Spectral optimization of chest CT angiography with reduced iodine load: experience in 80 patients evaluated with dual-source, dual-energy CT.

Marie-Aurélie Delesalle1, François Pontana, Alain Duhamel, Jean-Baptiste Faivre, Thomas Flohr, Nunzia Tacelli, Jacques Remy, Martine Remy-Jardin.   

Abstract

PURPOSE: To determine the energy levels that provide optimal imaging of thoracic circulation at dual-energy computed tomographic (CT) angiography with reduced iodine load in comparison with a standard technique.
MATERIALS AND METHODS: The institutional review board approved the study with waiver of patient consent. Eighty patients underwent a dual-source, dual-energy CT examination after administration of low-concentration contrast material (170 mg of iodine per milliliter), and eight series of images were reconstructed, including the original polychromatic images at 80 and 140 kV and six series of virtual monochromatic spectral images at 50, 60, 70, 80, 90, and 100 keV. For each vascular compartment, the energy level that provided optimal evaluation on virtual monochromatic spectral images was determined, and these series were compared with the polychromatic dual-energy images and with standard chest CT images that were used as controls. Comparisons between groups were performed by using the paired Student t test for continuous variables and the McNemar test for categorical variables. Comparisons between dual-energy and standard CT images were performed by using the unpaired Student t test for continuous variables and the χ(2) test for categorical variables.
RESULTS: For the aorta, pulmonary arteries, and veins, the reconstruction at 60 keV provided adequate attenuation without marked beam-hardening artifacts in 90% of patients, with the highest contrast-to-noise and signal-to-noise ratios, the lowest level of subjective noise, and no significant differences with images at 80 kV (mean energy, 54 keV). For the superior vena cava and brachiocephalic veins, the reconstructions at 100 keV enabled artifact-free analysis of the perivascular anatomic zone without a significant difference with images at 140 kV (mean energy, 92 keV). Compared with standard CT images acquired after administration of a 35% iodinated contrast agent, there was a statistically significant reduction in the frequency of artifacts around systemic veins at 100 keV (P < .001) and similar overall image quality for central vessels at 60 keV (P > .05).
CONCLUSION: An optimal analysis of thoracic circulation can be achieved on virtual monochromatic spectral images at 60 keV and 100 keV and on the original polychromatic images at 80 kV and 140 kV. SUPPLEMENTAL MATERIAL: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.12120195/-/DC1. RSNA, 2013

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23319663     DOI: 10.1148/radiol.12120195

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  41 in total

1.  Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging.

Authors:  Martin Beeres; Jesko Trommer; Claudia Frellesen; Nour-Eldin A Nour-Eldin; Jan E Scholtz; Eva Herrmann; Thomas J Vogl; Julian L Wichmann
Journal:  Int J Cardiovasc Imaging       Date:  2015-08-02       Impact factor: 2.357

2.  Individually tailored contrast enhancement in CT pulmonary angiography.

Authors:  Babs M F Hendriks; Madeleine Kok; Casper Mihl; Sebastiaan C A M Bekkers; Joachim E Wildberger; Marco Das
Journal:  Br J Radiol       Date:  2016-01-22       Impact factor: 3.039

3.  Assessment of an advanced virtual monoenergetic reconstruction technique in cerebral and cervical angiography with third-generation dual-source CT: Feasibility of using low-concentration contrast medium.

Authors:  Lu Zhao; Fengtan Li; Zewei Zhang; Zhang Zhang; Yingjian Jiang; Xinyu Wang; Jun Gu; Dong Li
Journal:  Eur Radiol       Date:  2018-04-13       Impact factor: 5.315

Review 4.  Dual energy computed tomography virtual monoenergetic imaging: technique and clinical applications.

Authors:  Tommaso D'Angelo; Giuseppe Cicero; Silvio Mazziotti; Giorgio Ascenti; Moritz H Albrecht; Simon S Martin; Ahmed E Othman; Thomas J Vogl; Julian L Wichmann
Journal:  Br J Radiol       Date:  2019-04-09       Impact factor: 3.039

5.  Ultra-low dose contrast CT pulmonary angiography in oncology patients using a high-pitch helical dual-source technology.

Authors:  Prabhakar Rajiah; Leslie Ciancibello; Ronald Novak; Jennifer Sposato; Luis Landeras; Robert Gilkeson
Journal:  Diagn Interv Radiol       Date:  2019-05       Impact factor: 2.630

Review 6.  State of the art: utility of multi-energy CT in the evaluation of pulmonary vasculature.

Authors:  Prabhakar Rajiah; Yuki Tanabe; Sasan Partovi; Alastair Moore
Journal:  Int J Cardiovasc Imaging       Date:  2019-05-02       Impact factor: 2.357

7.  Substantial iodine volume load reduction in CT angiography with dual-energy imaging: insights from a pilot randomized study.

Authors:  Patricia Carrascosa; Carlos Capunay; Gaston A Rodriguez-Granillo; Alejandro Deviggiano; Javier Vallejos; Jonathon A Leipsic
Journal:  Int J Cardiovasc Imaging       Date:  2014-07-23       Impact factor: 2.357

8.  Assessment of an Advanced Monoenergetic Reconstruction Technique in Dual-Energy Computed Tomography of Head and Neck Cancer.

Authors:  Moritz H Albrecht; Jan-Erik Scholtz; Johannes Kraft; Ralf W Bauer; Moritz Kaup; Patricia Dewes; Andreas M Bucher; Iris Burck; Jens Wagenblast; Thomas Lehnert; J Matthias Kerl; Thomas J Vogl; Julian L Wichmann
Journal:  Eur Radiol       Date:  2015-02-14       Impact factor: 5.315

9.  Contrast agent concentration optimization in CTA using low tube voltage and dual-energy CT in multiple vendors: a phantom study.

Authors:  Robbert W van Hamersvelt; Nienke G Eijsvoogel; Casper Mihl; Pim A de Jong; Arnold M R Schilham; Nico Buls; Marco Das; Tim Leiner; Martin J Willemink
Journal:  Int J Cardiovasc Imaging       Date:  2018-03-08       Impact factor: 2.357

10.  Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose.

Authors:  Benjamin Dubourg; Jérôme Caudron; Jean-Pierre Lestrat; Michael Bubenheim; Valentin Lefebvre; Matthieu Godin; Christophe Tron; Hélène Eltchaninoff; Fabrice Bauer; Jean-Nicolas Dacher
Journal:  Eur Radiol       Date:  2014-06-25       Impact factor: 5.315