Evaluation of hemoglobin A1c measurement by Capillarys 2 electrophoresis for detection of abnormal glucose tolerance in African immigrants to the United States.

Zhen Zhao1, Jeffrey Basilio1, Steven Hanson2, Randie R Little2, Anne E Sumner3, David B Sacks4.   

Abstract

BACKGROUND: Hemoglobin A1c (HbA1c) is used to monitor long-term glycemic control in individuals with diabetes, guide therapy, predict the risk of microvascular complications, and more recently to diagnose diabetes. An automated liquid-flow capillary electrophoresis method was recently developed to measure HbA1c using the Capillarys 2 Flex Piercing instrument.
METHODS: Analytical evaluation was performed at 2 clinical centers. A clinical analysis was conducted in 109 African-born individuals, 24% of whom have variant hemoglobin (HbAS or HbAC). Abnormal glucose tolerance (which includes both diabetes and prediabetes) was defined as 2h glucose of ≥ 140 mg/dl (7.8 mmol/l) during an oral glucose tolerance test.
RESULTS: Interlaboratory CVs were ≤ 2.1%. The method showed satisfactory correlation with 2 other analyzers that measure HbA1c by high-performance liquid chromatography. Neither labile HbA1c, carbamylated hemoglobin, uremia, bilirubin nor common hemoglobin variants (HbC/HbS/HbE) interfered. Forty-five individuals (41%) had abnormal glucose tolerance. The sensitivity of HbA1c for diagnosing abnormal glucose tolerance was 38%, 36% and 42% for total, normal and variant hemoglobin groups, respectively.
CONCLUSIONS: The analytical performance of HbA1c on the Capillarys 2 is suitable for clinical application. Variant hemoglobin in Africans did not interfere with the detection of abnormal glucose tolerance by HbA1c measured on the Capillarys 2. Published by Elsevier B.V.

Entities:  

Keywords:  Abnormal glucose tolerance; Africans; Capillarys 2; Diabetes; Hemoglobin A1c; Sickle cell

Mesh:

Substances:

Year:  2015        PMID: 25861848      PMCID: PMC4449818          DOI: 10.1016/j.cca.2015.03.025

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  38 in total

1.  Comparison of Sebia Capillarys Flex capillary electrophoresis with the BioRad Variant II high pressure liquid chromatography in the evaluation of hemoglobinopathies.

Authors:  Dina N Greene; Amy L Pyle; Judy S Chang; Carolyn Hoke; Thomas Lorey
Journal:  Clin Chim Acta       Date:  2012-04-10       Impact factor: 3.786

2.  Standards of medical care in diabetes--2014.

Authors: 
Journal:  Diabetes Care       Date:  2014-01       Impact factor: 19.112

3.  Multicenter evaluation of hemoglobin A1c assay on capillary electrophoresis.

Authors:  Mariela Marinova; Sara Altinier; Anna Caldini; Gabriella Passerini; Giorgio Pizzagalli; Marco Brogi; Martina Zaninotto; Ferruccio Ceriotti; Mario Plebani
Journal:  Clin Chim Acta       Date:  2013-06-20       Impact factor: 3.786

4.  Presidents' statement on WHO recommendation on HbA1c for diabetes diagnosis.

Authors:  Jean Claude Mbanya; Robert R Henry; Ulf Smith
Journal:  Diabetes Res Clin Pract       Date:  2011-07-28       Impact factor: 5.602

5.  First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis.

Authors:  Stéphane Jaisson; Nathalie Leroy; Julie Meurice; Emmanuelle Guillard; Philippe Gillery
Journal:  Clin Chem Lab Med       Date:  2012-10-01       Impact factor: 3.694

6.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Clin Chem       Date:  2011-05-26       Impact factor: 8.327

7.  Effects of hemoglobin C, D, E, and S traits on measurements of HbA1c by six methods.

Authors:  Chia-Ni Lin; Todd J Emery; Randie R Little; Steve E Hanson; Curt L Rohlfing; Stéphane Jaisson; Philippe Gillery; William L Roberts
Journal:  Clin Chim Acta       Date:  2012-01-08       Impact factor: 3.786

8.  Comparing analytic performance criteria: evaluation of HbA1c certification criteria as an example.

Authors:  Curt L Rohlfing; Curtis A Parvin; David B Sacks; Randie R Little
Journal:  Clin Chim Acta       Date:  2014-04-08       Impact factor: 3.786

9.  Evaluation of the Sebia CAPILLARYS 2 flex piercing for the measurement of HbA(1c) on venous and capillary blood samples.

Authors:  Olivier Heylen; Stefaan Van Neyghem; Sandra Exterbille; Charline Wehlou; Frans Gorus; Ilse Weets
Journal:  Am J Clin Pathol       Date:  2014-06       Impact factor: 2.493

10.  Metabolic syndrome does not detect metabolic risk in African men living in the U.S.

Authors:  Ugochi J Ukegbu; Darleen C Castillo; Michael G Knight; Madia Ricks; Bernard V Miller; Barbara M Onumah; Anne E Sumner
Journal:  Diabetes Care       Date:  2011-08-26       Impact factor: 19.112

View more
  3 in total

1.  Role of Glycated Proteins in the Diagnosis and Management of Diabetes: Research Gaps and Future Directions.

Authors:  Kerry J Welsh; M Sue Kirkman; David B Sacks
Journal:  Diabetes Care       Date:  2016-08       Impact factor: 19.112

2.  Single-Use Disposable Electrochemical Label-Free Immunosensor for Detection of Glycated Hemoglobin (HbA1c) Using Differential Pulse Voltammetry (DPV).

Authors:  Alireza Molazemhosseini; Luca Magagnin; Pasquale Vena; Chung-Chiun Liu
Journal:  Sensors (Basel)       Date:  2016-07-01       Impact factor: 3.576

3.  Haemoglobin J-Baltimore can be detected by HbA1c electropherogram but with underestimated HbA1c value.

Authors:  Valéry Brunel; Agnčs Lahary; Abdeslam Chagraoui; Christian Thuillez
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.