroscopic Surgery.

Authors:  Wiroon Laupattarakasem; Pat Laupattarakasem
Journal:  Arthrosc Tech       Date:  2016-08-08

2.  Suture Breakage Warning System for Robotic Surgery.

Authors:  Ahmad Abiri; Syed J Askari; Anna Tao; Yen-Yi Juo; Yuan Dai; Jake Pensa; Robert Candler; Erik P Dutson; Warren S Grundfest
Journal:  IEEE Trans Biomed Eng       Date:  2018-09-10       Impact factor: 4.538

3.  Bucket handle meniscus tears in low-resource settings can be successfully treated with a cost-effective technique.

Authors:  Allicia Ostoposides Imada; James J O'Hara; Ignacio L Proumen; Pablo S Molinari; Daniel C Wascher; Dustin L Richter; Robert C Schenck
Journal:  Int Orthop       Date:  2021-05-29       Impact factor: 3.075

4.  Tendon Collagen Crosslinking Offers Potential to Improve Suture Pullout in Rotator Cuff Repair: An Ex Vivo Sheep Study.

Authors:  Roland S Camenzind; Karl Wieser; Gion Fessel; Dominik C Meyer; Jess G Snedeker
Journal:  Clin Orthop Relat Res       Date:  2016-04-18       Impact factor: 4.176

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.