Nejc Mekis1, Dejan Zontar, Damijan Skrk.
Abstract
BACKGROUND: The aim of the study was to determine the influence of lead shielding on the dose to female breasts in conventional x-ray lumbar spine imaging. The correlation between the body mass index and the dose received by the breast was also investigated.Entities:
Keywords: breast dose; lead shielding; lumbar spine radiography; scattered radiation
Year: 2013 PMID: 23450158 PMCID: PMC3573831 DOI: 10.2478/raon-2013-0004
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
FIGURE 1.Image of the phantom with 340 ml implant size on the right and 500 ml implant size on the left.
FIGURE 2.The position of the TLDs on the implants.
Dose reduction on the phantom study with 340 ml implant size
| 0.07 (− 73%) | 0.02 (− 61%) | ||||
| 0.16 (− 65%) | 0.14 (− 84%) | 0.16 (− 73%) | 0.14 (− 71%) | 0.10 (− 78%) | 0.04 (− 59%) |
| 0.56 (− 67%) | 0.81 (− 95%) |
Dose reduction on the phantom study with 500 ml implant size
| 0.11 (− 96%) | 0.03 (− 86%) | ||||
| 0.28 (− 86%) | 0.16 (− 88%) | 0.11 (− 64%) | 0.32 (− 80%) | 0.06 (− 80%) | 0.06 (− 60%) |
| 0.65 (− 70%) | 0.024 (− 65%) |
Basic statistical characteristics of right breast dose measurement with and without the lead shielding
| Right breast without the lead shield | 0.45 | 0.39 | 0.25 | 0.11 | 1.23 |
| Right breast with the lead shield | 0.09 | 0.08 | 0.07 | 0.01 | 0.36 |
Basic statistical characteristics of left breast dose measurement with and without the lead shielding
| Left breast without the lead shield | 0.26 | 0.24 | 0.14 | 0.07 | 0.70 |
| Left breast with the lead shield | 0.06 | 0.05 | 0.04 | 0.01 | 0.17 |
FIGURE 3.The comparison of the dose between shielded and unshielded right breast.
FIGURE 4.The comparison of the dose between shielded and unshielded left breast.