| Literature DB >> 22941391 |
Kelli L Hancock1, Nadine R Lawson, Elizabeth C Ward.
Abstract
Device life of the Provox Vega Indwelling voice prosthesis is as yet untested outside Europe. The current study examined device life and reasons for replacement within an Australian clinical setting. Twenty-three participants were monitored for device life and reasons for replacement. Main outcome measure was days to failure of initial device. Average device life and reasons for replacement were secondary measures. Initial device life data revealed 67 % had functioning devices at 3 months, 52 % at 6 months and 29 % at 12 months. Average device life was 207 days (median of 222). The majority of devices (97 %) failed due to leakage through the prosthesis. The Provox Vega Indwelling voice prosthesis had favourable device life in this cohort of patients and in comparison to European data. Reasons for replacement were consistent with international literature.Entities:
Mesh:
Year: 2012 PMID: 22941391 PMCID: PMC3608863 DOI: 10.1007/s00405-012-2154-9
Source DB: PubMed Journal: Eur Arch Otorhinolaryngol ISSN: 0937-4477 Impact factor: 2.503
Fig. 1Details and justification for patient numbers and attrition throughout study
Fig. 2Kaplan–Meier survival curves of the patients’ initial (n = 21) and second (n = 17) Provox Vega
Summary Information regarding sizes and reasons for replacement of initial and second Vega within 12 month mark (n = 31)
| Device | Outer diameter | Length | Reasons for replacement ( |
|---|---|---|---|
| First Vega ( | 20 Fr: 11 | 4 mm: 1 | Leakage through: 16 (52 %) |
| 17 Fr: 6 | 6 mm: 1 | ||
| 8 mm: 7 | Increased speaking effort: 1 (3 %)a | ||
| 10 mm: 8 | |||
| Second Vega ( | 20 Fr: 8 | 6 mm: 2 | Leakage through: 14 (45 %) |
| 17 Fr: 6 | 8 mm: 3 | ||
| 10 mm: 7 | |||
| 12.5 mm: 2 |
a17Fr 8 mm Provox Vega