Literature DB >> 23179921

Extraction of defibrillator leads recalled for cable externalization and failure.

Divyang Patel1, Evan Adelstein, Jan Nemec, G Stuart Mendenhall, Raveen Bazaz, Sandeep Jain, Samir Saba.   

Abstract

BACKGROUND: The Riata implantable cardioverter-defibrillator lead was recalled by the Food and Drug Administration because of an increased rate of failure associated with cable externalization. Because of its mechanical separation, the Riata lead may be challenging to extract. We therefore examined the experience with Riata lead extractions at our institution.
METHODS: All patients implanted with the Riata lead underwent detailed review of their electronic medical records and operative notes. Procedural complications were ascertained by reviewing the medical records up to 30 days after Riata extraction.
RESULTS: From a total of 627 patients implanted with the Riata lead at our institution, 20 patients (age at time of extraction, 57 ± 11 years; 85 % men; lead dwell time, 48 ± 27 months) underwent lead extraction. Extraction was successful in 19 of 20 (95 %) patients and required the use of laser-powered sheaths in 18 (90 %) patients. Over a 30-day follow-up period, 1 of 20 (5 %) patients had a minor procedure-related complication consisting of a new pericardial effusion that did not require drainage.
CONCLUSIONS: Extraction of the Riata lead seems to be successful and safe and frequently requires the use of powered sheaths.

Entities:  

Mesh:

Year:  2012        PMID: 23179921     DOI: 10.1007/s10840-012-9751-9

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  17 in total

Review 1.  Evolution of the implantable cardioverter defibrillator.

Authors:  David S Cannom; Eric N Prystowsky
Journal:  J Cardiovasc Electrophysiol       Date:  2004-03

2.  Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA).

Authors:  Bruce L Wilkoff; Charles J Love; Charles L Byrd; Maria Grazia Bongiorni; Roger G Carrillo; George H Crossley; Laurence M Epstein; Richard A Friedman; Charles E H Kennergren; Przemyslaw Mitkowski; Raymond H M Schaerf; Oussama M Wazni
Journal:  Heart Rhythm       Date:  2009-05-22       Impact factor: 6.343

3.  Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead.

Authors:  Zaccaria Jalal; Nicolas Derval; Sylvain Ploux; Pierre Bordachar
Journal:  Heart Rhythm       Date:  2009-11-10       Impact factor: 6.343

4.  Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter-defibrillator lead.

Authors:  Melanie Maytin; Charles J Love; Avi Fischer; Roger G Carrillo; Juan D Garisto; Maria Grazia Bongiorni; Luca Segreti; Roy M John; Gregory F Michaud; Christine M Albert; Laurence M Epstein
Journal:  J Am Coll Cardiol       Date:  2010-07-02       Impact factor: 24.094

5.  Differences in encapsulating lead tissue in patients who underwent transvenous lead removal.

Authors:  Agnieszka Kołodzińska; Andrzej Kutarski; Łukasz Koperski; Marcin Grabowski; Barbara Małecka; Grzegorz Opolski
Journal:  Europace       Date:  2012-01-13       Impact factor: 5.214

6.  Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead.

Authors:  Robert G Hauser; Linda M Kallinen; Adrian K Almquist; Charles C Gornick; William T Katsiyiannis
Journal:  Heart Rhythm       Date:  2007-04-10       Impact factor: 6.343

7.  Laser-assisted extraction of pacemaker and defibrillator leads: the role of the cardiac surgeon.

Authors:  Jeffrey G Gaca; Brian Lima; Carmelo A Milano; Shu S Lin; R Duane Davis; James E Lowe; Peter K Smith
Journal:  Ann Thorac Surg       Date:  2009-05       Impact factor: 4.330

8.  Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads.

Authors:  A E Epstein; G N Kay; V J Plumb; S M Dailey; P G Anderson
Journal:  Circulation       Date:  1998-10-13       Impact factor: 29.690

9.  Large, single-center, single-operator experience with transvenous lead extraction: outcomes and changing indications.

Authors:  Samuel O Jones; Robert E Eckart; Christine M Albert; Laurence M Epstein
Journal:  Heart Rhythm       Date:  2008-01-17       Impact factor: 6.343

10.  Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads.

Authors:  Robert G Hauser; Raed Abdelhadi; Deepa McGriff; Linda Kallinen Retel
Journal:  Heart Rhythm       Date:  2012-03-23       Impact factor: 6.343

View more
  1 in total

1.  Managing patients with advisory defibrillator leads: what can we learn from published data?

Authors:  F A Bracke; B M van Gelder
Journal:  Neth Heart J       Date:  2015-04       Impact factor: 2.380