Literature DB >> 23229809

Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

S D A Valk1, D A M J Theuns, L Jordaens.   

Abstract

OBJECTIVE: Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580-1582 were studied, retrospectively, before the official recall.
METHODS: All 374 patients at the Erasmus Medical Center between July 2003 and December 2007 with a 1580, 1581 or 1582 shock lead.
RESULTS: The majority of the patients were male (78 %), with a median age of 60 years (IQR 52-70); primary prevention in 61 %. Median follow-up was 60.3 months (IQR 35.5-73.2), with 117 (31 %) patients dying. Electrical abnormalities (mainly noise, 65 %) were observed in 20/257 patients (7.8 %). Definite conductor externalisation was confirmed with fluoroscopy or chest X-ray in 16 patients, and in one after extraction. One patient presented with a drop in the high-voltage impedance trend with a short circuit of the ICD system during defibrillation testing, and needed to be shocked externally. In 8 more patients, conductor externalisation was found during an elective procedure. No predictors of externalisation could be found, except for the use of single coil (p = 0.02). Median time to conductor externalisation was 5 years (IQR 3.1-6.2). Lead externalisation was observed in 5.4 % (95 % CI 3.1-9.3) at 5 years and 22.7 % (95 % CI 13.6-36.6) at 8 years.
CONCLUSION: A high incidence of insulation defects associated with conductor externalisation in the Riata ICD lead family is observed. The mode of presentation is diverse. This type of insulation failure can lead to failure of therapy delivery.

Entities:  

Year:  2013        PMID: 23229809      PMCID: PMC3578533          DOI: 10.1007/s12471-012-0341-3

Source DB:  PubMed          Journal:  Neth Heart J        ISSN: 1568-5888            Impact factor:   2.380


  23 in total

1.  Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation.

Authors:  Martyn W Richards; Claire E Warren; Mark H Anderson
Journal:  Europace       Date:  2010-03-27       Impact factor: 5.214

2.  Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee.

Authors:  David H Birnie; Ratika Parkash; Derek V Exner; Vidal Essebag; Jeffrey S Healey; Atul Verma; Benoit Coutu; Teresa Kus; Iqwal Mangat; Felix Ayala-Paredes; Pablo Nery; George Wells; Andrew D Krahn
Journal:  Circulation       Date:  2012-02-06       Impact factor: 29.690

3.  ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons.

Authors:  Andrew E Epstein; John P DiMarco; Kenneth A Ellenbogen; N A Mark Estes; Roger A Freedman; Leonard S Gettes; A Marc Gillinov; Gabriel Gregoratos; Stephen C Hammill; David L Hayes; Mark A Hlatky; L Kristin Newby; Richard L Page; Mark H Schoenfeld; Michael J Silka; Lynne Warner Stevenson; Michael O Sweeney; Sidney C Smith; Alice K Jacobs; Cynthia D Adams; Jeffrey L Anderson; Christopher E Buller; Mark A Creager; Steven M Ettinger; David P Faxon; Jonathan L Halperin; Loren F Hiratzka; Sharon A Hunt; Harlan M Krumholz; Frederick G Kushner; Bruce W Lytle; Rick A Nishimura; Joseph P Ornato; Richard L Page; Barbara Riegel; Lynn G Tarkington; Clyde W Yancy
Journal:  Circulation       Date:  2008-05-15       Impact factor: 29.690

Review 4.  Premature failure of a Riata defibrillator lead without impedance change or inappropriate sensing: a case report and review of the literature.

Authors:  Angela Krebsbach; Fawaz Alhumaid; Charles A Henrikson; Hugh Calkins; Ronald D Berger; Alan Cheng
Journal:  J Cardiovasc Electrophysiol       Date:  2011-03-08

5.  Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead.

Authors:  Zaccaria Jalal; Nicolas Derval; Sylvain Ploux; Pierre Bordachar
Journal:  Heart Rhythm       Date:  2009-11-10       Impact factor: 6.343

6.  An ICD lead with failure of outer insulation goes undetected by regular measurements.

Authors:  Chi-Woh Chan; Chung-Seung Chiang
Journal:  Pacing Clin Electrophysiol       Date:  2011-07-11       Impact factor: 1.976

7.  Insulation defects of thin high-voltage ICD leads: an underestimated problem?

Authors:  Damir Erkapic; Gabor Z Duray; Tamas Bauernfeind; Salvatore De Rosa; Stefan H Hohnloser
Journal:  J Cardiovasc Electrophysiol       Date:  2011-04-01

Review 8.  ICD leads: design and chronic dysfunctions.

Authors:  Rainer Gradaus; Günter Breithardt; Dirk Böcker
Journal:  Pacing Clin Electrophysiol       Date:  2003-02       Impact factor: 1.976

9.  Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population.

Authors:  James G Porterfield; Linda M Porterfield; Karl H Kuck; Raffaele Corbisiero; Steven M Greenberg; Gerhard Hindricks; Oussama Wazni; Scott L Beau; John M Herre
Journal:  J Cardiovasc Electrophysiol       Date:  2009-11-17

10.  Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads.

Authors:  Robert G Hauser; Raed Abdelhadi; Deepa McGriff; Linda Kallinen Retel
Journal:  Heart Rhythm       Date:  2012-03-23       Impact factor: 6.343

View more
  5 in total

1.  Longitudinal follow-up of Riata leads reveals high annual incidence of new conductor externalization and electrical failure.

Authors:  Christian Steinberg; Jean-François Sarrazin; François Philippon; Jean Champagne; Marc-André Bouchard; Franck Molin; Isabelle Nault; Louis Blier; Gilles O'Hara
Journal:  J Interv Card Electrophysiol       Date:  2014-11-16       Impact factor: 1.900

Review 2.  [What must intensive care physicians know about implantable cardioverter defibrillator?].

Authors:  E Gatterer
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-09-20       Impact factor: 0.840

3.  Riata type implantable cardiac defibrillator leads: increasingly small, progressively troublesome.

Authors:  A Elvan; A Adiyaman
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

4.  Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening.

Authors:  Peem Lorvidhaya; Ivan Mendoza; Sharmila Sehli; Michael K Atalay; Michael H Kim
Journal:  J Interv Card Electrophysiol       Date:  2013-09-08       Impact factor: 1.900

5.  Riata silicone defibrillation lead with normal electrical measures at routine ambulatory check: The role of high-voltage shock testing.

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Cardiol       Date:  2016-11-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.