Literature DB >> 16639052

Pacemaker and ICD generator reliability: meta-analysis of device registries.

William H Maisel1.   

Abstract

CONTEXT: Despite there being millions of pacemaker and implantable cardioverter-defibrillator (ICD) generator implants worldwide, little is known about device reliability.
OBJECTIVES: To perform a meta-analysis of prospective pacemaker and ICD registries to determine annual rates of pacemaker and ICD malfunction and to identify trends in these rates. DATA SOURCES: MEDLINE (January 1966 to April 2005), the Cochrane Central Register of Controlled Trials (through second quarter 2005), the Cochrane Database of Systematic Reviews (through second quarter 2005), and a bibliographic review of secondary sources. Search terms included pacemaker, artificial; defibrillators, implantable; registries; performance; and malfunction. STUDY SELECTION: Eligible registries were prospective; reported the number of patients with pacemaker, ICD, or both; and allowed determination of the annual number of device malfunctions. Of 1007 references screened, 3 registries meeting the selection criteria were identified and included 2.1 million pacemaker person-years and 14,821 ICD person-years of observation. DATA EXTRACTION: Included the annual number of patients with pacemakers and ICDs at risk of device failure and the annual number of generator malfunctions (1983-2004 for pacemakers, 1988-2004 for ICDs). A device malfunction was defined as an integral component failure that required device explantation prior to reaching elective replacement. Failures of pacemaker and ICD electrodes were not included in the study. DATA SYNTHESIS: There were 2981 pacemaker and 384 ICD generator malfunctions. Pacemaker reliability improved markedly during the 1980s (P for trend <.001) and the pacemaker malfunction rate remained low during the remainder of the study. Implantable cardioverter-defibrillator reliability improved during the first 10 study years (P for trend <.001). From 1998-2002, however, the ICD malfunction rate increased more than 4-fold (P for trend <.001), before decreasing substantially in the latter 2 years of the study. Overall, the mean (SE) annual ICD malfunction rate was about 20-fold higher than the pacemaker malfunction rate (26.5 [3.8] vs 1.3 [0.1] malfunctions per 1000 person-years, P<.001). Battery malfunctions were the most common cause of device failure.
CONCLUSIONS: Pacemaker reliability has improved markedly. In contrast, after more than a decade of improving device reliability, the ICD malfunction rate transiently increased before experiencing substantial reductions in the latter 2 study years. Whether increasing device sophistication accounts for the observed decrease in reliability is not known. Continued monitoring of pacemaker and ICD performance is required.

Entities:  

Mesh:

Year:  2006        PMID: 16639052     DOI: 10.1001/jama.295.16.1929

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  28 in total

1.  ISHNE/EHRA expert consensus on remote monitoring of cardiovascular implantable electronic devices (CIEDs).

Authors:  Sergio Dubner; Angelo Auricchio; Jonathan S Steinberg; Panos Vardas; Peter Stone; Josep Brugada; Ryszard Piotrowicz; David L Hayes; Paulus Kirchhof; Günter Breithardt; Wojciech Zareba; Claudio Schuger; Mehmet K Aktas; Michal Chudzik; Suneet Mittal; Niraj Varma
Journal:  Ann Noninvasive Electrocardiol       Date:  2012-01       Impact factor: 1.468

Review 2.  Pacing technology: advances in pacing threshold management.

Authors:  Chu-pak Lau; Chung-wah Siu
Journal:  J Zhejiang Univ Sci B       Date:  2010-08       Impact factor: 3.066

3.  Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.

Authors:  Jay Sengupta; Arthur C Kendig; Marlene Goormastic; Eui-Seock Hwang; Elizabeth A Ching; Roy Chung; Bruce D Lindsay; Patrick J Tchou; Bruce L Wilkoff; Mark J Niebauer; David O Martin; Niraj Varma; Oussama Wazni; Walid Saliba; Mohamed Kanj; Mandeep Bhargava; Thomas Dresing; Tyler Taigen; Christopher Ingelmo; Mohamed Bassiouny; Edmond M Cronin; Bradley Wilsmore; John Rickard; Mina K Chung
Journal:  Heart Rhythm       Date:  2012-07-03       Impact factor: 6.343

4.  Seven times replacement of permanent cardiac pacemaker in 33 years to maintain adequate heart rate: a case report.

Authors:  Yinglu Hao; Yanping Li; Derong Liao; Ling Yang
Journal:  Ann Transl Med       Date:  2015-12

Review 5.  Modeling defibrillation of the heart: approaches and insights.

Authors:  Natalia Trayanova; Jason Constantino; Takashi Ashihara; Gernot Plank
Journal:  IEEE Rev Biomed Eng       Date:  2011

6.  Tunnel propagation of postshock activations as a hypothesis for fibrillation induction and isoelectric window.

Authors:  Takashi Ashihara; Jason Constantino; Natalia A Trayanova
Journal:  Circ Res       Date:  2008-01-24       Impact factor: 17.367

Review 7.  Radiology of cardiac devices and their complications.

Authors:  J Dipoce; A Bernheim; H Spindola-Franco
Journal:  Br J Radiol       Date:  2014-11-20       Impact factor: 3.039

Review 8.  [Fitness to drive in patients with cardiovascular implantable electronic devices].

Authors:  Dejan Mijic; Bernd Lemke; Harilaos Bogossian
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2019-05-09

9.  Focus on management of pacemaker and ICD advisories, recalls, and alerts.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Treat Options Cardiovasc Med       Date:  2006-09

Review 10.  Towards predictive modelling of the electrophysiology of the heart.

Authors:  Edward Vigmond; Fijoy Vadakkumpadan; Viatcheslav Gurev; Hermenegild Arevalo; Makarand Deo; Gernot Plank; Natalia Trayanova
Journal:  Exp Physiol       Date:  2009-03-06       Impact factor: 2.969

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.