Tighter accuracy standards within point-of-care blood glucose monitoring: how six commonly used systems compare.

Charlotte S Robinson1, Patrick Sharp.   

Abstract

INTRODUCTION: Blood glucose monitoring systems (BGMS) are used in the hospital environment to manage blood glucose levels in patients at the bedside. The International Organization for Standardization (ISO) 15197:2003 standard is currently used by regulatory bodies as a minimum requirement for the performance of BGMS, specific to self-testing. There are calls for the tightening of accuracy requirements and implementation of a standard specifically for point-of-care (POC) BGMS.
METHODS: The accuracy of six commonly used BGMS was assessed in a clinical setting, with 108 patients' finger stick capillary samples. Using the accuracy criteria from the existing standard and a range of tightened accuracy criteria, system performance was compared. Other contributors to system performance have been measured, including hematocrit sensitivity and meter error rates encountered in the clinical setting. RESULTS AND DISCUSSION: Five of the six BGMS evaluated met current accuracy criteria within the ISO 15197 standard. Only the Optium Xceed system had >95% of all readings within a tightened criteria of ±12.5% from the reference at glucose levels ≥72 mg/dl (4 mmol/liter) and ±9 mg/dl (0.5 mmol/liter) at glucose levels <72 mg/dl (4 mmol/liter). The Nova StatStrip Xpress had the greatest number of error messages observed; Optium Xceed the least. OneTouch Ultra2, Nova StatStrip Xpress, Accu-Chek Performa, and Contour TS products were all significantly influenced by blood hematocrit levels.
CONCLUSIONS: From evidence obtained during this clinical evaluation, the Optium Xceed system is most likely to meet future anticipated accuracy standards for POC BGMS. In this clinical study, the results demonstrated the Optium Xceed product to have the highest level of accuracy, to have the lowest occurrence of error messages, and to be least influenced by blood hematocrit levels.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22768885      PMCID: PMC3440041          DOI: 10.1177/193229681200600309

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  2 in total

1.  The food and drug administration is now preparing to establish tighter performance requirements for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

Review 2.  A review of standards and statistics used to describe blood glucose monitor performance.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2010-01-01
  2 in total
  11 in total

1.  Seven-Year Surveillance of the Clinical Performance of a Blood Glucose Test Strip Product.

Authors:  Steven Setford; Mike Grady; Stuart Phillips; Lesley Miller; Stephen Mackintosh; Hilary Cameron; Krisna Corrigall
Journal:  J Diabetes Sci Technol       Date:  2017-04-13

Review 2.  Challenges of inpatient blood glucose monitoring: standards, methods, and devices to measure blood glucose.

Authors:  Kathie L Hermayer; Aundrea S Loftley; Sumana Reddy; Satya Nandana Narla; Nina A Epps; Yusheng Zhu
Journal:  Curr Diab Rep       Date:  2015-03       Impact factor: 4.810

3.  Update on a Quality Initiative to Standardize Perioperative Care for Continuous Subcutaneous Insulin Infusion Therapy.

Authors:  Patricia A Mackey; Bithika M Thompson; Mary E Boyle; Heidi A Apsey; Karen M Seifert; Richard T Schlinkert; Joshua D Stearns; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2015-06-19

Review 4.  Existing and Emerging Technologies for Point-of-Care Testing.

Authors:  Andrew St John; Christopher P Price
Journal:  Clin Biochem Rev       Date:  2014-08

Review 5.  Perioperative Management of Patients with Diabetes and Hyperglycemia Undergoing Elective Surgery.

Authors:  Bithika M Thompson; Joshua D Stearns; Heidi A Apsey; Richard T Schlinkert; Curtiss B Cook
Journal:  Curr Diab Rep       Date:  2016-01       Impact factor: 4.810

Review 6.  Point-of-care blood glucose testing for diabetes care in hospitalized patients: an evidence-based review.

Authors:  Rajesh Rajendran; Gerry Rayman
Journal:  J Diabetes Sci Technol       Date:  2014-06-05

7.  Statistical transformation and the interpretation of inpatient glucose control data from the intensive care unit.

Authors:  George E Saulnier; Janna C Castro; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

8.  The Quantitative Relationship Between ISO 15197 Accuracy Criteria and Mean Absolute Relative Difference (MARD) in the Evaluation of Analytical Performance of Self-Monitoring of Blood Glucose (SMBG) Systems.

Authors:  Scott Pardo; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

9.  Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose.

Authors:  Hae-Il Park; Seong-Su Lee; Jang-Won Son; Hee-Sun Kwon; Sung Rae Kim; Hyojin Chae; Myungshin Kim; Yonggoo Kim; Soonjib Yoo
Journal:  J Clin Lab Anal       Date:  2016-04-13       Impact factor: 2.352

10.  Diabetes mellitus in Zambia and the Western Cape province of South Africa: Prevalence, risk factors, diagnosis and management.

Authors:  Sarah Lou Bailey; Helen Ayles; Nulda Beyers; Peter Godfrey-Faussett; Monde Muyoyeta; Elizabeth du Toit; John S Yudkin; Sian Floyd
Journal:  Diabetes Res Clin Pract       Date:  2016-05-07       Impact factor: 5.602

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.