[Postoperative assessment of daily energy expenditure. Comparison of two methods].

R Dummler1, A Zittermann, M Schäfer, M Emmerich.   

Abstract

BACKGROUND: The reference method for determining resting energy expenditure (REE) in clinical nutrition practice is measurement by indirect calorimetry; however, indirect calorimetry has some limitations, is expensive and not widely available. Therefore, the most used methods to estimate the caloric requirements in intensive care patients are predictive equations. The Harris-Benedict equations (HBE) are the most common formulae in the clinical setting. The SenseWear(®) armlet (SWA) is a noninvasive device that monitors skin temperature, heat flux, galvanic skin response and movement. These data as well as anthropometric characteristics are used to calculate REE. The aim of this study was to evaluate the levels of agreement and interchangeability of REE estimated by HBE (EEHBE) and measured by SWA (EESWA) in normometabolic patients after elective bowel resection with laparotomy. Furthermore, postsurgical pain therapy by continuous thoracic epidural anaesthesia (t-PDA) was compared with continuous intravenous pain therapy regarding EESWA in these patients.
METHODS: After obtaining approval by the ethics committee and written informed consent 57 patients participated in the study procedures. A total of 50 patients (23 male, 27 female) were finally included in the data analysis because 7 patients did not meet the criterion of > 80% on-body time of the SWA. Additional (a priori) exclusion criteria were metabolic or cardiopulmonary decompensation or postoperative mechanical ventilation. Before induction of general anesthesia 26 patients received a thoracic epidural catheter. Immediately after surgery the SWA was placed on the right upper arm of each patient for 24 h. A continuous pain therapy was started either an epidural application of ropivacain 0.2% and sufentanil or in the other 24 patients an intravenous infusion of metamizol and tramadol. RESULTS AND DISCUSSION: The data showed good agreement between EESWA and EEHBE. The mean on-body time was found to be 22.94±4.77 h. There were no significant differences between EESWA and EEHBE (p>0.05) corresponding to a high Pearson's coefficient of correlation of 0.985. The mean bias (EESWA-EEHBE) was -0.569±0.378 kcal/kgBW/24 h reflecting a minimal systematic underestimation of REE by SWA of -2.9% compared to EEHBE. The Bland-Altman plot shows interchangeability of EESWA and EEHBE. It was noted that 94% of the data points (47 out of 50 patients) were within ±2 SD and the remaining 3 data points were lying close to the 95% interval. The same results (no significant differences between EESWA and EEHBE) were obtained after differentiation of EEHBE into low (<18 kcal/kgBW/24 h, n=9), medium (18-21 kcal/kgBW/24 h, n=30) and high (>21 kcal/kgBW/24 h, n=11) energy ranges. There were no significant differences in EESWA regarding postsurgical pain therapy regimens.
CONCLUSIONS: The SWA showed reliable concordance with daily REE estimated by HBE in normometabolic postsurgery patients. This noninvasive, convenient and easy to handle device may be helpful in determining energy requirements as part of metabolic monitoring. Further research is needed to validate the method in patients with severe metabolic disturbances. The energetic requirements of patients with postoperative t-PDA were not different from those with intravenous pain therapy.

Entities:  

Mesh:

Year:  2013        PMID: 23319271     DOI: 10.1007/s00101-012-2120-3

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  17 in total

Review 1.  Evidence basis for regional anesthesia in multidisciplinary fast-track surgical care pathways.

Authors:  Francesco Carli; Henrik Kehlet; Gabriele Baldini; Andrew Steel; Karen McRae; Peter Slinger; Thomas Hemmerling; Francis Salinas; Joseph M Neal
Journal:  Reg Anesth Pain Med       Date:  2011 Jan-Feb       Impact factor: 6.288

Review 2.  Indirect calorimetry: methodology, instruments and clinical application.

Authors:  Eduardo E Moreira da Rocha; Valéria Girard F Alves; Rosana Barcellos V da Fonseca
Journal:  Curr Opin Clin Nutr Metab Care       Date:  2006-05       Impact factor: 4.294

3.  Non-analgetic effects of thoracic epidural anaesthesia.

Authors:  Alf Kozian; Thomas Schilling; Thomas Hachenberg
Journal:  Curr Opin Anaesthesiol       Date:  2005-02       Impact factor: 2.706

4.  Evaluation of a portable device to measure daily energy expenditure in free-living adults.

Authors:  Maxime St-Onge; Diane Mignault; David B Allison; Rémi Rabasa-Lhoret
Journal:  Am J Clin Nutr       Date:  2007-03       Impact factor: 7.045

5.  A new device for measuring resting energy expenditure (REE) in healthy subjects.

Authors:  Marcella Malavolti; Angelo Pietrobelli; Manfredo Dugoni; Marco Poli; Elisa Romagnoli; Paolo De Cristofaro; Nino C Battistini
Journal:  Nutr Metab Cardiovasc Dis       Date:  2006-03-20       Impact factor: 4.222

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Graphical interpretation of analytical data from comparison of a field method with reference method by use of difference plots.

Authors:  P H Petersen; D Stöckl; O Blaabjerg; B Pedersen; E Birkemose; L Thienpont; J F Lassen; J Kjeldsen
Journal:  Clin Chem       Date:  1997-11       Impact factor: 8.327

8.  Epidural blockade suppresses lipolysis during major abdominal surgery.

Authors:  Ralph Lattermann; Franco Carli; Thomas Schricker
Journal:  Reg Anesth Pain Med       Date:  2002 Sep-Oct       Impact factor: 6.288

9.  Poor agreement between a portable armband and indirect calorimetry in the assessment of resting energy expenditure.

Authors:  S Bertoli; A Posata; A Battezzati; A Spadafranca; G Testolin; G Bedogni
Journal:  Clin Nutr       Date:  2008-02-13       Impact factor: 7.324

10.  Nutrition therapy in the critical care setting: what is "best achievable" practice? An international multicenter observational study.

Authors:  Naomi E Cahill; Rupinder Dhaliwal; Andrew G Day; Xuran Jiang; Daren K Heyland
Journal:  Crit Care Med       Date:  2010-02       Impact factor: 7.598

View more
  2 in total

1.  Evaluation of a non-invasive multisensor accelerometer for calculating energy expenditure in ventilated intensive care patients compared to indirect calorimetry and predictive equations.

Authors:  Janine Krüger; Matthias Kraft; Matthias Gründling; Sigrun Friesecke; Simone Gärtner; Lena J Vogt; Nicole Schüler; Antje Steveling; Markus M Lerch; Ali A Aghdassi
Journal:  J Clin Monit Comput       Date:  2016-09-14       Impact factor: 2.502

2.  Evaluating the predictive factors of resting energy expenditure and validating predictive equations for Chinese obese children.

Authors:  Lin Zhang; Ran Chen; Rong Li; Meng-Ying Chen; Rong Huang; Xiao-Nan Li
Journal:  World J Pediatr       Date:  2018-03-07       Impact factor: 2.764

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.