Literature DB >> 22904606

The Latest Lessons Learned from Retrieval Analyses of Ultra-High Molecular Weight Polyethylene, Metal-on-Metal, and Alternative Bearing Total Disc Replacements.

Steven M Kurtz1, Jeffrey M Toth, Ryan Siskey, Lauren Ciccarelli, Dan Macdonald, Jorge Isaza, Todd Lanman, Ilona Punt, Marla Steinbeck, Jan Goffin, André van Ooij.   

Abstract

Knowledge regarding the in vivo performance and periposthetic tissue response of cervical and lumbar total disc replacements (TDRs) continues to expand. This review addresses the following four main questions: 1) What are the latest lessons learned from polyethylene in large joints and how are they relevant to current TDRs? 2) What are the latest lessons learned regarding adverse local tissue reactions from metal-on-metal, CoCr bearings in large joints and how are they relevant to current TDRs? 3) What advancements have been made in understanding the in vivo performance of alternative biomaterials, such as stainless steel and polycarbonate urethane, for TDRs in the past five years? 4) How has retrieval analysis of all these various artificial disc bearing technologies advanced the state of the art in preclinical testing of TDRs? The study of explanted artificial discs and their associated tissues can help inform bearing selection as well as the design of future generations of disc arthroplasty. Analyzing retrieved artificial discs is also essential for validating preclinical test methods.

Entities:  

Year:  2012        PMID: 22904606      PMCID: PMC3418706          DOI: 10.1053/j.semss.2011.11.011

Source DB:  PubMed          Journal:  Semin Spine Surg        ISSN: 1040-7383


  32 in total

1.  Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants.

Authors:  H J Wilke; K Wenger; L Claes
Journal:  Eur Spine J       Date:  1998       Impact factor: 3.134

2.  Polyethylene wear and rim fracture in total disc arthroplasty.

Authors:  Steven M Kurtz; André van Ooij; Raymond Ross; Jan de Waal Malefijt; John Peloza; Lauren Ciccarelli; Marta L Villarraga
Journal:  Spine J       Date:  2006-12-06       Impact factor: 4.166

3.  Longitudinal prospective long-term radiographic follow-up after treatment of single-level cervical disk disease with the Bryan Cervical Disc.

Authors:  Joris Walraevens; Philippe Demaerel; Paul Suetens; Frank Van Calenbergh; Johan van Loon; Jozef Vander Sloten; Jan Goffin
Journal:  Neurosurgery       Date:  2010-09       Impact factor: 4.654

4.  A clinical analysis of 4- and 6-year follow-up results after cervical disc replacement surgery using the Bryan Cervical Disc Prosthesis.

Authors:  Jan Goffin; Johan van Loon; Frank Van Calenbergh; Bailey Lipscomb
Journal:  J Neurosurg Spine       Date:  2010-03

Review 5.  History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.

Authors:  Steven M Kurtz; Heather A Gawel; Jasmine D Patel
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

6.  Risk factors for inflammatory pseudotumour formation following hip resurfacing.

Authors:  S Glyn-Jones; H Pandit; Y-M Kwon; H Doll; H S Gill; D W Murray
Journal:  J Bone Joint Surg Br       Date:  2009-12

7.  The epidemiology of bearing surface usage in total hip arthroplasty in the United States.

Authors:  Kevin J Bozic; Steven Kurtz; Edmund Lau; Kevin Ong; Vanessa Chiu; Thomas P Vail; Harry E Rubash; Daniel J Berry
Journal:  J Bone Joint Surg Am       Date:  2009-07       Impact factor: 5.284

8.  Metal-on-metal bearing surfaces.

Authors:  Joshua J Jacobs; Robert M Urban; Nadim J Hallab; Anastasia K Skipor; Alfons Fischer; Markus A Wimmer
Journal:  J Am Acad Orthop Surg       Date:  2009-02       Impact factor: 3.020

9.  The Bryan Cervical Disc: wear properties and early clinical results.

Authors:  Paul A Anderson; Rick C Sasso; Jeffrey P Rouleau; Cathy S Carlson; Jan Goffin
Journal:  Spine J       Date:  2004 Nov-Dec       Impact factor: 4.166

10.  Retrieval analysis of motion preserving spinal devices and periprosthetic tissues.

Authors:  Steven M Kurtz; Marla Steinbeck; Allyson Ianuzzi; André van Ooij; Ilona M Punt; Jorge Isaza; E R S Ross
Journal:  SAS J       Date:  2009-12-01
View more
  10 in total

1.  UHMWPE wear debris and tissue reactions are reduced for contemporary designs of lumbar total disc replacements.

Authors:  Sai Y Veruva; Todd H Lanman; Jorge E Isaza; Daniel W MacDonald; Steven M Kurtz; Marla J Steinbeck
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

Review 2.  Which design and biomaterial factors affect clinical wear performance of total disc replacements? A systematic review.

Authors:  Sai Y Veruva; Marla J Steinbeck; Jeffrey Toth; Dominik D Alexander; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 3.  Cervical disc replacement - emerging equivalency to anterior cervical discectomy and fusion.

Authors:  Aaron J Buckland; Joseph F Baker; Ryan P Roach; Jeffrey M Spivak
Journal:  Int Orthop       Date:  2016-04-08       Impact factor: 3.075

4.  A comparison of the inflammatory host response to particulate debris adjacent to unlocked and locked screws of a growth guidance system for early onset scoliosis.

Authors:  Jeffrey M Toth; Felix Ankomah; Noriaki Kawakami; Koki Uno
Journal:  Eur Spine J       Date:  2022-06-13       Impact factor: 2.721

5.  Cervical Artificial Disc Replacement Versus Fusion for Cervical Degenerative Disc Disease: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2019-02-19

6.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

7.  ISASS Policy Statement - Lumbar Artificial Disc.

Authors:  Jack Zigler; Rolando Garcia
Journal:  Int J Spine Surg       Date:  2015-03-12

Review 8.  Material Science in Cervical Total Disc Replacement.

Authors:  Martin H Pham; Vivek A Mehta; Alexander Tuchman; Patrick C Hsieh
Journal:  Biomed Res Int       Date:  2015-10-07       Impact factor: 3.411

9.  Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study.

Authors:  Ryan Siskey; Lauren Ciccarelli; Melissa K C Lui; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2016-11       Impact factor: 4.176

10.  Tuning C-C sp2/sp3 ratio of DLC films in FCVA system for biomedical application.

Authors:  Xi Rao; Jihan Yang; Zilin Chen; Yidie Yuan; Qiubing Chen; Xue Feng; Lizhao Qin; Yongping Zhang
Journal:  Bioact Mater       Date:  2020-02-20
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.