Janine Krüger1, Matthias Kraft1, Matthias Gründling2, Sigrun Friesecke3, Simone Gärtner1, Lena J Vogt1, Nicole Schüler1, Antje Steveling1, Markus M Lerch4, Ali A Aghdassi1.
Abstract
Continuous measurement of resting energy expenditure (REE) in critically ill patients remains challenging but is required to prevent malnutrition. SenseWear Pro 3 Armband (SWA) is a research grade accelerometer for assessment of REE with the advantage of easy handling. In a prospective study we compared SWA with indirect calorimetry (IC) and predictive equations in critically ill, ventilated patients. REE was measured by SWA, IC and calculated by predictive formulas. Potential confounding factors that influence REE were also recorded. Results of SenseWear Armband and indirect calorimetry were compared using the Bland-Altman method. 34 ICU patients were investigated. SWA underestimated resting energy expenditure compared to IC with a mean bias of ΔREE = -253.6 ± 333.2 kcal, equivalent to -11.7 % (p = 0.025). This underestimation was seen in both, medical (-14.9 %) and surgical (-12.9 %) patients and the bias was greater in patients with fever (-19.0 %), tachycardia (-18.7 %) or tachypnea (-26.2 %). Differences were also noted when SWA was compared to predictive formulas. At present, SWA cannot be regarded as an alternative to indirect calorimetry. Individual measurements are often inaccurate and should be used with caution until improved algorithms, based on the results of this study, have been implemented.Entities:
Keywords: Critically ill; Energy expenditure; Indirect calorimetry; Predictive equations; SenseWear Pro Armband
Mesh:
Year: 2016 PMID: 27628058 DOI: 10.1007/s10877-016-9934-5
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502