Use of the iPhone for radiographic evaluation of hallux valgus.

Tolga Ege1, Ozkan Kose, Kenan Koca, Bahtiyar Demiralp, Mustafa Basbozkurt.   

Abstract

OBJECTIVE: The purpose of this study was to compare the measurements made using a smartphone accelerometer and computerized measurements as a reference in a series of 32 hallux valgus patients.
MATERIALS AND METHODS: Two observers used an iPhone to measure the hallux valgus angle (HVA), intermetatarsal angle (IMA), and distal metatarsal articular angle (of anteroposterior foot radiographs in 32 patients with symptomatic hallux valgus on a computer screen. Digital angular measurements on the computer were set as the reference standard for analysis and comparison. The difference between computerized measurements and all iPhone measurements, and the difference between the first and second iPhone measurements for each observer were calculated. Inter- and intraobserver reliability of the smartphone measurement method was also tested.
RESULTS: The variability of all measurements was similar for the iPhone and the computer-assisted techniques. The concordance between iPhone and computer-assisted angular measurements was excellent for the HVA, IMA, and DMAA. The maximum mean difference between the two techniques was 1.25 ± 1.02° for HVA, 0.92 ± 0.92° for IMA, and 1.10 ± 0.82° for DMAA. The interobserver reliability was excellent for HVA, IMA, and DMAA. The maximum mean difference between observers was 1.31 ± 0.89° for HVA, 0.90 ± 0.92° for IMA, and 0.78 ± 0.87° for DMAA. The intraobserver reliability was excellent for HVA, IMA, and DMAA.
CONCLUSIONS: We conclude that the Hallux Angles software for the iPhone can be used for measurement of hallux valgus angles in clinical practice and even for research purposes. It is an accurate and reproducible method.

Entities:  

Mesh:

Year:  2012        PMID: 22669733     DOI: 10.1007/s00256-012-1455-9

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  14 in total

1.  Surgery for hallux valgus. The expectations of patients and surgeons.

Authors:  W Schneider; K Knahr
Journal:  Int Orthop       Date:  2001       Impact factor: 3.075

2.  Use of the iPhone for Cobb angle measurement in scoliosis.

Authors:  Matthew Shaw; Clayton J Adam; Maree T Izatt; Paul Licina; Geoffrey N Askin
Journal:  Eur Spine J       Date:  2011-11-09       Impact factor: 3.134

Review 3.  Modern techniques in hallux abducto valgus surgery.

Authors:  Bradly W Bussewitz; Tim Levar; Christopher F Hyer
Journal:  Clin Podiatr Med Surg       Date:  2011-04       Impact factor: 1.231

Review 4.  The pathogenesis of hallux valgus.

Authors:  A M Perera; Lyndon Mason; M M Stephens
Journal:  J Bone Joint Surg Am       Date:  2011-09-07       Impact factor: 5.284

5.  Radiographic angles in hallux valgus: comparison between manual and computer-assisted measurements.

Authors:  Subodh Srivastava; Nachiappan Chockalingam; Tarek El Fakhri
Journal:  J Foot Ankle Surg       Date:  2010-09-15       Impact factor: 1.286

6.  Roger A. Mann Award . The reliability of angular measurements in hallux valgus deformities. .

Authors:  M J Coughlin; E Freund
Journal:  Foot Ankle Int       Date:  2001-05       Impact factor: 2.827

Review 7.  Hallux valgus: effectiveness and safety of minimally invasive surgery. A systematic review.

Authors:  Nicola Maffulli; Umile Giuseppe Longo; Andrea Marinozzi; Vincenzo Denaro
Journal:  Br Med Bull       Date:  2010-08-14       Impact factor: 4.291

8.  Radiographic angles in hallux valgus: differences between measurements made manually and with a computerized program.

Authors:  Carlos Piqué-Vidal; Ignaci Maled-García; Juanjo Arabi-Moreno; Joan Vila
Journal:  Foot Ankle Int       Date:  2006-03       Impact factor: 2.827

9.  Hallux valgus: demographics, etiology, and radiographic assessment.

Authors:  Michael J Coughlin; Caroll P Jones
Journal:  Foot Ankle Int       Date:  2007-07       Impact factor: 2.827

10.  Measuring hallux valgus: a comparison of conventional radiography and clinical parameters with regard to measurement accuracy.

Authors:  S Resch; L Ryd; A Stenström; K Johnsson; K Reynisson
Journal:  Foot Ankle Int       Date:  1995-05       Impact factor: 2.827

View more
  8 in total

1.  Sensitivity of fNIRS measurement to head motion: an applied use of smartphones in the lab.

Authors:  Xu Cui; Joseph M Baker; Ning Liu; Allan L Reiss
Journal:  J Neurosci Methods       Date:  2015-02-14       Impact factor: 2.390

2.  REPRODUCIBILITY OF SCHATZKER CLASSIFICATION THROUGH SMARTPHONE APPLICATIONS.

Authors:  Mauro Rodrigues Dos Santos; Junichiro Sado; Roberto Medeiros de Sousa; Osman Rodrigues Roriz
Journal:  Acta Ortop Bras       Date:  2016 Nov-Dec       Impact factor: 0.513

3.  Reliability and validity of angle measurements using radiograph and smartphone applications: experimental research on protractor.

Authors:  Takenori Awatani; Taisuke Enoki; Ikuhiro Morikita
Journal:  J Phys Ther Sci       Date:  2017-10-21

4.  Non-weightbearing compared with weightbearing x-rays in hallux valgus decision-making.

Authors:  Andrzej Boszczyk; Sławomir Kwapisz; Maciej Kiciński; Bartłomiej Kordasiewicz; Henryk Liszka
Journal:  Skeletal Radiol       Date:  2020-04-21       Impact factor: 2.199

5.  Reliability of different smartphones measuring the hallux valgus parameters in a new rapid method: a follow-up study.

Authors:  Lin Wang; Chao Zhang; Hao Liang; Jun Zhang; Weiyang Zhong; Zenghui Zhao; Tianji Huang; Xiaoji Luo
Journal:  BMC Musculoskelet Disord       Date:  2022-04-02       Impact factor: 2.362

6.  Development of a Manual Measurement Device for Measuring Hallux Valgus Angle in Patients with Hallux Valgus.

Authors:  Guoli Li; Jinsong Shen; Edward Smith; Chetna Patel
Journal:  Int J Environ Res Public Health       Date:  2022-07-26       Impact factor: 4.614

Review 7.  Telemedicine Workplace Environments: Designing for Success.

Authors:  Elizabeth A Krupinski
Journal:  Healthcare (Basel)       Date:  2014-02-24

Review 8.  Human Factors and Human-Computer Considerations in Teleradiology and Telepathology.

Authors:  Elizabeth A Krupinski
Journal:  Healthcare (Basel)       Date:  2014-02-19
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.