Death due to an implantable cardioverter defibrillator.

Anne Messali1, Olivier Thomas, Michel Chauvin, Philippe Coumel, Antoine Leenhardt.   

Abstract

Inappropriate therapy due to noise oversensing caused a true ventricular fibrillation (VF) and death of a patient. A 49-year-old patient with a history of dilated cardiomyopathy received a double-chamber implantable cardioverter defibrillator (ICD) in 1991 for a sustained inducible ventricular tachycardia (VT). One appropriate shock delivered in 1994 terminated an episode of VT. The generator was replaced in 1995 and in 2000, and was connected to the initial leads. Three months after the second replacement, the patient received six consecutive shocks related to detection of noise interpreted as VF. Unfortunately, the sixth shock triggered a true VF, which was not treated due to end of the therapeutic sequence, and the patient died. The causes of the dysfunction are discussed.

Entities:  

Mesh:

Year:  2004        PMID: 15333095     DOI: 10.1046/j.1540-8167.2004.03611.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  4 in total

1.  Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.

Authors:  Jay Sengupta; Arthur C Kendig; Marlene Goormastic; Eui-Seock Hwang; Elizabeth A Ching; Roy Chung; Bruce D Lindsay; Patrick J Tchou; Bruce L Wilkoff; Mark J Niebauer; David O Martin; Niraj Varma; Oussama Wazni; Walid Saliba; Mohamed Kanj; Mandeep Bhargava; Thomas Dresing; Tyler Taigen; Christopher Ingelmo; Mohamed Bassiouny; Edmond M Cronin; Bradley Wilsmore; John Rickard; Mina K Chung
Journal:  Heart Rhythm       Date:  2012-07-03       Impact factor: 6.343

2.  Reduction of the inappropriate ICD therapies by implementing a new fuzzy logic-based diagnostic algorithm.

Authors:  Michał Lewandowski; Andrzej Przybylski; Wiesław Kuźmicz; Hanna Szwed
Journal:  Ann Noninvasive Electrocardiol       Date:  2013-09       Impact factor: 1.468

3.  Life-threatening and life-saving inappropriate implantable cardioverter defibrillator shocks.

Authors:  Jordi Pérez-Rodon; David Doiny; Berta Miranda; Nuria Rivas-Gandara; Ivo Roca-Luque; Jaume Francisco-Pascual; Rosa Maria Lidón; David García-Dorado; Angel Moya Mitjans
Journal:  Clin Case Rep       Date:  2017-03-08

Review 4.  ICD discrimination of SVT versus VT with 1:1 V-A conduction: A review of the literature.

Authors:  Rhanderson N Cardoso; Chris Healy; Juan Viles-Gonzalez; James O Coffey
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.