High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads.

Sunthosh V Parvathaneni1, Christopher R Ellis, Jeffrey N Rottman.   

Abstract

BACKGROUND: Inside-out abrasion with externalization of sensing ring or high-voltage cables in St Jude Medical Riata implantable cardioverter-defibrillator leads has been reported. The prevalence of extruded cables, rate of electrical abnormalities, and predictors of failure in Riata leads are unknown.
OBJECTIVES: To estimate the incidence of lead failure in the St Jude Medical Riata implantable cardioverter-defibrillator leads and to propose a standard for the fluoroscopic assessment of insulation breakdown.
METHODS: Patients undergoing cine-fluoroscopy on Riata implantable cardioverter-defibrillator leads at our institution before January 25, 2012, were included (n = 87). Leads were graded as types 0-3 (0 = normal, 1 = abnormal conductor spacing, 2 ≤1 cm cable extrusion, 3 = >1 cm length extrusion). Comparison to extracted leads (n = 15) was documented. Device interrogation data were used for electrical analysis.
RESULTS: The mean time from implant was 5.9 ± 3.45 years. Structural lead failure with externalized cables was seen in 33.3% (29 of 87) of the patients. Thirty-one percent (9 of 29) of the leads with exposed cables showed electrical failure, and 29.7% (19 of 64) of the leads with normal electrical data contained externalized cables. Time from implant ≥5 years predicted structural lead failure (P < 0.05). X-ray grade compared with extracted leads demonstrated a sensitivity and specificity of 86% and 100%, respectively.
CONCLUSIONS: Cine-fluoroscopy using a simple scale correlated with the structural integrity of extracted Riata leads. A high percentage of leads with extrusion showed electrical failure. Leads ≥5 years from implant showed a high rate of externalized cables. A large independent multicenter study to determine the prevalence and clinical sequelae of Riata lead failures is warranted.
Copyright © 2012 Heart Rhythm Society. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22449742     DOI: 10.1016/j.hrthm.2012.03.049

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  14 in total

1.  Longitudinal follow-up of Riata leads reveals high annual incidence of new conductor externalization and electrical failure.

Authors:  Christian Steinberg; Jean-François Sarrazin; François Philippon; Jean Champagne; Marc-André Bouchard; Franck Molin; Isabelle Nault; Louis Blier; Gilles O'Hara
Journal:  J Interv Card Electrophysiol       Date:  2014-11-16       Impact factor: 1.900

2.  Riata lead up to date.

Authors:  Nobuhiro Nishii
Journal:  J Cardiol Cases       Date:  2017-02-01

3.  Cable externalization at the proximal portion of the superior vena cava coil in Riata implantable cardioverter defibrillator leads.

Authors:  Tomonori Katsuki; Hiroshi Furusho; Takashi Kusayama; Shinichiro Takashima; Takeshi Kato; Hisayoshi Murai; Soichiro Usui; Shuichi Kaneko; Masayuki Takamura
Journal:  J Cardiol Cases       Date:  2016-09-28

4.  Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening.

Authors:  Peem Lorvidhaya; Ivan Mendoza; Sharmila Sehli; Michael K Atalay; Michael H Kim
Journal:  J Interv Card Electrophysiol       Date:  2013-09-08       Impact factor: 1.900

5.  What did we learn from Riata™?

Authors:  Avi Fischer; C Jenney
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-09

6.  Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads.

Authors:  Liza P Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; John D Ferguson; J Michael Mangrum; Sandeep Kamath; Pamela K Mason
Journal:  J Interv Card Electrophysiol       Date:  2012-12-20       Impact factor: 1.900

7.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

8.  [Recommendations of the Working Group of Arrhythmias of the German Society of Cardiology on the approach to patients with Riata® and Riata ST® leads (St. Jude Medical). Nucleus of the Working Group of Arrhythmias of the German Society of Cardiology].

Authors:  C W Israel; D Bänsch; D Böcker; C Butter; J Chun; I Deisenhofer; L Eckardt; J-C Geller; T Hanke; T Klingenheben; C Piorkowski; B Schumacher
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06

9.  Extraction of defibrillator leads recalled for cable externalization and failure.

Authors:  Divyang Patel; Evan Adelstein; Jan Nemec; G Stuart Mendenhall; Raveen Bazaz; Sandeep Jain; Samir Saba
Journal:  J Interv Card Electrophysiol       Date:  2012-11-21       Impact factor: 1.900

10.  Secondary use of clinical data: the Vanderbilt approach.

Authors:  Ioana Danciu; James D Cowan; Melissa Basford; Xiaoming Wang; Alexander Saip; Susan Osgood; Jana Shirey-Rice; Jacqueline Kirby; Paul A Harris
Journal:  J Biomed Inform       Date:  2014-02-14       Impact factor: 6.317

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.