| Literature DB >> 22955659 |
Stephen F Kry1, Jimmy Jones, Nathan L Childress.
Abstract
The goal of this work was to develop and evaluate an end-to-end test for determining and verifying image-guided radiation therapy setup accuracy relative to the radiation isocenter. This was done by placing a cube phantom with a central tungsten sphere directly on the treatment table and offset from isocenter either by 5.0 mm in the longitudinal, lateral, and vertical dimensions or by a random amount. A high-resolution cone-beam CT image was acquired and aligned with the tungsten sphere in the reference CT image. The table was shifted per this alignment, and megavoltage anterior-posterior and lateral images were acquired with the electronic portal imaging device. Agreement between the radiation isocenter (based on the MV field) and the center of the sphere (i.e., the alignment point based on kV imaging) was determined for each image via Winston-Lutz analysis. This procedure was repeated 10 times to determine short-term reproducibility, and then repeated daily for 51 days in a clinical setting. The short-term reproducibility test yielded a mean 3D vector displacement of 0.9 ± 0.15 mm between the imaging-based isocenter and the radiation isocenter, with a maximum displacement of 1.1 mm. The clinical reproducibility test yielded a mean displacement of1.1 ± 0.4 mm with a maximum of 2.0 mm when the cube was offset by 5.0 mm, and a mean displacement of 0.9 ± 0.3 mm with a maximum of 1.8 mm when the cube was offset by a random amount. These differences were observed in all directions and were independent of the magnitude of the couch shift. This test was quick and easy to implement clinically and highlighted setup inaccuracies in an image-guided radiation therapy environment.Entities:
Mesh:
Year: 2012 PMID: 22955659 PMCID: PMC5718216 DOI: 10.1120/jacmp.v13i5.3939
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Custom plastic cube phantom used for IGRT tests measuring 5 cm per side with an 8 mm tungsten ball in the center.
Figure 2A sample Winston‐Lutz image and analysis. The left panel is a raw image from the electronic portal imaging device. The right panel shows analysis of the Winston‐Lutz image, including differences (Δ) between the center of the radiation field and the tungsten ball centroid.
Residual difference between IGRT‐aligned isocenter (tungsten sphere centroid) and radiation field isocenter. Data are results of 10 sequential exposures of the phantom (no movement between exposures). Range is the absolute value of the difference between the maximum and minimum values for each set of exposures at each angle.
|
| ||||||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
| Standard Deviation | 0.03 | 0.03 | 0.02 | 0.01 | 0.03 | 0.03 | 0.01 | 0.02 |
| Range | 0.08 | 0.11 | 0.08 | 0.02 | 0.09 | 0.12 | 0.04 | 0.04 |
LAT: lateral dimension; LNG: longitudinal dimension; VRT: vertical dimension.
Residual difference between IGRT‐aligned isocenter (tungsten sphere centroid) and radiation field isocenter, along with the standard deviation and magnitude of the maximum difference. Data are results of 10 IGRT‐positioned phantom exposures conducted with two different linear accelerators for a total of 20 exposures. The standard deviation of the mean (standard error) for both linear accelerators was 0.15 mm.
|
|
| ||||
|---|---|---|---|---|---|
|
|
|
|
|
|
|
| 1 Mean | 0.0 | 0.0 | 0.0 | 0.5 | 0.9 |
| Std. Deviation | 0.7 | 0.4 | 0.4 | 0.4 | 0.2 |
| Max | 0.9 | 0.9 | 0.7 | 0.9 | 1.0 |
| 2 Mean | 0.4 | 0.3 | 0.6 | 0.7 | 0.9 |
| Std. Deviation | 0.3 | 0.2 | 0.1 | 0.2 | 0.1 |
| Max | 0.7 | 0.5 | 0.7 | 0.9 | 1.1 |
LAT: lateral dimension; LNG: longitudinal dimension; VRT: vertical dimension.
Figure 3The residual difference between the IGRT‐aligned isocenter (the tungsten sphere centroid) and radiation treatment isocenter for 51 days of Winston‐Lutz analysis. Each day the phantom was imaged twice, once with an initial setup offset of 5.0 mm from isocenter, and one with a random offset.
Residual difference between IGRT‐aligned isocenter (tungsten sphere centroid) and radiation field isocenter, along with the standard deviation and magnitude of the maximum difference. Data are results of 51 IGRT‐positioned phantom exposures. The phantom was set up daily with either a 5.0 mm offset from the isocenter or a random offset of 2.0–23.0 mm in each dimension. The standard deviation of the mean (standard error) of the 3D vector was 0.08 mm.
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| 5.0 mm | Mean | 0.5 | 0.2 | 0.1 | 0.7 | 1.1 |
| Std. Deviation | 0.5 | 0.7 | 0.4 | 0.7 | 0.4 | |
| Max | 1.5 | 1.5 | 0.5 | 1.9 | 2.0 | |
| Random | Mean | 0.1 | 0.2 | 0.0 | 0.6 | 0.9 |
| Std. Deviation | 0.5 | 0.6 | 0.5 | 0.6 | 0.3 | |
| Max | 1.0 | 1.5 | 0.7 | 2.0 | 1.8 | |
LAT: lateral dimension; LNG: longitudinal dimension; VRT: vertical dimension.