Survival rates as a guide to implanted cardioverter-defibrillator replacement strategies for device recalls--adding statistical insight to clinical intuition.

Lorne J Gula1, David Massel, Andrew D Krahn, Allan C Skanes, Raymond Yee, George J Klein.   

Abstract

BACKGROUND: Recalls and advisories of implanted cardioverter-defibrillators (ICDs) have become an unfortunate reality of cardiac rhythm management. With a paucity of data available on which to base replacement decisions, our goal is to model the potential risks and benefits of ICD generator replacement. The estimated risks are varied through a wide range to determine the potential range of outcomes. METHODS AND
RESULTS: Using initial estimates of risk derived from real data on 2915 advisory devices from 17 implanting centers, a decision analysis and Markov model were used to estimate survival according to device replacement decision. Survival rates at 5 years with and without device replacement were estimated at 60.38% and 60.66%, respectively. This difference was not significantly different on comparative analysis, using variability determined by Monte Carlo simulation. One-way and two-way sensitivity analyses are presented, demonstrating the minimal effect of varying estimates of risk. Only variation in risk of device failure had a differential effect on survival, with a survival benefit at 7 years if annual risk of device failure is at least 1.8%. Little differential effect on survival was demonstrated by variation of estimates of arrhythmia risk, nonarrhythmic mortality, and postprocedure infection rate.
CONCLUSIONS: Survival rates with a generator replacement or nonreplacement strategy in response to ICD recalls are similar and decrease nearly in parallel over time. The main factor with differential effect on survival is risk of device failure, although the level of this risk required to confer a survival advantage to a replacement strategy is quite large.

Entities:  

Mesh:

Year:  2007        PMID: 17239686     DOI: 10.1016/j.ahj.2006.10.029

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  4 in total

1.  The effect of device advisories on implantable cardioverter-defibrillator therapy.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Cardiol Rep       Date:  2010-09       Impact factor: 2.931

2.  Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.

Authors:  Jay Sengupta; Arthur C Kendig; Marlene Goormastic; Eui-Seock Hwang; Elizabeth A Ching; Roy Chung; Bruce D Lindsay; Patrick J Tchou; Bruce L Wilkoff; Mark J Niebauer; David O Martin; Niraj Varma; Oussama Wazni; Walid Saliba; Mohamed Kanj; Mandeep Bhargava; Thomas Dresing; Tyler Taigen; Christopher Ingelmo; Mohamed Bassiouny; Edmond M Cronin; Bradley Wilsmore; John Rickard; Mina K Chung
Journal:  Heart Rhythm       Date:  2012-07-03       Impact factor: 6.343

3.  Formation of a national network for rapid response to device and lead advisories: The Canadian Heart Rhythm Society Device Advisory Committee.

Authors:  Andrew D Krahn; Christopher S Simpson; Ratika Parkash; Raymond Yee; Jean Champagne; Jeffrey S Healey; Doug Cameron; Bernard Thibault; Iqwal Mangat; Stanley Tung; Laurence Sterns; David H Birnie; Derek V Exner; Soori Sivakumaran; Ted Davies; Benoit Coutu; Eugene Crystal; Kevin Wolfe; Atul Verma; Elizabeth A Stephenson; Shubhayan Sanatani; Robert Gow; Sean Connors; Felix Ayala Paredes; Mike Turabian; Teresa Kus; Vidal Essebag; Martin Gardner
Journal:  Can J Cardiol       Date:  2009-07       Impact factor: 5.223

4.  Canadian Registry of Electronic Device Outcomes (CREDO): The Abbott ICD Premature Battery Depletion Advisory, a Multicentre Cohort Study.

Authors:  Jason Davis; Bernard Thibault; Iqwal Mangat; Benoit Coutu; Matthew Bennett; Francois Philippon; Roopinder Sandhu; Laurence Sterns; Vidal Essebag; Pablo Nery; George Wells; Raymond Yee; Derek Exner; Andrew Krahn; Ratika Parkash
Journal:  CJC Open       Date:  2020-09-12