Alleviating artifacts in 1H MRI thermometry by single scan spatiotemporal encoding.

Rita Schmidt1, Lucio Frydman.   

Abstract

OBJECTIVE: Recent years have seen an increased interest in combining MRI thermometry with devices capable of destroying malignancies by heat ablation. Expected from the MR protocols are accurate and fast thermal characterizations, providing real time feedback on restricted tissue volumes and/or rapidly moving organs like liver. This article explores the potential advantages of relying on spatiotemporally encoded (SPEN) sequences for retrieving real-time thermometric images based on the water's proton resonance frequency (PRF) shifts.
MATERIALS AND METHODS: Hybrid spatiotemporal/k-space encoding single-scan MRI experiments were implemented on animal and human scanners, and their abilities to deliver single- and multi-slice real-time thermometric measurements based on PRF-derived phase maps in phantoms and in vivo, were compared against echo planar imaging (EPI) and gradient-echo counterparts.
RESULTS: Under comparable acquisition conditions, SPEN exhibited advantages vis-à-vis EPI in terms of dealing with inhomogeneous magnetic field distortions, with shifts arising due to changes in the central frequency offsets, with PRF distributions, and for zooming into restricted fields-of-view without special pulse sequence provisions.
CONCLUSION: This work confirms the ability of SPEN sequences, particularly when implemented under fully-refocused conditions, to exploit their built-in robustness to shift- and field-derived inhomogeneities for monitoring thermal changes in real-time under in vitro and in vivo conditions.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23443884     DOI: 10.1007/s10334-013-0372-9

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  29 in total

Review 1.  Magnetic resonance temperature imaging for guidance of thermotherapy.

Authors:  B Quesson; J A de Zwart; C T Moonen
Journal:  J Magn Reson Imaging       Date:  2000-10       Impact factor: 4.813

2.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

3.  Single-scan multidimensional magnetic resonance.

Authors:  Assaf Tal; Lucio Frydman
Journal:  Prog Nucl Magn Reson Spectrosc       Date:  2010-04-22       Impact factor: 9.795

4.  Spatial encoding and the single-scan acquisition of high definition MR images in inhomogeneous fields.

Authors:  Assaf Tal; Lucio Frydman
Journal:  J Magn Reson       Date:  2006-07-14       Impact factor: 2.229

5.  Readout-segmented EPI for rapid high resolution diffusion imaging at 3 T.

Authors:  Samantha J Holdsworth; Stefan Skare; Rexford D Newbould; Raphael Guzmann; Nikolas H Blevins; Roland Bammer
Journal:  Eur J Radiol       Date:  2007-11-05       Impact factor: 3.528

6.  RASER: a new ultrafast magnetic resonance imaging method.

Authors:  Ryan Chamberlain; Jang-Yeon Park; Curt Corum; Essa Yacoub; Kamil Ugurbil; Clifford R Jack; Michael Garwood
Journal:  Magn Reson Med       Date:  2007-10       Impact factor: 4.668

Review 7.  MR thermometry.

Authors:  Viola Rieke; Kim Butts Pauly
Journal:  J Magn Reson Imaging       Date:  2008-02       Impact factor: 4.813

8.  Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences.

Authors:  A Kickhefel; Jörg Roland; Clifford Weiss; Fritz Schick
Journal:  Phys Med       Date:  2010-01-21       Impact factor: 2.685

9.  Noninvasive magnetic resonance thermography of recurrent rectal carcinoma in a 1.5 Tesla hybrid system.

Authors:  Johanna Gellermann; Waldemar Wlodarczyk; Bert Hildebrandt; Hildegard Ganter; Anett Nicolau; Beate Rau; Wolfgang Tilly; Horst Fähling; Jacek Nadobny; Roland Felix; Peter Wust
Journal:  Cancer Res       Date:  2005-07-01       Impact factor: 12.701

10.  Use of frequency-modulated radiofrequency pulses in MR imaging experiments.

Authors:  D Kunz
Journal:  Magn Reson Med       Date:  1986-06       Impact factor: 4.668

View more
  2 in total

1.  Improved MRI thermometry with multiple-echo spirals.

Authors:  Michael Marx; Kim Butts Pauly
Journal:  Magn Reson Med       Date:  2015-08-29       Impact factor: 4.668

2.  Robust diffusion tensor imaging by spatiotemporal encoding: Principles and in vivo demonstrations.

Authors:  Eddy Solomon; Gilad Liberman; Noam Nissan; Lucio Frydman
Journal:  Magn Reson Med       Date:  2016-03-10       Impact factor: 4.668

  2 in total