BACKGROUND AND PURPOSE: Selecting a lower tube current for CT fluoroscopic spine injections is a method of radiation dose reduction. Ideally tube current should be tailored to the patient's body habitus, but a greater influence on tube current may be the proceduralist's personal preference. The purpose of this study was to compare tube current and fluoroscopy time of different proceduralists for lumbar spine CT-guided selective nerve root blocks, and to correlate image quality to patient diameter and tube current. MATERIALS AND METHODS: Eighty CT-guided SNRBs performed by 4 proceduralists were retrospectively reviewed for tube current and fluoroscopy time. Patient body habitus was evaluated by measuring anteroposterior diameters on scout images. Image quality was evaluated objectively and subjectively: noise was measured in the psoas muscle and images were graded on a 3-point scale. RESULTS: The mean tube current was 59 ± 20 mA and mean fluoroscopy time was 10.4 ± 7.5 seconds. The mean tube current between proceduralists differed by almost 2-fold, and there was greater than 2-fold difference in mean fluoroscopy time (P < .0001 and .01, respectively). Mean AP body size was 27 ± 5 cm. When categoric data of tube current and AP diameter were analyzed, only AP diameter was a statistically significant variable influencing image noise (P = .009). Twenty of 23 patients with AP diameter ≤30 cm had adequate to excellent image quality, even with lower tube current of ≤40 mA. CONCLUSIONS: Wide variability in tube current selection between proceduralists calls for a more objective method of selecting tube current to minimize radiation dose. Body size, measured by AP diameter, had the greatest influence on image quality. This could be used to identify patients for lower tube current selection.
BACKGROUND AND PURPOSE: Selecting a lower tube current for CT fluoroscopic spine injections is a method of radiation dose reduction. Ideally tube current should be tailored to the patient's body habitus, but a greater influence on tube current may be the proceduralist's personal preference. The purpose of this study was to compare tube current and fluoroscopy time of different proceduralists for lumbar spine CT-guided selective nerve root blocks, and to correlate image quality to patient diameter and tube current. MATERIALS AND METHODS: Eighty CT-guided SNRBs performed by 4 proceduralists were retrospectively reviewed for tube current and fluoroscopy time. Patient body habitus was evaluated by measuring anteroposterior diameters on scout images. Image quality was evaluated objectively and subjectively: noise was measured in the psoas muscle and images were graded on a 3-point scale. RESULTS: The mean tube current was 59 ± 20 mA and mean fluoroscopy time was 10.4 ± 7.5 seconds. The mean tube current between proceduralists differed by almost 2-fold, and there was greater than 2-fold difference in mean fluoroscopy time (P < .0001 and .01, respectively). Mean AP body size was 27 ± 5 cm. When categoric data of tube current and AP diameter were analyzed, only AP diameter was a statistically significant variable influencing image noise (P = .009). Twenty of 23 patients with AP diameter ≤30 cm had adequate to excellent image quality, even with lower tube current of ≤40 mA. CONCLUSIONS: Wide variability in tube current selection between proceduralists calls for a more objective method of selecting tube current to minimize radiation dose. Body size, measured by AP diameter, had the greatest influence on image quality. This could be used to identify patients for lower tube current selection.
Authors: Jenny K Hoang; Andreia R Gafton; James D Eastwood; Luke F Chen; Lynne M Hurwitz Journal: AJR Am J Roentgenol Date: 2011-11 Impact factor: 3.959
Authors: L F Donnelly; K H Emery; A S Brody; T Laor; V M Gylys-Morin; C G Anton; S R Thomas; D P Frush Journal: AJR Am J Roentgenol Date: 2001-02 Impact factor: 3.959
Authors: Jenny K Hoang; Terry T Yoshizumi; Greta Toncheva; Linda Gray; Andreia R Gafton; Billy K Huh; James D Eastwood; Christopher D Lascola; Lynne M Hurwitz Journal: AJR Am J Roentgenol Date: 2011-10 Impact factor: 3.959
Authors: Xiang Li; Ehsan Samei; W Paul Segars; Gregory M Sturgeon; James G Colsher; Donald P Frush Journal: Med Phys Date: 2008-12 Impact factor: 4.071
Authors: Sarabjeet Singh; Mannudeep K Kalra; Michael A Moore; Randheer Shailam; Bob Liu; Thomas L Toth; Ellen Grant; Sjirk J Westra Journal: Radiology Date: 2009-05-12 Impact factor: 11.105
Authors: Ronald J Viola; Giao B Nguyen; Terry T Yoshizumi; Sandra S Stinnett; Jenny K Hoang; Peter G Kranz Journal: Interv Neuroradiol Date: 2014-10-17 Impact factor: 1.610
Authors: Kathrin Kizina; Benjamin Stolte; Andreas Totzeck; Saskia Bolz; Michael Fleischer; Christoph Mönninghoff; Nika Guberina; Denise Oldenburg; Michael Forsting; Christoph Kleinschnitz; Tim Hagenacker Journal: Front Neurol Date: 2019-11-05 Impact factor: 4.003