Literature DB >> 22772136

Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.

Jay Sengupta1, Arthur C Kendig, Marlene Goormastic, Eui-Seock Hwang, Elizabeth A Ching, Roy Chung, Bruce D Lindsay, Patrick J Tchou, Bruce L Wilkoff, Mark J Niebauer, David O Martin, Niraj Varma, Oussama Wazni, Walid Saliba, Mohamed Kanj, Mandeep Bhargava, Thomas Dresing, Tyler Taigen, Christopher Ingelmo, Mohamed Bassiouny, Edmond M Cronin, Bradley Wilsmore, John Rickard, Mina K Chung.   

Abstract

BACKGROUND: A significant proportion of implantable cardioverter-defibrillators (ICDs) have been subject to Food and Drug Administration (FDA) advisories. The impact of device advisories on mortality or patient care is poorly understood. Although estimated risks of ICD generators under advisory are low, dependency on ICD therapies to prevent sudden death justifies the assessment of long-term mortality.
OBJECTIVE: To test the association of FDA advisory status with long-term mortality.
METHODS: The study was a retrospective, single-center review of clinical outcomes, including device malfunctions, in patients from implantation to either explant or death. Patients with ICDs first implanted at Cleveland Clinic between August 1996 and May 2004 who became subject to FDA advisories on ICD generators were identified. Mortality was determined by using the Social Security Death Index.
RESULTS: In 1644 consecutive patients receiving first ICD implants, 704 (43%) became subject to an FDA advisory, of which 172 (10.5%) were class I and 532 (32.3%) were class II. ICDs were explanted before advisory notifications in 14.0% of class I and 10.1% of class II advisories. Among ICDs under advisory, 28 (4.0%) advisory-related and 15 non-advisory- related malfunctions were documented. Over a median follow-up of 70 months, 814 patients died. Kaplan-Meier 5-year survival rate was 65.6% overall, and 64.2, 61.1, and 69.3% in patients with no, class I, and class II advisories, respectively (P = .17).
CONCLUSIONS: ICD advisories impacted 43% of the patients. Advisory-related malfunctions affected 4% within the combined advisory group. Based on a conservative management strategy, ICDs under advisory were not associated with increased mortality over a background of significant disease-related mortality.
Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22772136      PMCID: PMC3459308          DOI: 10.1016/j.hrthm.2012.07.002

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  15 in total

1.  Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories.

Authors:  Paul A Gould; Andrew D Krahn
Journal:  JAMA       Date:  2006-04-26       Impact factor: 56.272

2.  Pacemaker and ICD generator reliability: meta-analysis of device registries.

Authors:  William H Maisel
Journal:  JAMA       Date:  2006-04-26       Impact factor: 56.272

3.  Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry.

Authors:  Jeanne E Poole; Marye J Gleva; Theofanie Mela; Mina K Chung; Daniel Z Uslan; Richard Borge; Venkateshwar Gottipaty; Timothy Shinn; Dan Dan; Leon A Feldman; Hanscy Seide; Stuart A Winston; John J Gallagher; Jonathan J Langberg; Kevin Mitchell; Richard Holcomb
Journal:  Circulation       Date:  2010-10-04       Impact factor: 29.690

4.  Pacemaker and ICD malfunction--an incomplete picture.

Authors:  Bruce L Wilkoff
Journal:  JAMA       Date:  2006-04-26       Impact factor: 56.272

5.  Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) and the International Coalition of Pacing and Electrophysiology Organizations (COPE).

Authors:  Mark D Carlson; Bruce L Wilkoff; William H Maisel; Mark D Carlson; Kenneth A Ellenbogen; Leslie A Saxon; Eric N Prystowsky; Joseph S Alpert; Michael E Cain; Elizabeth A Ching; Anne B Curtis; D Wyn Davies; Stephen C Hammill; Robert G Hauser; Rachel Lampert; Douglas P Zipes
Journal:  Heart Rhythm       Date:  2006-10       Impact factor: 6.343

6.  Multicenter experience with failed and recalled implantable cardioverter-defibrillator pulse generators.

Authors:  Robert G Hauser; David L Hayes; Andrew E Epstein; David S Cannom; Stephen C Vlay; Susan L Song; G Frank O Tyers
Journal:  Heart Rhythm       Date:  2006-02-28       Impact factor: 6.343

7.  Recalls and safety alerts involving pacemakers and implantable cardioverter-defibrillator generators.

Authors:  W H Maisel; M O Sweeney; W G Stevenson; K E Ellison; L M Epstein
Journal:  JAMA       Date:  2001-08-15       Impact factor: 56.272

8.  Heart disease and stroke statistics--2006 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee.

Authors:  Thomas Thom; Nancy Haase; Wayne Rosamond; Virginia J Howard; John Rumsfeld; Teri Manolio; Zhi-Jie Zheng; Katherine Flegal; Christopher O'Donnell; Steven Kittner; Donald Lloyd-Jones; David C Goff; Yuling Hong; Robert Adams; Gary Friday; Karen Furie; Philip Gorelick; Brett Kissela; John Marler; James Meigs; Veronique Roger; Stephen Sidney; Paul Sorlie; Julia Steinberger; Sylvia Wasserthiel-Smoller; Matthew Wilson; Philip Wolf
Journal:  Circulation       Date:  2006-01-11       Impact factor: 29.690

9.  Survival rates as a guide to implanted cardioverter-defibrillator replacement strategies for device recalls--adding statistical insight to clinical intuition.

Authors:  Lorne J Gula; David Massel; Andrew D Krahn; Allan C Skanes; Raymond Yee; George J Klein
Journal:  Am Heart J       Date:  2007-02       Impact factor: 4.749

10.  Death due to an implantable cardioverter defibrillator.

Authors:  Anne Messali; Olivier Thomas; Michel Chauvin; Philippe Coumel; Antoine Leenhardt
Journal:  J Cardiovasc Electrophysiol       Date:  2004-08
View more
  4 in total

1.  Use and Abuse of Internal Cardioverter Defibrillators for Primary Prevention.

Authors:  Joshua R Silverstein; Demosthenes G Katritsis; Mark E Josephson
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09

Review 2.  Adverse events following implantable cardioverter defibrillator implantation: a systematic review.

Authors:  Rebecca Persson; Amy Earley; Ann C Garlitski; Ethan M Balk; Katrin Uhlig
Journal:  J Interv Card Electrophysiol       Date:  2014-06-20       Impact factor: 1.900

3.  Perspectives in managing recalls of cardiac implantable electronic devices.

Authors:  Niraj Varma
Journal:  Indian Pacing Electrophysiol J       Date:  2016-12-14

4.  Health care policy and regulatory implications on medical device innovations: a cardiac rhythm medical device industry perspective.

Authors:  Arjun Sharma; Anthony Blank; Parashar Patel; Kenneth Stein
Journal:  J Interv Card Electrophysiol       Date:  2013-03-10       Impact factor: 1.900

  4 in total