Accuracy of continuous glucose monitoring during exercise in type 1 diabetes pregnancy.

Kavita Kumareswaran1, Daniela Elleri, Janet M Allen, Karen Caldwell, Marianna Nodale, Malgorzata E Wilinska, Stephanie A Amiel, Roman Hovorka, Helen R Murphy.   

Abstract

BACKGROUND: Performance of continuous glucose monitors (CGMs) may be lower when glucose levels are changing rapidly, such as occurs during physical activity. Our aim was to evaluate accuracy of a current-generation CGM during moderate-intensity exercise in type 1 diabetes (T1D) pregnancy. SUBJECTS AND METHODS: As part of a study of 24-h closed-loop insulin delivery in 12 women with T1D (disease duration, 17.6 years; glycosylated hemoglobin, 6.4%) during pregnancy (gestation, 21 weeks), we evaluated the Freestyle Navigator(®) sensor (Abbott Diabetes Care, Alameda, CA) during afternoon (15:00-18:00 h) and morning (09:30-12:30 h) exercise (55 min of brisk walking on a treadmill followed by a 2-h recovery), compared with sedentary conditions (18:00-09:00 h). Plasma (reference) glucose, measured at regular 15-30-min intervals with the YSI Ltd. (Fleet, United Kingdom) model YSI 2300 analyzer, was used to assess CGM performance.
RESULTS: Sensor accuracy, as indicated by the larger relative absolute difference (RAD) between paired sensor and reference glucose values, was lower during exercise compared with rest (median RAD, 11.8% vs. 18.4%; P<0.001). These differences remained significant when correcting for plasma glucose relative rate of change (P<0.001). Analysis by glucose range showed lower accuracy during hypoglycemia for both sedentary (median RAD, 24.4%) and exercise (median RAD, 32.1%) conditions. Using Clarke error grid analysis, 96% of CGM values were clinically safe under resting conditions compared with only 87% during exercise.
CONCLUSIONS: Compared with sedentary conditions, accuracy of the Freestyle Navigator CGM was lower during moderate-intensity exercise in pregnant women with T1D. This difference was particularly marked in hypoglycemia and could not be solely explained by the glucose rate of change associated with physical activity.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23445170      PMCID: PMC3598434          DOI: 10.1089/dia.2012.0292

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  32 in total

1.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Physical activity/exercise and diabetes mellitus.

Authors:  Bernard Zinman; Neil Ruderman; Barbara N Campaigne; John T Devlin; Stephen H Schneider
Journal:  Diabetes Care       Date:  2003-01       Impact factor: 19.112

3.  Use of the noninvasive GlucoWatch Biographer during exercise of varying intensity.

Authors:  Tanja Nunnold; Sheri R Colberg; Matthew T Herriott; C Thomas Somma
Journal:  Diabetes Technol Ther       Date:  2004-08       Impact factor: 6.118

4.  Evaluation of factors affecting CGMS calibration.

Authors:  Bruce A Buckingham; Craig Kollman; Roy Beck; Andrea Kalajian; Rosanna Fiallo-Scharer; Michael J Tansey; Larry A Fox; Darrell M Wilson; Stuart A Weinzimer; Katrina J Ruedy; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

5.  Graphical and numerical evaluation of continuous glucose sensing time lag.

Authors:  Boris P Kovatchev; Devin Shields; Marc Breton
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

6.  Point accuracy of interstitial continuous glucose monitoring during exercise in type 1 diabetes.

Authors:  Jane E Yardley; Ronald J Sigal; Glen P Kenny; Michael C Riddell; Leif E Lovblom; Bruce A Perkins
Journal:  Diabetes Technol Ther       Date:  2012-11-08       Impact factor: 6.118

7.  Continuous glucose monitoring system during physical exercise in adolescents with type 1 diabetes.

Authors:  P Adolfsson; S Nilsson; B Lindblad
Journal:  Acta Paediatr       Date:  2011-10-10       Impact factor: 2.299

8.  Endurance athletes and type 1 diabetes.

Authors:  Mercy Devadoss; Laura Kennedy; Nancie Herbold
Journal:  Diabetes Educ       Date:  2011-02-15       Impact factor: 2.140

9.  FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.

Authors:  McGarraugh Geoffrey; Ronald Brazg; Weinstein Richard
Journal:  J Diabetes Sci Technol       Date:  2011-01-01

10.  Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems.

Authors:  Satish K Garg; James Smith; Christie Beatson; Benita Lopez-Baca; Mary Voelmle; Peter A Gottlieb
Journal:  Diabetes Technol Ther       Date:  2009-02       Impact factor: 6.118

View more
  10 in total

Review 1.  Different insulin types and regimens for pregnant women with pre-existing diabetes.

Authors:  Sinéad M O'Neill; Louise C Kenny; Ali S Khashan; Helen M West; Rebecca Md Smyth; Patricia M Kearney
Journal:  Cochrane Database Syst Rev       Date:  2017-02-03

Review 2.  CGM, Pregnancy, and Remote Monitoring.

Authors:  Sarit Polsky; Rachel Garcetti
Journal:  Diabetes Technol Ther       Date:  2017-06       Impact factor: 6.118

Review 3.  Quantifying the acute changes in glucose with exercise in type 1 diabetes: a systematic review and meta-analysis.

Authors:  Fernando García-García; Kavita Kumareswaran; Roman Hovorka; M Elena Hernando
Journal:  Sports Med       Date:  2015-04       Impact factor: 11.136

4.  Predicting Plasma Glucose From Interstitial Glucose Observations Using Bayesian Methods.

Authors:  Alexander Hildenbrand Hansen; Anne Katrine Duun-Henriksen; Rune Juhl; Signe Schmidt; Kirsten Nørgaard; John Bagterp Jørgensen; Henrik Madsen
Journal:  J Diabetes Sci Technol       Date:  2014-03-06

5.  A comparison of adipose tissue interstitial glucose and venous blood glucose during postprandial resistance exercise in patients with type 2 diabetes.

Authors:  Timothy D Heden; Ying Liu; Jill A Kanaley
Journal:  J Appl Physiol (1985)       Date:  2017-09-28

6.  Adding heart rate signal to a control-to-range artificial pancreas system improves the protection against hypoglycemia during exercise in type 1 diabetes.

Authors:  Marc D Breton; Sue A Brown; Colleen Hughes Karvetski; Laura Kollar; Katarina A Topchyan; Stacey M Anderson; Boris P Kovatchev
Journal:  Diabetes Technol Ther       Date:  2014-04-04       Impact factor: 6.118

7.  Closed-loop artificial pancreas systems: physiological input to enhance next-generation devices.

Authors:  Yogish C Kudva; Rickey E Carter; Claudio Cobelli; Rita Basu; Ananda Basu
Journal:  Diabetes Care       Date:  2014       Impact factor: 19.112

8.  Accuracy of Continuous Glucose Monitoring before, during, and after Aerobic and Anaerobic Exercise in Patients with Type 1 Diabetes Mellitus.

Authors:  Lyvia Biagi; Arthur Bertachi; Carmen Quirós; Marga Giménez; Ignacio Conget; Jorge Bondia; Josep Vehí
Journal:  Biosensors (Basel)       Date:  2018-03-09

9.  Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes.

Authors:  Eleanor M Scott; Rudy W Bilous; Alexandra Kautzky-Willer
Journal:  Diabetes Technol Ther       Date:  2018-02-22       Impact factor: 6.118

10.  Accuracy of the Dexcom G6 Glucose Sensor during Aerobic, Resistance, and Interval Exercise in Adults with Type 1 Diabetes.

Authors:  Florian H Guillot; Peter G Jacobs; Leah M Wilson; Joseph El Youssef; Virginia B Gabo; Deborah L Branigan; Nichole S Tyler; Katrina Ramsey; Michael C Riddell; Jessica R Castle
Journal:  Biosensors (Basel)       Date:  2020-09-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.