Literature DB >> 10908211

Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans.

H Tamai1, K Igaki, E Kyo, K Kosuga, A Kawashima, S Matsui, H Komori, T Tsuji, S Motohara, H Uehata.   

Abstract

BACKGROUND: Although metallic stents are effective in preventing acute occlusion and reducing late restenosis after coronary angioplasty, many concerns still remain. Compared with metallic stents, poly-l-lactic acid (PLLA) stents are biodegradable and can deliver drugs locally. The aim of this study was to evaluate the feasibility, safety, and efficacy of the PLLA stent. METHODS AND
RESULTS: Fifteen patients electively underwent PLLA Igaki-Tamai stent implantation for coronary artery stenoses. The Igaki-Tamai stent is made of a PLLA monopolymer, has a thickness of 0.17 mm, and has a zigzag helical coil pattern. A balloon-expandable covered sheath system was used, and the stent expanded by itself to its original size with an adequate temperature. A total of 25 stents were successfully implanted in 19 lesions in 15 patients, and angiographic success was achieved in all procedures. No stent thrombosis and no major cardiac event occurred within 30 days. Coronary angiography and intravascular ultrasound were serially performed 1 day, 3 months, and 6 months after the procedure. Angiographically, both the restenosis rate and target lesion revascularization rate per lesion were 10.5%; the rates per patient were 6.7% at 6 months. Intravascular ultrasound findings revealed no significant stent recoil at 1 day, and they revealed stent expansion at follow-up. No major cardiac event, except for repeat angioplasty, developed within 6 months.
CONCLUSIONS: Our preliminary experience suggests that coronary PLLA biodegradable stents are feasible, safe, and effective in humans. Long-term follow-up with more patients will be required to validate the long-term efficacy of PLLA stents.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10908211     DOI: 10.1161/01.cir.102.4.399

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  95 in total

Review 1.  Surgical palliation in advanced disease: recent developments.

Authors:  Geoffrey P Dunn
Journal:  Curr Oncol Rep       Date:  2002-05       Impact factor: 5.075

2.  A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal-results 6-18 months after implantation into New Zealand white rabbits.

Authors:  M Peuster; P Wohlsein; M Brügmann; M Ehlerding; K Seidler; C Fink; H Brauer; A Fischer; G Hausdorf
Journal:  Heart       Date:  2001-11       Impact factor: 5.994

3.  Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology?

Authors:  B Heublein; R Rohde; V Kaese; M Niemeyer; W Hartung; A Haverich
Journal:  Heart       Date:  2003-06       Impact factor: 5.994

4.  Drug-eluting stents.

Authors:  Xiaodong Ma; Tim Wu; Michael P Robich; Xingwei Wang; Hao Wu; Bryan Buchholz; Stephen McCarthy
Journal:  Int J Clin Exp Med       Date:  2010-07-15

Review 5.  Emerging applications of nanotechnology for the diagnosis and management of vulnerable atherosclerotic plaques.

Authors:  Shann S Yu; Ryan A Ortega; Brendan W Reagan; John A McPherson; Hak-Joon Sung; Todd D Giorgio
Journal:  Wiley Interdiscip Rev Nanomed Nanobiotechnol       Date:  2011-08-10

6.  Tissue biocompatibility of new biodegradable drug-eluting stent materials.

Authors:  Ilkka Uurto; Andres Kotsar; Taina Isotalo; Joonas Mikkonen; Paula M Martikainen; Minna Kellomäki; Pertti Törmälä; Teuvo L J Tammela; Martti Talja; Juha-Pekka Salenius
Journal:  J Mater Sci Mater Med       Date:  2007-04-17       Impact factor: 3.896

7.  In vitro study of drug-eluting stent coatings based on poly(L-lactide) incorporating cyclosporine A - drug release, polymer degradation and mechanical integrity.

Authors:  Katrin Sternberg; Sven Kramer; Claudia Nischan; Niels Grabow; Thomas Langer; Gerhard Hennighausen; Klaus-Peter Schmitz
Journal:  J Mater Sci Mater Med       Date:  2007-03-27       Impact factor: 3.896

8.  Unconstrained recovery characterization of shape-memory polymer networks for cardiovascular applications.

Authors:  Christopher Michael Yakacki; Robin Shandas; Craig Lanning; Bryan Rech; Alex Eckstein; Ken Gall
Journal:  Biomaterials       Date:  2007-02-02       Impact factor: 12.479

9.  Enhanced endothelialization on surface modified poly(L-lactic acid) substrates.

Authors:  Hao Xu; Rajendrasing Deshmukh; Richard Timmons; Kytai Truong Nguyen
Journal:  Tissue Eng Part A       Date:  2010-12-18       Impact factor: 3.845

10.  Novel biodegradable stents for benign esophageal strictures following endoscopic submucosal dissection.

Authors:  Yasuharu Saito; Toyohiko Tanaka; Akira Andoh; Hideki Minematsu; Kazunori Hata; Tomoyuki Tsujikawa; Norihisa Nitta; Kiyoshi Murata; Yoshihide Fujiyama
Journal:  Dig Dis Sci       Date:  2007-08-23       Impact factor: 3.199

View more