Design and development of a new micro-beam treatment planning system: effectiveness of algorithms of optimization and dose calculations and potential of micro-beam treatment.

Hidenobu Tachibana1, Hiroyuki Kojima, Noritaka Yusa, Satoshi Miyajima, Akihisa Tsuda, Takashi Yamashita.   

Abstract

A new treatment planning system (TPS) was designed and developed for a new treatment system, which consisted of a micro-beam-enabled linac with robotics and a real-time tracking system. We also evaluated the effectiveness of the implemented algorithms of optimization and dose calculations in the TPS for the new treatment system. In the TPS, the optimization procedure consisted of the pseudo Beam's-Eye-View method for finding the optimized beam directions and the steepest-descent method for determination of beam intensities. We used the superposition-/convolution-based (SC-based) algorithm and Monte Carlo-based (MC-based) algorithm to calculate dose distributions using CT image data sets. In the SC-based algorithm, dose density scaling was applied for the calculation of inhomogeneous corrections. The MC-based algorithm was implemented with Geant4 toolkit and a phase-based approach using a network-parallel computing. From the evaluation of the TPS, the system can optimize the direction and intensity of individual beams. The accuracy of the dose calculated by the SC-based algorithm was less than 1% on average with the calculation time of 15 s for one beam. However, the MC-based algorithm needed 72 min for one beam using the phase-based approach, even though the MC-based algorithm with the parallel computing could decrease multiple beam calculations and had 18.4 times faster calculation speed using the parallel computing. The SC-based algorithm could be practically acceptable for the dose calculation in terms of the accuracy and computation time. Additionally, we have found a dosimetric advantage of proton Bragg peak-like dose distribution in micro-beam treatment.

Entities:  

Mesh:

Year:  2012        PMID: 22544809     DOI: 10.1007/s12194-012-0153-6

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  10 in total

1.  Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions.

Authors:  W Schneider; T Bortfeld; W Schlegel
Journal:  Phys Med Biol       Date:  2000-02       Impact factor: 3.609

2.  Pseudo beam's-eye-view as applied to beam orientation selection in intensity-modulated radiation therapy.

Authors:  A Pugachev; L Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-12-01       Impact factor: 7.038

3.  Testing of the analytical anisotropic algorithm for photon dose calculation.

Authors:  Ann Van Esch; Laura Tillikainen; Jukka Pyykkonen; Mikko Tenhunen; Hannu Helminen; Sami Siljamäki; Jyrki Alakuijala; Marta Paiusco; Mauro Lori; Dominique P Huyskens
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

4.  A technique for generating phase-space-based Monte Carlo beamlets in radiotherapy applications.

Authors:  K Bush; I A Popescu; S Zavgorodni
Journal:  Phys Med Biol       Date:  2008-08-18       Impact factor: 3.609

5.  The importance of tissue segmentation for dose calculations for kilovoltage radiation therapy.

Authors:  Magdalena Bazalova; Edward E Graves
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

6.  Fast calculation of the exact radiological path for a three-dimensional CT array.

Authors:  R L Siddon
Journal:  Med Phys       Date:  1985 Mar-Apr       Impact factor: 4.071

7.  A convolution method of calculating dose for 15-MV x rays.

Authors:  T R Mackie; J W Scrimger; J J Battista
Journal:  Med Phys       Date:  1985 Mar-Apr       Impact factor: 4.071

8.  The accuracy of dose calculations by anisotropic analytical algorithms for stereotactic radiotherapy in nasopharyngeal carcinoma.

Authors:  M W K Kan; J Y C Cheung; L H T Leung; B M F Lau; P K N Yu
Journal:  Phys Med Biol       Date:  2010-12-22       Impact factor: 3.609

9.  Impact of inhomogeneity corrections on dose coverage in the treatment of lung cancer using stereotactic body radiation therapy.

Authors:  George X Ding; Dennis M Duggan; Bo Lu; Dennis E Hallahan; Anthony Cmelak; Arnold Malcolm; Jared Newton; Matthew Deeley; Charles W Coffey
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

10.  Clinical implications of the anisotropic analytical algorithm for IMRT treatment planning and verification.

Authors:  Christopher M Bragg; Katrina Wingate; John Conway
Journal:  Radiother Oncol       Date:  2008-02-04       Impact factor: 6.280

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.