Abstract
The ultrasonic cardiac output monitor is a non-invasive, continuous-wave Doppler monitor that can be used to measure cardiac output. Its accuracy and precision in peri-operative and critical care has been assessed with varied results. We conducted a meta-analysis to calculate a pooled mean bias, precision and percentage error for this device. Six studies were suitable for inclusion into a pooled parametric analysis consisting of 320 measurements. The mean weighted bias was -0.39 l.min(-1) (95% CI -0.25 to -0.53 l.min(-1) ), precision 1.27 l.min(-1) and percentage error 42.7% (95% CI 38.5-46.9%). The ultrasonic cardiac output monitor achieved similar agreement with bolus thermodilution to that of other minimally invasive methods of peri-operative cardiac output monitoring, and may have a useful role in cardiac output monitoring. AnaesthesiaMesh:
Year: 2012 PMID: 22928650 DOI: 10.1111/j.1365-2044.2012.07311.x
Source DB: PubMed Journal: Anaesthesia ISSN: 0003-2409 Impact factor: 6.955