Literature DB >> 23453091

Determining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: rationale and design of the MagnaSafe Registry.

Robert J Russo1.   

Abstract

BACKGROUND: Until recently, the presence of a permanent pacemaker or an implantable cardioverter-defibrillator has been a relative contraindication for the performance of magnetic resonance imaging (MRI). A number of small studies have shown that MRI can be performed with minimal risk when patients are properly monitored and device programming is modified appropriately for the procedure. However, the risk of performing MRI for patients with implanted cardiac devices has not been sufficiently evaluated to advocate routine clinical use. The aim of the present protocol is to prospectively determine the rate of adverse clinical events and device parameter changes in patients with implanted non-MRI-conditional cardiac devices undergoing clinically indicated nonthoracic MRI at 1.5 T.
METHODS: The MagnaSafe Registry is a multicenter, prospective cohort study of up to 1500 MRI examinations in patients with pacemakers or implantable cardioverter-defibrillators implanted after 2001 who undergo clinically indicated nonthoracic MRI following a specific protocol to ensure that preventable potential adverse events are mitigated. Adverse events and changes in device parameter measurements that may be associated with the imaging procedure will be documented.
RESULTS: Through August 2012, 701 MRI studies have been performed, representing 47% of the total target enrollment.
CONCLUSIONS: The results of this registry will provide additional documentation of the risk of MRI and will further validate a clinical protocol for screening and the performance of clinically indicated MRI for patients with implanted cardiac devices.
Copyright © 2013 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23453091     DOI: 10.1016/j.ahj.2012.12.004

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  14 in total

Review 1.  Safety of implanted cardiac devices in an MRI environment.

Authors:  Esra Gucuk Ipek; Saman Nazarian
Journal:  Curr Cardiol Rep       Date:  2015-07       Impact factor: 2.931

2.  Feasibility of Cardiac Magnetic Resonance Wideband Protocol in Patients With Implantable Cardioverter Defibrillators and Its Utility for Defining Scar.

Authors:  Amita Singh; Keigo Kawaji; Neha Goyal; Noreen T Nazir; Andrew Beaser; Virginia O'Keefe-Baker; Karima Addetia; Roderick Tung; Peng Hu; Victor Mor-Avi; Amit R Patel
Journal:  Am J Cardiol       Date:  2019-01-31       Impact factor: 2.778

Review 3.  MRI for patients with cardiac implantable electrical devices.

Authors:  Grant V Chow; Saman Nazarian
Journal:  Cardiol Clin       Date:  2014-02-12       Impact factor: 2.213

4.  MRI of Patients with Cardiac Implantable Electronic Devices.

Authors:  Jessica A Martinez; Daniel B Ennis
Journal:  Curr Cardiovasc Imaging Rep       Date:  2019-05-27

5.  Feasibility of MRI in patients with non-Pacemaker/Defibrillator metallic devices and abandoned leads.

Authors:  Prabhakaran P Gopalakrishnan; Loretta Gevenosky; Robert W W Biederman
Journal:  J Biomed Sci Eng       Date:  2021-03-09

6.  Magnetic resonance imaging in patients with a subcutaneous implantable cardioverter-defibrillator.

Authors:  Jirí Keller; Petr Neužil; Josef Vymazal; Marek Janotka; Jirí Brada; Radovan Žáček; Roman Vopálka; Jirí Weichet; Vivek Y Reddy
Journal:  Europace       Date:  2015-02-16       Impact factor: 5.214

Review 7.  MRI-conditional pacemakers: current perspectives.

Authors:  António M Ferreira; Francisco Costa; António Tralhão; Hugo Marques; Nuno Cardim; Pedro Adragão
Journal:  Med Devices (Auckl)       Date:  2014-05-07

8.  Safe performance of magnetic resonance of the heart in patients with magnetic resonance conditional pacemaker systems: the safety issue of the ESTIMATE study.

Authors:  Christian G Wollmann; Karin Thudt; Bernd Kaiser; Erich Salomonowitz; Harald Mayr; Sebastian Globits
Journal:  J Cardiovasc Magn Reson       Date:  2014-05-06       Impact factor: 5.364

9.  Issues with radiofrequency heating in MRI.

Authors:  Jihong Wang
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

Review 10.  Cardiovascular implantable electronic devices: a review of the dangers and difficulties in MR scanning and attempts to improve safety.

Authors:  Pei Ghim Poh; Charlene Liew; Colin Yeo; Le Roy Chong; Andrew Tan; Angeline Poh