Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images.

Ruijiang Li1, Edward Mok, Bin Han, Albert Koong, Lei Xing.   

Abstract

PURPOSE: To evaluate the geometric accuracy of beam targeting in external surrogate-based gated volumetric modulated arc therapy (VMAT) using kilovoltage (kV) x-ray images acquired during dose delivery.
METHODS: Gated VMAT treatments were delivered using a Varian TrueBeam STx Linac for both physical phantoms and patients. Multiple gold fiducial markers were implanted near the target. The reference position was created for each implanted marker, representing its correct position at the gating threshold. The gating signal was generated from the RPM system. During the treatment, kV images were acquired immediately before MV beam-on at every breathing cycle, using the on-board imaging system. All implanted markers were detected and their 3D positions were estimated using in-house developed software. The positioning error of a marker is defined as the distance of the marker from its reference position for each frame of the images. The overall error of the system is defined as the average over all markers. For the phantom study, both sinusoidal motion (1D and 3D) and real human respiratory motion was simulated for the target and surrogate. In the baseline case, the two motions were synchronized for the first treatment fraction. To assess the effects of surrogate-target correlation on the geometric accuracy, a phase shift of 5% and 10% between the two motions was introduced. For the patient study, intrafraction kV images of five stereotactic body radiotherapy (SBRT) patients were acquired for one or two fractions.
RESULTS: For the phantom study, a high geometric accuracy was achieved in the baseline case (average error: 0.8 mm in the superior-inferior or SI direction). However, the treatment delivery is prone to geometric errors if changes in the target-surrogate relation occur during the treatment: the average error was increased to 2.3 and 4.7 mm for the phase shift of 5% and 10%, respectively. Results obtained with real human respiratory curves show a similar trend. For a target with 3D motion, the technique is able to detect geometric errors in the left-right (LR) and anterior-posterior (AP) directions. For the patient study, the average intrafraction positioning errors are 0.8, 0.9, and 1.4 mm and 95th percentile errors are 1.7, 2.1, and 2.7 mm in the LR, AP, and SI directions, respectively.
CONCLUSIONS: The correlation between external surrogate and internal target motion is crucial to ensure the geometric accuracy of surrogate-based gating. Real-time guidance based on kV x-ray images overcomes the potential issues in surrogate-based gating and can achieve accurate beam targeting in gated VMAT.

Entities:  

Mesh:

Year:  2012        PMID: 22559639      PMCID: PMC3344884          DOI: 10.1118/1.4704729

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Respiratory correlated cone beam CT.

Authors:  Jan-Jakob Sonke; Lambert Zijp; Peter Remeijer; Marcel van Herk
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

2.  Accuracy of daily image guidance for hypofractionated liver radiotherapy with active breathing control.

Authors:  Laura A Dawson; Cynthia Eccles; Jean-Pierre Bissonnette; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

Review 3.  Overview of image-guided radiation therapy.

Authors:  Lei Xing; Brian Thorndyke; Eduard Schreibmann; Yong Yang; Tian-Fang Li; Gwe-Ya Kim; Gary Luxton; Albert Koong
Journal:  Med Dosim       Date:  2006       Impact factor: 1.482

4.  Four-dimensional cone-beam computed tomography using an on-board imager.

Authors:  Tianfang Li; Lei Xing; Peter Munro; Christopher McGuinness; Ming Chao; Yong Yang; Bill Loo; Albert Koong
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

5.  A Bayesian approach to real-time 3D tumor localization via monoscopic x-ray imaging during treatment delivery.

Authors:  Ruijiang Li; Benjamin P Fahimian; Lei Xing
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

6.  Low-dose megavoltage cone-beam CT for radiation therapy.

Authors:  Jean Pouliot; Ali Bani-Hashemi; Josephine Chen; Michelle Svatos; Farhad Ghelmansarai; Matthias Mitschke; Michele Aubin; Ping Xia; Olivier Morin; Kara Bucci; Mack Roach; Paco Hernandez; Zirao Zheng; Dimitre Hristov; Lynn Verhey
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-02-01       Impact factor: 7.038

7.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy.

Authors:  H Shirato; S Shimizu; T Kunieda; K Kitamura; M van Herk; K Kagei; T Nishioka; S Hashimoto; K Fujita; H Aoyama; K Tsuchiya; K Kudo; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-11-01       Impact factor: 7.038

Review 8.  Stereotactic body radiation therapy in multiple organ sites.

Authors:  Robert D Timmerman; Brian D Kavanagh; L Chinsoo Cho; Lech Papiez; Lei Xing
Journal:  J Clin Oncol       Date:  2007-03-10       Impact factor: 44.544

9.  Clinical feasibility of using an EPID in CINE mode for image-guided verification of stereotactic body radiotherapy.

Authors:  Ross I Berbeco; Fred Hacker; Dan Ionascu; Harvey J Mamon
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-01       Impact factor: 7.038

10.  A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging.

Authors:  Weihua Mao; Nadeem Riaz; Louis Lee; Rodney Wiersma; Lei Xing
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

View more
  17 in total

1.  Evaluation of reproducibility of tumor repositioning during multiple breathing cycles for liver stereotactic body radiotherapy treatment.

Authors:  Ludovic Bedos; Olivier Riou; Norbert Aillères; Antoine Braccini; Jessica Molinier; Carmen Llacer Moscardo; David Azria; Pascal Fenoglietto
Journal:  Rep Pract Oncol Radiother       Date:  2016-11-12

2.  Evaluation of a template-based algorithm for markerless lung tumour localization on single- and dual-energy kilovoltage images.

Authors:  Alec M Block; Rakesh Patel; Murat Surucu; Matthew M Harkenrider; John C Roeske
Journal:  Br J Radiol       Date:  2016-10-12       Impact factor: 3.039

3.  First study of on-treatment volumetric imaging during respiratory gated VMAT.

Authors:  Kihwan Choi; Lei Xing; Albert Koong; Ruijiang Li
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

4.  Simultaneous modulated accelerated radiation therapy for esophageal cancer: a feasibility study.

Authors:  Wu-Zhe Zhang; Jian-Zhou Chen; De-Rui Li; Zhi-Jian Chen; Hong Guo; Ting-Ting Zhuang; Dong-Sheng Li; Ming-Zhen Zhou; Chuang-Zhen Chen
Journal:  World J Gastroenterol       Date:  2014-10-14       Impact factor: 5.742

5.  Data-Driven Respiratory Gating Outperforms Device-Based Gating for Clinical 18F-FDG PET/CT.

Authors:  Matthew D Walker; Andrew J Morgan; Kevin M Bradley; Daniel R McGowan
Journal:  J Nucl Med       Date:  2020-04-03       Impact factor: 10.057

6.  Making the right software choice for clinically used equipment in radiation oncology.

Authors:  Hilke Vorwerk; Klemens Zink; Daniela Michaela Wagner; Rita Engenhart-Cabillic
Journal:  Radiat Oncol       Date:  2014-06-23       Impact factor: 3.481

7.  Commissioning of a motion system to investigate dosimetric consequences due to variability of respiratory waveforms.

Authors:  Ashley J Cetnar; Joshua James; Brain Wang
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

8.  Dosimetric evaluation of four-dimensional dose distributions of CyberKnife and volumetric-modulated arc radiotherapy in stereotactic body lung radiotherapy.

Authors:  Mark K H Chan; Dora L W Kwong; Gilbert M L Law; Eric Tam; Anthony Tong; Venus Lee; Sherry C Y Ng
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

9.  Robust optimization of VMAT for lung cancer: Dosimetric implications of motion compensation techniques.

Authors:  Ben R Archibald-Heeren; Mikel V Byrne; Yunfei Hu; Meng Cai; Yang Wang
Journal:  J Appl Clin Med Phys       Date:  2017-08-08       Impact factor: 2.102

10.  Amplitude gating for a coached breathing approach in respiratory gated 10 MV flattening filter-free VMAT delivery.

Authors:  Francis Viel; Richard Lee; Ermias Gete; Cheryl Duzenli
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.