Literature DB >> 22358174

Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study.

Marcus Christian Müller1, Peter Belei, Peter H Pennekamp, Koroush Kabir, Dieter C Wirtz, Christof Burger, Oliver Weber.   

Abstract

PURPOSE: Medial femoral neck fractures are common, and closed reduction and internal fixation by three cannulated screws is an accepted method for the surgical treatment. Computer navigation for screw placement may reduce fluoroscopy time, the number of guidewire passes and optimise screw placement.
METHODS: In the context of a sawbone study, a computer-assisted planning and navigation system based on 3D-imaging for guidewire placement in the femoral neck was tested to improve screw placement. Three screws were inserted into 12, intact, femoral sawbones using the conventional technique and into 12, intact, femoral sawbones guided by the computer-based navigation system. Guidewire and subsequent screw placement in the femoral neck were evaluated.
RESULTS: Use of the navigation system resulted in a significant reduction of the number of drilling attempts (p≤0.05) and achieved optimised accuracy of implant placement by attaining significantly better screw parallelism (p≤0.05) and significantly enlarged neck-width coverage by the three screws (p≤0.0001). Computer assistance significantly increased the number of fluoroscopic images (p≤0.001) and the operation time (p≤0.0001).
CONCLUSIONS: Three-dimensional computer-assisted navigation improves accuracy of cannulated screw placement in femoral neck while increasing the number of fluoroscopic images and operation time. Additional studies including fractured sawbones and cadaver models with the goal of reducing operation time are indispensable before introduction of this navigation system into clinical practice.

Entities:  

Mesh:

Year:  2012        PMID: 22358174      PMCID: PMC3385907          DOI: 10.1007/s00264-012-1496-7

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  21 in total

1.  [Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-C-3 D)].

Authors:  C Rock; D Kotsianos; U Linsenmaier; T Fischer; R Brandl; F Vill; S Wirth; R Kaltschmidt; E Euler; K J Pfeifer; M Reiser
Journal:  Rofo       Date:  2002-02

2.  Intracapsular femoral neck fractures: the importance of early reduction and stable osteosynthesis.

Authors:  J Szita; P Cserháti; U Bosch; J Manninger; T Bodzay; K Fekete
Journal:  Injury       Date:  2002-12       Impact factor: 2.586

3.  [Iso-C(3D0-assisted) navigated implantation of pedicle screws in thoracic lumbar vertebrae].

Authors:  K Wendl; J von Recum; A Wentzensen; P A Grützner
Journal:  Unfallchirurg       Date:  2003-11       Impact factor: 1.000

Review 4.  Image guidance in pelvic and acetabular surgery--expectations, success and limitations.

Authors:  Ulrich Stöckle; Klaus Schaser; Benjamin König
Journal:  Injury       Date:  2007-04-02       Impact factor: 2.586

5.  Subtrochanteric fracture after cannulated screw fixation of femoral neck fractures: a report of four cases.

Authors:  Peter Kloen; Iván F Rubel; John P Lyden; David L Helfet
Journal:  J Orthop Trauma       Date:  2003-03       Impact factor: 2.512

6.  Evaluation of a fluoroscopy-based navigation system enabling a virtual radiation-free preview of X-ray images for placement of cannulated hip screws. A cadaver study.

Authors:  M C Müller; P Belei; M De La Fuente; M Strake; O Weber; C Burger; K Radermacher; D C Wirtz
Journal:  Comput Aided Surg       Date:  2011

7.  [3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography].

Authors:  D Kotsianos; C Rock; E Euler; S Wirth; U Linsenmaier; R Brandl; W Mutschler; K J Pfeifer
Journal:  Unfallchirurg       Date:  2001-09       Impact factor: 1.000

8.  [A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation].

Authors:  D Kendoff; T Hüfner; M Citak; C Maier; F Wesemeier; A Pearle; C Krettek
Journal:  Unfallchirurg       Date:  2006-10       Impact factor: 1.000

9.  Experimental validation of noninvasive referencing in navigated procedures on long bones.

Authors:  Daniel Kendoff; Aleksander Bogojević; Musa Citak; Mustafa Citak; Christian Maier; Georg Maier; Christian Krettek; Tobias Hüfner
Journal:  J Orthop Res       Date:  2007-02       Impact factor: 3.494

10.  Three-dimensional fluoroscopy for evaluation of articular reduction and screw placement in calcaneal fractures.

Authors:  Daniel Kendoff; Musa Citak; Michael Gardner; Maurico Kfuri; Barbara Thumes; Christian Krettek; Tobias Hüfner
Journal:  Foot Ankle Int       Date:  2007-11       Impact factor: 2.827

View more
  6 in total

1.  Computer-Assisted Orthopedic and Trauma Surgery.

Authors:  Timo Stübig; Henning Windhagen; Christian Krettek; Max Ettinger
Journal:  Dtsch Arztebl Int       Date:  2020-11-20       Impact factor: 5.594

2.  TiRobot-Assisted Percutaneous Cannulated Screw Fixation in the Treatment of Femoral Neck Fractures: A Minimum 2-Year Follow-up of 50 Patients.

Authors:  Zong-Dong Zhu; Cheng-Wei Xiao; Bo Tan; Xiao-Ming Tang; Dan Wei; Jia-Bin Yuan; Jiang Hu; Liao Feng
Journal:  Orthop Surg       Date:  2021-01-15       Impact factor: 2.071

Review 3.  Computer-assisted cannulated screw internal fixation versus conventional cannulated screw internal fixation for femoral neck fractures: a systematic review and meta-analysis.

Authors:  Qing-Hao Cheng; Peng-Biao Li; Ting-Ting Lu; Shi-Fang Guo; Wen-Fei Di; Ke-Hu Yang; Yao-Wen Qian
Journal:  J Orthop Surg Res       Date:  2021-11-22       Impact factor: 2.359

4.  A Novel Guidewire Aiming Device to Improve the Accuracy of Guidewire Insertion in Femoral Neck Fracture Surgery Using Cannulated Screw Fixation.

Authors:  Wenjing Yin; Haitao Xu; Peijun Xu; Tu Hu; Zhiquan An; Changqing Zhang; Jiagen Sheng
Journal:  Med Sci Monit       Date:  2016-08-16

5.  Introduction of a guide based on a femoral neck section for fixation with multiple screws: a cadaveric study.

Authors:  Qiuliang Zhu; Bin Xu; Jinzhu Lv; Maohua Yan
Journal:  BMC Musculoskelet Disord       Date:  2018-04-04       Impact factor: 2.362

6.  Three-dimensional morphological analysis of the femoral neck torsion angle-an anatomical study.

Authors:  Ru-Yi Zhang; Xiu-Yun Su; Jing-Xin Zhao; Jian-Tao Li; Li-Cheng Zhang; Pei-Fu Tang
Journal:  J Orthop Surg Res       Date:  2020-05-27       Impact factor: 2.359

  6 in total