Next generation drug-eluting stents: focus on bioabsorbable platforms and polymers.

Brendan Doyle1, David R Holmes.   

Abstract

The success of drug-eluting stents in preventing restenosis has shifted the focus of new stent development toward enhancing long term safety and efficacy of these devices, while simultaneously eliminating the need for indefinite dual antiplatelet therapy. A technical advance fulfilling these aims would hold tremendous potential to reduce morbidity, mortality and economic costs associated with the percutaneous treatment of coronary artery disease. An attractive approach is the use of bioabsorbable stent designs. These may include stents with different bioabsorbable drugs, bioabsorbable polymers or even bioabsorbable metallic backbones. A device that could achieve excellent acute and long-term results, but disappear completely within months (thereby avoiding the need for prolonged dual antiplatelet therapy), would be a tremendous advance. Too good to be true? We explore here the scientific rationale and prospects for success with this exciting concept.

Entities:  

Keywords:  bioabsorbable; biodegradable; percutaneous coronary intervention; polymer; stent

Year:  2009        PMID: 22915914      PMCID: PMC3417859          DOI: 10.2147/mder.s5617

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


Introduction

The success of drug-eluting stents (DES) in preventing restenosis has shifted the focus of new stent development toward enhancing long term safety and efficacy of these devices, whilst simultaneously eliminating the need for indefinite dual antiplatelet therapy. A technical advance fulfilling these aims would hold tremendous potential to reduce morbidity, mortality and economic costs associated with the percutaneous treatment of coronary artery disease. Myocardial infarction (MI) occurring late after percutaneous coronary intervention (PCI) with DES is a complex problem. Numerous mechanisms have been described which may be responsible for acute thrombotic occlusion within the treated segment. Those cited most often include hypersensitivity reactions to the drug/polymer coating, inadequate endothelialization of the stent, and aneurysmal remodeling of the vessel wall leading to stent malapposition. Although these processes are biologically distinct, the end result is the same; the risk for generation of a localized thrombogenic milieu which may not resolve, due to the permanent presence of the metallic platform and polymer. In this context, the role of dual antiplatelet therapy in contemporary practice has become a contentious issue. Although current recommendations for prolonged use of aspirin and a thienopyridine were directed to address some of the safety problems which have been highlighted, the efficacy of this approach has yet to be proven in randomized trials. Furthermore, this strategy increases the long-term bleeding risk and expense associated with PCI, as well as complicating subsequent medical care (such as the need for invasive surgical procedures, or co-existing requirements for anticoagulation with warfarin). The emerging issue of antiplatelet drug resistance is yet another concern, although the true extent to which this phenomenon exists and contributes to late adverse events following PCI is still unknown. An attractive approach to solving many of these problems is the concept of bioabsorbable stent designs. These may include stents with different bioabsorbable drugs, bioabsorbable polymers or even bioabsorbable metallic backbones. Such a device that could achieve excellent acute and longterm results, but disappear completely within months and avoid the need for prolonged dual antiplatelet therapy, would be a tremendous advance. Too good to be true? We explore here the scientific rationale and prospects for success with this exciting concept.

Late myocardial infarction after PCI

Reports of late thrombosis of DES after these devices were approved for use in routine clinical practice prompted a major controversy about their safety. Evidence of incomplete re-endothelialization, aneurysmal vessel wall remodeling and local hypersensitivity reactions in DES-treated vessels suggested that impaired post-intervention healing of the vessel wall may provide a substrate for late (30 days to 1 year) and very late (>1 year) thrombotic events.1–5 Bare metal stents (BMS) do not possess an anti-proliferative coating and although the risk of restenosis is higher with BMS, this was traditionally perceived as a relatively benign complication.6–8 This led to a resurgence in use of BMS at the height of the DES thrombosis controversy. Such assumptions regarding the benign nature of restenosis would tend to support an approach to future stent development centered upon refinement of the drug–polymer combination, rather than a fundamental re-evaluation of the need for a durable metallic platform. Such intense focus on safety led ultimately to a number of important new perspectives on late adverse events, and how these may relate to the index procedure and the kind of device that was used. In particular, the concepts that (1) late stent thrombosis did not occur with BMS and (2) that BMS restenosis was benign was examined in a study of 4503 patients treated with BMS and dual antiplatelet therapy at the Mayo Clinic between 1994 and 2000.9 The cumulative incidence of stent thrombosis was 0.5% at 30 days (95% confidence interval [CI]; 0.3%–0.7%), 0.8% at 1 year (95% CI; 0.6%–1.1%) and 2.0% at 10 years (95% CI; 1.5%–2.5%). Risk of late (30 days to 1 year) and very late (>1 year) BMS thrombosis was increased among patients considered ‘off label’ for BMS use (P = 0.024). Mortality was markedly increased following late and very late BMS thrombosis, particularly during the first 30 days (hazard ratios: 22 [95% CI 3.1–159] and 40 [95% CI 15–107] respectively). The 10 year incidence of clinical restenosis was 18.1% (95% CI; 16.5%–19.7%); many of these patients presented with an acute coronary syndrome and some presented with MI in 2.1% (95% CI; 1.6%–2.6%). Restenosis presenting with MI was associated with increased mortality compared to no restenosis (hazard ratio 2.37, P < 0.001) and to restenosis with a non-MI presentation (hazard ratio 2.42, P < 0.001). The implications of these findings for future stent development are important, as they introduce doubt about the notion that safety can be enhanced (and the need for indefinite dual-antiplatelet therapy avoided) simply by modifying the drug–polymer coating on the bare metal stent platform. The big question is whether the long-term presence of the stent platform itself is actually a help or sometimes a hindrance in our efforts to improve outcomes. As technology in biodegradable materials has evolved, the potential to manufacture a drug-eluting stent that is partially or completely bioabsorbable has now taken center stage as a novel and extremely exciting therapeutic concept.

Eliminating the permanent polymer

The first generation of polymer-based DES rapidly replaced bare metal stents as the treatment of choice for percutaneous coronary revascularization. However, the long-term safety of polymer-based DES has been called into question because of concerns about late stent thrombosis secondary to impaired arterial healing, characterized by delayed re-endothelialization and persistence of fibrin.2,7,10 Emerging evidence suggests that drug delivery polymers may play an important role in the pathophysiology of impaired healing by provoking inflammatory cell infiltration and/or causing long-term drug sequestration within the arterial wall.5,11,12 In addition, there are some data to indicate that the use of polymer-based DES for interventions involving multiple stents (such as bifurcations or overlapping stents) may further exacerbate local arterial toxicity, which occurs when drug and polymer concentrations are substantially increased.13–15 Furthermore, recent clinical data demonstrates ongoing reduction in luminal calibre beyond 6 to 8 months postintervention16 – the time point at which completion of vessel-wall healing was observed in the bare metal stent era.17 The common thread linking both late stent thrombosis and late erosion of luminal caliber (‘late luminal creep’) seems to be the existence of a persistent inflammatory response within the coronary vessel-wall following polymer-coated DES implantation.16 A possible culprit appears to be the residual presence of permanent polymer following completion of its functional role.2,5,18

Bioabsorbable polymers

Various strategies have been proposed to eliminate the potential harm associated with permanent polymer-based DES. One of these solutions is the use of bioabsorbable polymer. A recent study compared the healing and inflammatory responses of polymer-free bare-metal stents (BMS), polymer-free sirolimus-eluting stents (SES) and polymer-free sirolimus-eluting stents plus estradiol (SES + ED) to Cypher® drug-eluting stents (CDES) in a rabbit model of overlapping stent placement.10 Twenty-eight rabbits received 2 overlapping stents in each iliac artery: SES, SES + ED, BMS, or CDES, and vessels were harvested at 28 days for histology and scanning electron microscopy. Although similar at nonoverlapping segments, neointimal thickness within the overlap site of CDES was significantly less than in SES, SES + ED, and BMS (0.07 ± 0.04 mm vs 0.16 ± 0.03 mm, 0.14 ± 0.03 mm, and 0.15 ± 0.03 mm, P < 0.0001). Endothelialization was greater in SES, SES + ED, and BMS compared with CDES in nonoverlapping sections (80.0% ± 5.0% vs 95.3% ± 5.0%, 97.5% ± 2.5%, and 96.7% ± 3.8%; P = 0.0028) and overlapping sections (85.8% ± 2.9% vs 90.8% ± 6.3%, 89.2% ± 6.3%, and 48.3% ± 2.9%; P < 0.0001). The number of luminal eosinophils was also less in overlapping sections of SES, SES + ED, and BMS versus CDES but was similar in nonoverlapping sections. The authors concluded that polymer-free stents coated with SES or SES + ED result in less robust neointimal suppression, but markedly improved arterial healing compared with CDES in the rabbit model.10 A number of biodegradable polymer-based DES have been tested in humans. One such device is the EXCEL sirolimus eluting stent (JW Medical Systems, China, Registration certification no. 2005DI3461514) which uses a novel polylactic acid (PLA) material. This material is gradually biodegraded within approximately 6 months, converting to water and carbon dioxide upon breakdown. The stent platform is a laser-cut, 316L stainless steel, open cell design stent with strut thickness of 0.0047 inches. The PLA polymer is abluminally coated (only applied to the surface of the stent facing the arterial wall). Total sirolimus dosage varies from 195 to 376 μg per stent according to the stent length and diameter, and is evenly mixed within the PLA coating. In a study of 100 consecutive patients (153 lesions) receiving 6 months of dual-antiplatelet therapy treated at a single center, the EXCEL stent was associated with an encouragingly low 4.0% major adverse cardiovascular event rate at 12 months.19 During 24 months of follow up there were no cases of stent thrombosis, cardiac death or myocardial infarction. Quantitative coronary angiography at 12 months demonstrated 5.4% in-segment restenosis (of which 3.6% was in-stent). These preliminary data therefore provide encouraging initial safety and efficacy for this approach. A much larger study with a biodegradable PLA polymer (BioMatrix Flex®; Biosensors Inc, Newport Beach, CA, USA) used a biolimus-eluting stent versus a sirolimus-eluting stent with durable polymer (Cypher SELECT Cypher SELECT; Cordis, Miami Lakes, Fl, USA) has provided further encouraging data on the use of this novel polymer.20 1707 patients with chronic stable coronary disease or acute coronary syndromes were enrolled in a multicenter, assessor-blind, noninferiority study in 10 European centers. Biolimus-eluting stents were found to be noninferior for the primary endpoint (composite of cardiac death, MI or clinically driven target lesion revascularization) at 9 months (79 [9%] patients vs 89 [11%], rate ratio 0.88 [95% CI 0.64–1.19], P for non-inferiority = 0.003, P for superiority = 0.39). Frequency of cardiac death (14 [1.6%] vs 21 [2.5%], p for superiority = 0.22), myocardial infarction (49 [5.7%] vs 39 [4.6%], P = 0.30), and clinically-indicated target vessel revascularization (38 [4.4%] vs 47 [5.5%], P = 0.29) were similar for both stent types. 168 (79%) patients in the biolimus-eluting group and 167 (78%) in the sirolimus-eluting group had data for angiographic follow-up available. Biolimus-eluting stents were noninferior to sirolimus-eluting stents in in-stent percentage diameter stenosis (20.9% vs 23.3%, difference −2.2% [95% CI −6.0 to 1.6], P for noninferiority = 0.001, P for superiority = 0.26). The results indicate that a stent eluting biolimus from a biodegradable polymer represents a safe and effective alternative to a stent eluting sirolimus from a durable polymer in patients with chronic stable coronary artery disease or acute coronary syndromes. Nevertheless, the trial provides no insight as to whether the bioabsorbable polymer which characterizes this device will distinguish this biolimus-eluting stent from other stent platforms currently approved by the US Food and Drug Administration (FDA), in terms of increased safety (reduced rates of stent thrombosis) or shorter duration of dual antiplatelet therapy. Further studies will be required to address these issues with particular emphasis on longer follow up (significantly beyond the 9 months timeframe taken for the polymer to dissolve, during which time – in theory – greater benefit should accrue). In contrast to the encouraging data presented above, there have also been some notable failures in this arena. The MAHOROBA study examining a tacrolimus-eluting bioabsorbable polymer-coated cobalt chromium stent demonstrated failure of this device to prevent neointimal hyperplasia, with a composite MACE (major adverse cardiac event) rate of 23.4% among a pilot study of 47 patients.21 Tests of a stent with a biodegradable polylactic-co-glycolic acid polymer releasing paclitaxel have shown this to be safe but less effective than an approved and widely used paclitaxel-eluting stent (TAXUS®), presumably because of ineffective pharmacokinetics associated with the novel polymer.22 Recent data examining the potential to reduce anti-proliferative drug dose in combination with a bioabsorbable polymer have indicated this may be another means of achieving greater safety without compromising efficacy.23 In a porcine study of low- and high-dose sirolimus release (25, 40, or 100 μg) from hydroxyapatite (Hap) with Cypher (Cordis; Johnson & Johnson, Warren, New Jersey, USA) (111 μg sirolimus), similar efficacy in prevention of neointimal hyperplasia was demonstrated, with increasing signs of delayed vascular healing in a dose-dependent manner. Cypher induced the highest amount of fibrinoid (considered to be a marker of delayed healing).23

Polymer-free DES

An alternative to use of a bioabsorbable polymer is complete avoidance of the use of polymer, applying the drug directly to the stent. Producing the same predictable and controlled release of drug over a sustained period is clearly more challenging without recourse to a polymer, but early human data of polymer-free DES have been encouraging. One such study compared a novel polymer-free rapamycin- and probucol-eluting stent (Dual-DES) with a commercially available permanent polymer-based sirolimus-eluting stent (SES; Cypher) and with a zotarolimus- eluting (ZES; Endeavor) stent.16 This device utilizes a sand-blasting technique to generate micropores on the thin strut (87 μm) stent surface which can then be directly coated using a mixture of the active compounds and a shellac resin (0.07%) with no need for a polymer. A total of 1007 patients undergoing coronary stenting of de novo lesions, in native vessels, were randomized to treatment with SES (n = 335), Dual-DES (n = 333), or ZES (n = 339). The primary endpoint was binary angiographic restenosis at 6- to 8-month follow-up angiography. Secondary endpoints were angiographic in-stent late loss; and target lesion revascularization (TLR), death/MI and stent thrombosis at 12 months. Follow-up angiographic data were available for 828 (82.2%) patients. There was a significant difference in both binary restenosis and TLR across treatment groups (P = 0.003 and P < 0.001, respectively). Binary restenosis in the Dual-DES group (11.0%) was significantly lower than that in the ZES group (19.3%; P = 0.002) but comparable with that in the SES group (12.0%; P = 0.68). Similarly, TLR with Dual-DES (6.8%) was significantly lower than ZES (13.6%; P = 0.001) but not different to that of SES (7.2%; P = 0.83). These differences were mirrored in the extent of late loss across the groups. No differences were observed between stent groups in terms of death/MI or stent thrombosis. These data support an approach to DES design that utilizes novel polymer-free drug dosing, while maintaining high antirestenotic efficacy. The potential of this approach to increase safety whilst reducing the need for prolonged dual antiplatelet therapy remains to be tested in larger cohorts however, with wider and more clinically relevant inclusion criteria. Such studies now being planned include the NEVO II trial. This randomized study will compare the NEVO™ Sirolimus-eluting Coronary Stent to the XIENCE V™ Everolimus-eluting Coronary Stent and aims to enroll 2000 patients. The NEVO stent utilizes a reservoir technology incorporating hundreds of small reservoirs, each acting as a depot into which drug(s) may be loaded.

Bioabsorbable stent platforms

An ambitious project to develop a fully biodegradable coronary stent is now underway. There are numerous theoretical advantages to such a device. The benefits of initial scaffolding to prevent acute recoil and treat dissection could still be maintained, but without the subsequent presence of a permanent prosthesis in the artery which may act as a substrate for late thrombotic events, or impede subsequent surgical revascularization. The potential to safely discontinue dual-antiplatelet therapy within a reasonable timeframe of several months post- PCI is another enormous attraction. As noninvasive imaging techniques such as CT improve, the avoidance of a permanent metallic stent in the coronary artery (associated with blooming artifact) is also likely to be advantageous. The technological challenge should not be underestimated. The required characteristics of sufficient mechanical strength to prevent negative vessel remodeling and avoid stent deformity/strut fractures, while still maintaining the capacity for full dissolution of the stent in a timely manner, will require tradeoffs between these competing aims. Whether such tradeoffs can actually be made in such as way as to improve overall safety and efficacy when compared to existing (or polymer-refined) DES is far from certain. The first animal studies of biodegradable stents were performed in the late 1980s. Stack et al24 developed a prototype poly-L-lactide (PLA) stent that could withstand up to 1000 mmHg of crush pressure. They designed the stent to maintain its radial strength for 1 month, with complete resorption of the device complete by 9 months post-implant. There were several reasons why this and similar biodegradable devices did not prosper and eventually succumbed to their bare-metal stent competitors. These included difficulties in manufacturing compatible biodegradable materials that had molecular weights capable of limiting inflammation and restenosis;25 concerns about loss of radial strength over time that increased the risk for stent fracture and migration of degraded products;26,27 and concerns about long-term biocompatibility of breakdown products.28 Finally, swelling of the biodegradable stent as it absorbs water to dissolve was also thought to exacerbate neointimal hyperplasia.29 Biodegradable stents that are currently in development fall broadly into one of two categories: those manufactured entirely from polymers versus devices that are made from bioabsorbable metals.30

Polymer-based biodegradable stents

Of the polymeric stents, greatest interest has focused on use of PLA as the polymer of choice. Polylactide polymers are already used in a variety of medical settings, including orthopedic prostheses and absorbable sutures. This polymer is metabolized through the Kreb’s cycle (degraded to carbon dioxide and water over a period of 1–2 years) with small residual particles <2 μm in diameter capable of being phagocytosed by macrophages. The Igaki-Tamai® stent (Igaki Medical Planning Co. Ltd., Kyoto, Japan) was the first PLA stent tested in humans.31 This stent was deployed using a thermal balloon expandable covered sheath system, which itself posed a number of potential problems. First, the heat required for rapid expansion (65–75 ºC for a few seconds) may produce arterial wall necrosis leading to proliferation of smooth muscle cells.32 Second, temperatures above 55 ºC increase platelet adhesion raising concerns regarding stent thrombosis.33 Initial experience however in a pilot study of 16 patients was encouraging, with late loss similar to that expected with a bare metal stent.31 In a larger study of 50 patients with 63 lesions (including lengths of 13.5 mm and a reference vessel diameter of 2.95 mm) only one patient had a thrombotic event resulting in a Q-wave myocardial infarct at 1-year follow-up.30 In the same study, quantitative coronary angiography at 3, 6, and 12 months demonstrated percentage diameter stenoses of 12 ± 8, 38 ± 23, and 33 ± 23, respectively. Restenosis rates were 21% at 6 months and 19% at 12 months; the incidence of restenosis seemed to peak between 3 and 6 months, after which it decreased. The target lesion revascularization rate was 12% at 6 months and 17% at 12 months. There were no deaths or surgical revascularizations. At 4-year follow-up, 18% of the patients had repeat percutaneous coronary intervention (somewhat higher than seen with DES), there was one death, and the MACE-free event survival, as determined by a Kaplan Meier curve, was 82%. Further animal studies have since been performed using drug-coated versions of this device which successfully limited restenosis,28,34 although human studies are still awaited. Another device combining a PLA platform with the capability to elute everolimus has been developed by the Biosensors International company (Singapore). In this iteration, the PLA was cross-linked rather than linear (as in the Igaki-Tamai stent), a design feature thought to encourage preferential surface rather than bulk degredation. This should in theory ensure a more gradual change in the mechanical properties of the stent over time. Indeed, in animal models, there was no apparent loss in mechanical strength over the first 4 to 6 weeks following implantation.30 The only biodegradable drug-eluting PLA stent currently in clinical trials is the bioabsorbable everolimus-eluting coronary stent (BVS; Abbott Vascular, Santa Clara, CA, USA). This device has a backbone of circumferential hoops of highly crystalline PLA, with a strut thickness of 150 μm, linked directly together or held by straight bridges24 achieving a radial strength similar to the Multi-link Vision metal stent (Abbott Vascular). At both ends of the stent are two adjacent radio-opaque metal markers to facilitate correct fluoroscopic positioning. On the surface, there is a thin coating of a 1:1 mixture of everolimus 8.2 μg/mm and amorphous PLA matrix, allowing 80% of the drug to be released in 30 days. This is similar to the release patterns of the Xience V™ (Abbott Vascular) and Cypher drug-eluting stents. Animal studies comparing the BVS PLA everolimus-eluting coronary stent with a control DES (Cypher) have shown that the inflammatory response is comparable at 6 and 9 months and significantly reduced at 12 months.35 There was complete luminal endothelialization by 28 days with a thin complete neointima composed of compact smooth muscle cells in a proteoglycan/collagen matrix overlying widely patent lumens. Two year results from the ABSORB (a bioabsorbable everolimus-eluting coronary stent system) trial were recently published.36 In this study, 30 patients with a single de-novo coronary artery lesion were followed up for 2 years clinically and with multiple imaging methods: multislice CT, angiography, intravascular ultrasound, derived morphology parameters (virtual histology, palpography, and echogenicity), and optical coherence tomography (OCT). Clinical data were obtained from 29 of 30 patients. At 2 years, the device was safe with no cardiac deaths, ischaemia-driven target lesion revascularisations, or stent thromboses recorded, and only one MI (non-Q wave). 18-month multislice CT (assessed in 25 patients) showed a mean diameter stenosis of 19% (SD 9). At 2-year angiography, the in-stent late loss of 0.48 mm (SD 0.28) and the diameter stenosis of 27% (11) did not differ from the findings at 6 months. The luminal area enlargement on OCT and intravascular ultrasound between 6 months and 2 years was due to a decrease in plaque size without change in vessel size. At 2 years, 34.5% of strut locations presented no discernible features by OCT, confirming decreases in echogenicity and in radiofrequency backscattering; the remaining apparent struts were fully apposed. Additionally, vasomotion occurred at the stented site and adjacent coronary artery in response to vasoactive agents indicating that more normal endothelial function was present. The authors concluded that at 2 years after implantation the stent was partially bioabsorbed, had vasomotion restored and restenosis prevented, and was clinically safe, suggesting freedom from late thrombosis. Although these data are encouraging, the small numbers of patients enrolled in the trial and the simplicity of the lesions treated means that a great deal of work remains to be done before these devices can be considered for use in routine clinical practice. Another biodegradable PLA-based stent in development is an offspring of the Genous™ Bio-engineered cobalt-chromium R-stent™ (OrbusNeich, Fort Lauderdale, FL, USA). This device was designed to capture circulating endothelial progenitor cells (EPC) from the circulation by coating the surface of the stent with immobilized antibodies which bind to EPC-surface antigens. This approach should in theory lead to more rapid and complete endothelialization of the device. The bioabsorbable iteration replaces the metallic platform with a PLA-platform. Tantalum markers at both ends of the stent facilitate accurate placement under fluoroscopy. The abluminal surface is coated with a prohealing drug held in a bioabsorbable matrix, and the luminal side of the stent has the EPC-capturing antibodies. Incorporated within a hybrid-copolymer backbone is the antiproliferative drug with a programmable release profile. This stent is still in development and at this time no preclinical data have been published.30 A variation on the PLA-based design is the REVA stent (Boston Scientific Corporation, Natick, MA, USA). It is made from monomeric units of the common amino acid L-tyrosine, chemically modified to incorporate iodine molecules so as to make this stent radio-opaque. This polymer degrades to carbon dioxide and water in a similar fashion to PLA. It also forms ethanol. The stent struts are designed with a ‘slide and lock’ feature that enables strut thickness to be reduced without compromising radial integrity. Preclinical data have shown that at 30 days there is complete endothelialization of the bare REVA stent.30 Restenosis assessed by intravascular ultrasound was similar when compared to a bare metal stent at 12 months follow up. Interestingly, when compared to bare metal stents which displayed a relatively fixed luminal area over time, the REVA stent showed an increase in the luminal area from 3.65 mm2 at 1 month to 8.28 mm2 at 1 year. Boston Scientific is now developing a further evolution of this device which incorporates paclitaxel, whereby 50% of the antirestenotic drug will be eluted by day 10, and 90% by 3 months. Enrollment in the RESORB trial (REVA Endovascular Study of a Bioabsorbable Coronary Stent) to test this device in 60 patients with short de novo lesions is expected to commence in late 2009. Another device currently in development is the IDEAL™ stent (Bioabsorbable Therapeutics Inc., Menlo Park, CA, USA) which incorporates salicylic acid into the backbone of a polyanhydride ester (PAE) polymeric stent. The stent surface is coated with sirolimus, with compete elution of the drug by 30 days. The addition of salicylic acid in this device is designed to limit inflammation. When compared to bare metal stents, the PAE-salicylic acid polymer reduced angiographic restenosis at 30 day follow up30. Furthermore, the IDEAL stent reduced the inflammatory response and neointimal growth observed when compared to the Cypher stents. These preliminary findings from animal studies are encouraging and clinical data are eagerly awaited.

Biodegradable metallic stents

An absorbable metallic stent manufactured from magnesium has been developed by Biotronik (Berlin, Germany). This absorbable magnesium stent (AMS) is composed of 93% magnesium and 7% rare-earth metals laser cut from a single tube of alloy. It degrades within the body over a 2- to 3-month timeframe, forming inorganic salts containing calcium, chloride, oxide, sulfates and phosphates. The deposition of calcium within the vessel wall as a byproduct of magnesium degredation is clearly of some concern, and will be a particular focus of safety monitoring as this device is evaluated. The first generation stents had mechanical properties similar to current bare metal stents exhibiting high collapse pressure (0.8 bar), low elastic recoil (<8%) and minimum shortening after inflation (<5%). This device was first tested in the PROGRESS-AMS trial.37 A total of 63 patients (44 men; mean age 61.3) in 8 centers with single de novo lesions in a native coronary artery were enrolled in a multicenter, nonrandomized prospective study. Follow-up included coronary angiography and intravascular ultrasound at 4 months and clinical assessment at 6 months and 12 months. The primary endpoint was cardiac death, nonfatal MI, or clinically driven target lesion revascularization at 4 months. Among 63 patients a total of 71 stents were deployed, 10 to 15 mm in length and 3.0 to 3.5 mm in diameter. Diameter stenosis was reduced from 61.5 (SD 13.1%) to 12.6 (5.6%) with an acute gain of 1.41 mm (0.46 mm) and in-stent late loss of 1.08 mm (0.49 mm). The ischemia-driven target lesion revascularization rate was 23.8% after 4 months, and the overall target lesion revascularization rate was 45% after 1 year. No MI, subacute or late thrombosis, or death occurred. Angiography at 4 months showed an increased diameter stenosis of 48.4 (17.0%). After serial intravascular ultrasound examinations, only small remnants of the original struts were visible. This study demonstrated the feasibility of a magnesium-based bioabsorbable stent with acceptable safety and efficacy in this initial small and highly selected cohort. Further evolution of the AMS device has targeted greater radial strength through modification of the magnesium-alloy composition. This AMS-2 stent should achieve a collapse pressure of 1.5 bar (compared to 0.8 with AMS-1), thereby limiting recoil to only 3%. A drug-eluting version of this device is also now in development. Iron has been proposed as an alternative metal to magnesium as a basis for bioabsorbable metallic stents. However, problems with the slow degredation time for iron-based devices (several years) and concerns regarding systemic iron-related toxicity have so far hampered progress with this approach.

Conclusions and caveats

There are a number of assumptions that underpin the hypothesis that bioabsorbable stents will lead to improved outcomes for patients when compared to a refined version of conventional DES. It is worth considering each of these in turn.

Assumptions on dual antiplatelet therapy

Currently, indefinite dual antiplatelet therapy is often routinely prescribed for all but the simplest DES procedures in the belief (or hope) that this approach reduces risk for late ischemic events, leading to better overall outcomes. It is assumed that by using a bioabsorbable stent the need for prolonged dual antiplatelet therapy will be eliminated with similar or better outcomes achieved. It is important to emphasize that there are currently no data for or against the use of prolonged dual antiplatelet therapy (DAPT) in patients treated with DES. It is entirely possible that this approach produces the best outcomes for patients who require PCI, but not necessarily because of a reduction in stent thrombosis. It is possible that prolonged DAPT is an excellent way to treat high risk patients with coronary disease who have required intervention, regardless of stent-specific complications. Perhaps the benefits may extend to territories not treated with PCI or even to other vascular beds (such as in the prevention of stroke). If such data demonstrating a net benefit of prolonged DAPT on non-stent related outcomes were to emerge, then perhaps the biggest theoretical advantage of bioabsorbable DES would become irrelevant. Having the option of safely discontinuing DAPT in a patient who had received a bioabsorbable device would still remain as a potential advantage, but this too is an area where there remains tremendous uncertainty; an assumption that ischemic events following discontinuation of DAPT would be lower after use of a bioabsorbable stent should certainly not be taken for granted. Even among patients with acute coronary syndromes who are not treated with medical therapy alone (and no coronary stent) there is an excess of ischemic events after clopidogrel discontinuation.38

Biodegradable = biocompatible?

It is worth emphasizing that a biodegradable polymer is not necessarily a biocompatible polymer. During degradation, polymers are broken down into molecules that are metabolized and removed from the body via normal metabolic pathways. In addition to these degradation products, many degradable materials release by-products that may reduce biocompatibility. PLA is an attractive polymer because its degradation involves hydrolysis of its ester bonds, which produces lactic acid, which is ultimately converted to water and lactic acid. However several studies have also suggested that PLA can produce an inflammatory response.25 Despite all the publicity, the fact remains that late stent thrombosis is a rare complication of conventional DES and by switching to bioabsorbable DES we may simply be replacing one mechanism for this rare complication with another. Calcification of the vessel wall during magnesium degradation remains an additional concern.

Incremental gains in the field of PCI

Advances in PCI over the past 2 decades have been remarkable. Although some obstacles and problems remain, overall the outcomes for an increasingly complex and aging population who are being treated with PCI are excellent. Even in these patients, the absolute specific frequency of following adverse events is low, requiring very large numbers of patients to document statistically significant and more important, statistically meaningful improvements in outcome with new device strategies. In the future, incremental advances will require increasingly greater effort for smaller and smaller rewards – this at a time when health budgets are coming under intense scrutiny and pressure. Whereas previously there has been widespread application of very expensive new technologies across the spectrum of age and co-morbidities (often without proof of efficacy in groups excluded from randomized trials), it is not expected that in the future this approach will continue. Comparative efficacy studies of current (or polymer free) DES against bioabsorbable DES may demonstrate gains that are sufficiently small in absolute terms not to justify their widespread use in broad populations. It seems more likely that bioabsorbable DES may initially be used as a niche application, for example among patients who cannot take prolonged DAPT. As the elderly population of the western world grows however, and patients of increasing complexity are successfully treated with PCI, it is still reasonable to expect that niche markets of sufficient size will emerge to support the use and further development of exciting technologies such as this. Perhaps the biggest question remaining is whether the presence of a permanent scaffold beyond a several-month period after PCI is indeed superfluous (as proponents of bioabsorbable stents contend). At the present time, this is something that cannot be answered by the available data. Once the initial technical challenges have been overcome (in particular balancing the need for radial strength against the ability of the device to fully dissolve) these questions can be addressed by large scale, randomized, controlled trials.
  38 in total

1.  Biodegradable stents : "fulfilling the mission and stepping away".

Authors:  A Colombo; E Karvouni
Journal:  Circulation       Date:  2000-07-25       Impact factor: 29.690

2.  Long-term outcomes after stenting of bifurcation lesions with the "crush" technique: predictors of an adverse outcome.

Authors:  Angela Hoye; Ioannis Iakovou; Lei Ge; Carlos A G van Mieghem; Andrew T L Ong; John Cosgrave; Giuseppe M Sangiorgi; Flavio Airoldi; Matteo Montorfano; Iassen Michev; Alaide Chieffo; Mauro Carlino; Nicola Corvaja; Jiro Aoki; Gaston A Rodriguez Granillo; Marco Valgimigli; Georgios Sianos; Willem J van der Giessen; Pim J de Feyter; Ron T van Domburg; Patrick W Serruys; Antonio Colombo
Journal:  J Am Coll Cardiol       Date:  2006-04-24       Impact factor: 24.094

3.  Drug-eluting stents and late adverse clinical outcomes lessons learned, lessons awaited.

Authors:  Sotirios Tsimikas
Journal:  J Am Coll Cardiol       Date:  2006-04-04       Impact factor: 24.094

4.  Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk.

Authors:  Michael Joner; Aloke V Finn; Andrew Farb; Erik K Mont; Frank D Kolodgie; Elena Ladich; Robert Kutys; Kristi Skorija; Herman K Gold; Renu Virmani
Journal:  J Am Coll Cardiol       Date:  2006-05-05       Impact factor: 24.094

5.  Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans.

Authors:  H Tamai; K Igaki; E Kyo; K Kosuga; A Kawashima; S Matsui; H Komori; T Tsuji; S Motohara; H Uehata
Journal:  Circulation       Date:  2000-07-25       Impact factor: 29.690

Review 6.  Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings.

Authors:  Jun-ichi Kotani; Masaki Awata; Shinsuke Nanto; Masaaki Uematsu; Fusako Oshima; Hitoshi Minamiguchi; Gary S Mintz; Seiki Nagata
Journal:  J Am Coll Cardiol       Date:  2006-04-04       Impact factor: 24.094

7.  Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries.

Authors:  W J van der Giessen; A M Lincoff; R S Schwartz; H M van Beusekom; P W Serruys; D R Holmes; S G Ellis; E J Topol
Journal:  Circulation       Date:  1996-10-01       Impact factor: 29.690

8.  Dose-dependent smooth muscle cell proliferation induced by thermal injury with pulsed infrared lasers.

Authors:  P C Douek; R Correa; R Neville; E F Unger; M Shou; S Banai; V J Ferrans; S E Epstein; M B Leon; R F Bonner
Journal:  Circulation       Date:  1992-10       Impact factor: 29.690

9.  Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious?

Authors:  Renu Virmani; Giulio Guagliumi; Andrew Farb; Giuseppe Musumeci; Niccolo Grieco; Teresio Motta; Laurian Mihalcsik; Maurizio Tespili; Orazio Valsecchi; Frank D Kolodgie
Journal:  Circulation       Date:  2004-01-26       Impact factor: 29.690

10.  Lowering the dose of sirolimus, released from a nonpolymeric hydroxyapatite coated coronary stent, reduces signs of delayed healing.

Authors:  Wim J van der Giessen; Oana Sorop; Patrick W Serruys; Ilona Peters-Krabbendam; Heleen M M van Beusekom
Journal:  JACC Cardiovasc Interv       Date:  2009-04       Impact factor: 11.195

View more
  1 in total

1.  Biodegradable polymer versus second-generation durable polymer drug-eluting stents in patients with coronary artery disease: A meta-analysis.

Authors:  James J Wu; Joshua A H Way; Probal Roy; Andy Yong; Harry Lowe; Leonard Kritharides; David Brieger
Journal:  Health Sci Rep       Date:  2018-10-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.