Reduced-dose low-voltage chest CT angiography with Sinogram-affirmed iterative reconstruction versus standard-dose filtered back projection.

François Pontana1, Julien Pagniez, Alain Duhamel, Thomas Flohr, Jean-Baptiste Faivre, Colm Murphy, Jacques Remy, Martine Remy-Jardin.   

Abstract

PURPOSE: To evaluate image quality of low-voltage chest computed tomographic (CT) angiography with raw data-based iterative reconstruction (sonogram-affirmed iterative reconstruction) in comparison with image quality of standard-dose standard-voltage filtered back projection (FBP) CT.
MATERIALS AND METHODS: This prospective study was approved by the institutional review board, and the informed consent requirement was waived. Eighty consecutive patients who were referred for follow-up chest CT angiography underwent reduced-dose CT (hereafter, T2 examination) under technical conditions similar to those of the initial examination (hereafter, T1 examination), except the voltage selection was reduced by 20 kV with adaptation of the tube current to ensure a 50% reduction in CT dose index, and regular FBP was replaced by iterative reconstruction with sonogram-affirmed iterative reconstruction. The two techniques were compared by using paired tests (Student t test, Wilcoxon test, or McNemar test, according to the nature of variables).
RESULTS: When compared with standard-dose T1 studies, reduced-dose T2 images showed: (a) significantly less objective noise at the level of the trachea on mediastinal and lung parenchymal images (P < .001) and no significant difference in objective noise at the level of the aorta on mediastinal images (P = .507); (b) significantly higher signal-to-noise and contrast-to-noise (P < .001) ratios; (c) similar visual perception of noise on mediastinal (P = .132) and lung (P = .366) images, mainly rated as moderate; and (d) similar overall subjective image quality (P = .405).
CONCLUSION: Raw data-based iterative reconstruction yielded equivalent subjective and improved objective image quality of low-voltage half-dose CT angiograms compared with standard-dose FBP CT images for an average dose-length product of less than 80 mGy · cm in this population. SUPPLEMENTAL MATERIAL: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.12120414/-/DC1. © RSNA, 2013.

Entities:  

Mesh:

Year:  2013        PMID: 23297336     DOI: 10.1148/radiol.12120414

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  27 in total

1.  Aortic CT angiography dose reduction: investigation of optimal noise index and iterative algorithm strength in combination with low kV.

Authors:  Osvaldo Rampado; Stefania Busso; Domenica Garabello; Ezio Marengo; Marco Valerio; Simona Capello; Simona Veglia; Ottavio Davini; Roberto Ropolo
Journal:  Radiol Med       Date:  2015-12-16       Impact factor: 3.469

Review 2.  Imaging in patients after cardiac transplantation and in patients with ventricular assist devices.

Authors:  Bhanu Gupta; Dany Jacob; Randall Thompson
Journal:  J Nucl Cardiol       Date:  2015-04-02       Impact factor: 5.952

3.  Multidetector CT and three-dimensional CT angiography of upper extremity arterial injury.

Authors:  Jan Fritz; David T Efron; Elliot K Fishman
Journal:  Emerg Radiol       Date:  2014-12-11

4.  CT image quality in sinogram affirmed iterative reconstruction phantom study - is there a point of diminishing returns?

Authors:  Juan C Infante; Yu Liu; Cynthia K Rigsby
Journal:  Pediatr Radiol       Date:  2016-11-28

5.  Impact of iterative reconstruction on the diagnosis of acute pulmonary embolism (PE) on reduced-dose chest CT angiograms.

Authors:  François Pontana; Simon Henry; Alain Duhamel; Jean-Baptiste Faivre; Nunzia Tacelli; Julien Pagniez; Jacques Remy; Martine Remy-Jardin
Journal:  Eur Radiol       Date:  2015-01-31       Impact factor: 5.315

6.  High-pitch computed tomography pulmonary angiography with iterative reconstruction at 80 kVp and 20 mL contrast agent volume.

Authors:  Guang Ming Lu; Song Luo; Felix G Meinel; Andrew D McQuiston; Chang Sheng Zhou; Xiang Kong; Yan E Zhao; Ling Zheng; U Joseph Schoepf; Long Jiang Zhang
Journal:  Eur Radiol       Date:  2014-08-07       Impact factor: 5.315

7.  What is the preferred strength setting of the sinogram-affirmed iterative reconstruction algorithm in abdominal CT imaging?

Authors:  Andrew D Hardie; Rachel M Nelson; Robert Egbert; William J Rieter; Sameer V Tipnis
Journal:  Radiol Phys Technol       Date:  2014-08-28

8.  Prospective randomised comparison of diagnostic confidence and image quality with normal-dose and low-dose CT pulmonary angiography at various body weights.

Authors:  Zsolt Szucs-Farkas; Boglarka Megyeri; Andreas Christe; Peter Vock; Johannes T Heverhagen; Sebastian T Schindera
Journal:  Eur Radiol       Date:  2014-05-28       Impact factor: 5.315

9.  Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT.

Authors:  Daniele Marin; Jose J Pratts-Emanuelli; Achille Mileto; Daniela B Husarik; Mustafa R Bashir; Rendon C Nelson; Daniel T Boll
Journal:  Eur Radiol       Date:  2014-10-03       Impact factor: 5.315

10.  Repeated head CT in the neurosurgical intensive care unit: feasibility of sinogram-affirmed iterative reconstruction-based ultra-low-dose CT for surveillance.

Authors:  I Corcuera-Solano; A H Doshi; A Noor; L N Tanenbaum
Journal:  AJNR Am J Neuroradiol       Date:  2014-02-20       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.