| Literature DB >> 23318392 |
Annette Haworth1, Duncan J Butler, Lisa Wilfert, Martin A Ebert, Stephen P Todd, Anna J M Hayton, Tomas Kron.
Abstract
For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23318392 PMCID: PMC5714053 DOI: 10.1120/jacmp.v14i1.4037
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Schematic overhead view of the Nucletron‐type 077.211 jig and Farmer chamber (not to scale). This jig was used to determine the AKR of the source. A modified form of this jig was used for TLD calibration Methods 1 and 2, however the source‐to‐detector distance was 50 mm for these measurements (i.e., catheter 1‐catheter 2 distance was 100 mm)
Figure 2Water phantom used in the Australian Brachytherapy Audit. This design is based on the work of Roué et al.( )
Figure 3The energy response of the TLDs, relative to 6 MV, was determined by irradiating the TLDs in a range of energies and interpolating a value at 258 k V, the mean energy of in water at a depth of 50 mm.( ) Curve fit is based on the methodology of Kron et al.( ) and is derived from the measurements made with kV, , and megavoltage X‐rays, and does not include the measurements made with the source.
Values and uncertainty estimates (1 σ) for Method 1, irradiation in air. Dwell times were calculated to deliver 1 Gy. The measured dose was 1.022 Gy (i.e., 2.2% higher than predicted). The uncertainty in d is estimated to be 0.5 mm, resulting in a 2% uncertainty in the calculated absorbed dose. The total uncertainty is the sum in quadrature of all the components and does not include the uncertainty in the TLD reading (0.7%).
|
|
|
|
|
|---|---|---|---|
|
| |||
| AKR (at 1 m) |
|
| 1 |
| Distance | 50 | mm | 2 |
| Correction for catheter attenuation, | 1.0017 | 0.2 | |
| Correction for TLD holder attenuation, | 1.013 | 2 | |
| Ratio of mass energy absorption coefficients, | 1.11 | 3 | |
| Dwell time for 1 Gy = | 258.2 | s | 4% |
|
| |||
|
| |||
|
|
| C | < 0.1 |
|
| 1.006 | 0.1 | |
|
|
|
| 1 |
| Ratio of mass energy absorption coefficients, | 1.11 | 3 | |
|
| 1.035 | 0.4 | |
| Distance | 50 | mm | 2 |
|
| 1.022 | Gy | 4% |
The results of the Brachytherapy Audit for the different TLD calibration methods. ‘Calculated’ values are derived from the AKR of the source. ‘Measured’ values are determined with an ionization chamber at the position of TLD irradiation.
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 1 | 1.04 | 1.06 | 1.02 | 1.03 | 1.04 | 1.03 | 1.08 |
| 2 | 1.03 | 1.06 | 1.01 | 1.02 | 1.04 | 1.02 | 1.07 |
| 3 | 1.01 | 1.03 | 0.99 | 1.00 | 1.01 | 1.00 | 1.05 |
| 4 | 1.03 | 1.05 | 1.00 | 1.02 | 1.03 | 1.01 | 1.07 |
| 5 | 1.00 | 1.02 | 0.98 | 0.99 | 1.00 | 0.99 | 1.04 |
| 6 | 1.01 | 1.03 | 0.99 | 1.00 | 1.01 | 1.00 | 1.05 |
| 7 | 0.99 | 1.02 | 0.97 | 0.98 | 1.00 | 0.98 | 1.03 |
| Average | 1.02 | 1.04 | 0.99 | 1.01 | 1.02 | 1.00 | 1.06 |
Values and uncertainty estimates (1 σ) for Method 2, irradiation in the Nucletron jig in water. Dwell times were calculated to deliver 1 Gy. The measured dose was 1.011 Gy (i.e., 1.1% higher than predicted). The uncertainty in d is estimated to be 0.5 mm, resulting in a 2% uncertainty in the calculated absorbed dose.
|
|
|
|
|
|---|---|---|---|
|
| |||
| AKR (at 1 m) |
|
| 1 |
| Distance | 50 | mm | 2 |
| Dose at 50 mm per unit air kerma at 1 m, | 446 | 2 | |
| Dwell time for 1 | 253.3 | s | 3% |
|
| |||
|
|
| C | < 0.1 |
|
| 1.006 | 0.1 | |
|
|
|
| 1 |
| Average stopping power ratio, water to air, | 1.137 | 0.3 | |
|
| 0.99 | 0.5 | |
|
| 0.98 | 0.5 | |
|
| 1.035 | 1 | |
| Distance | 50 | mm | 2 |
|
| 1.011 | Gy | 3% |
Values and uncertainty estimates (1 σ) for Method 3, irradiation in a modified audit phantom in water. Dwell times were calculated to deliver 1 Gy using the treatment planning system. The measured dose was 0.985 Gy (i.e., 1.5% lower than predicted). The uncertainty in d is estimated to be 0.3 mm, resulting in a 1% uncertainty in the calculated absorbed dose. The total uncertainty is the sum in quadrature of all the components and does not include the uncertainty in the TLD reading (0.7%).
|
|
|
|
|
|---|---|---|---|
|
| |||
| AKR (at 1 m) |
|
| 1.5 |
| Distance | 58.3 | mm | 1 |
| Dose at 58.3 mm per unit air kerma at 1 m, | 317.95 | 2 | |
| Dwell time for 1 | 592.8 | s | 3% |
|
| |||
|
|
| C |
|
|
| 1.006 | 0.1 | |
|
|
|
| 1 |
| Average stopping power ratio, water to air, | 1.137 | 0.3 | |
|
| 0.99 | 0.5 | |
|
| 0.98 | 0.5 | |
|
| 1.035 | 1 | |
| Distance | 58.3 | mm | 1 |
|
| 0.985 | Gy | 2% |
Values and uncertainty estimates (1 SD) for Method 4, irradiation in 6 MV photons. The total uncertainty is the sum of all errors taken in quadrature and does not include the uncertainty in the TLD reading (0.7%).
|
|
|
| ||
|---|---|---|---|---|
| Measurement | Measured dose (TRS 398) | 0.999 | Gy | 1.5 |
| SSD/Depth | 1000/100 | mm | 0.6 | |
| Energy Correction Factor | 1.03 | 5 | ||
| Total | 5% |