Design of a phantom for the quality control of high dose rate 192Ir source used in brachytherapy.

Ricardo Ochoa1, Fredy Gómez, Ivaldo H Ferreira, Federico Gutt, Carlos E de Almeida.   

Abstract

BACKGROUND AND
PURPOSE: A new phantom is proposed for measuring the strength of 192Ir high dose rate sources and for verification of the dose calculated by the treatment planning system. The complete formalism and measurement procedure for this phantom is described, as well as the preliminary results obtained in a number of centers around Brazil.
MATERIALS AND METHODS: The measurements are performed using powder thermoluminescent dosimeter capsules; the source strength is measured in air and the verification of the dose calculation algorithm in water phantom. The correction factors required to take into account the specificities related to the geometry and the phantom materials have been assessed using the PENELOPE Monte Carlo code and experimental methods. The dedicated phantom, constructed to use as part of a QA program, in this case specifically for high dose rate 192Ir brachytherapy sources, allows simultaneous irradiation of three thermoluminescent dosimeter capsules, requiring only one source stop (dwell positions).
RESULTS: The phantom was mailed to seven radiotherapy institutions in Brazil, and the results show its usefulness in verifying the source air kerma and correctness of treatment planning dose calculation in water phantom.
CONCLUSIONS: The comparison made between the phantom measurements, the well-type ionization chamber, and source specifications stated by the hospital (most of the times provided by the source manufacturer) agreed within 3% showing the quality in the HDR dose delivery in Brazilian radiotherapy centers.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17267060     DOI: 10.1016/j.radonc.2007.01.005

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  5 in total

1.  The Design of an Audit Test for 60Co Brachytherapy Treatment Planning System.

Authors:  Mohammad Javad Tahmasebi Birgani; Hadis Khorshidsavar; Ali Bagheri; Amir Danyaei; Neda Abdalvand
Journal:  J Med Signals Sens       Date:  2022-05-12

2.  A feasibility study of Fricke dosimetry as an absorbed dose to water standard for 192Ir HDR sources.

Authors:  Carlos Eduardo deAlmeida; Ricardo Ochoa; Marilene Coelho de Lima; Mariano Gazineu David; Evandro Jesus Pires; José Guilherme Peixoto; Camila Salata; Mario Antônio Bernal
Journal:  PLoS One       Date:  2014-12-18       Impact factor: 3.240

3.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30

4.  Comparison of TLD calibration methods for 192Ir dosimetry.

Authors:  Annette Haworth; Duncan J Butler; Lisa Wilfert; Martin A Ebert; Stephen P Todd; Anna J M Hayton; Tomas Kron
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

5.  The effect of tandem-ovoid titanium applicator on points A, B, bladder, and rectum doses in gynecological brachytherapy using 192Ir.

Authors:  Mohammad Hosein Sadeghi; Sedigheh Sina; Amir Mehdizadeh; Reza Faghihi; Vahed Moharramzadeh; Ali Soleimani Meigooni
Journal:  J Contemp Brachytherapy       Date:  2018-02-28
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.