Literature DB >> 17004045

[A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation].

D Kendoff1, T Hüfner, M Citak, C Maier, F Wesemeier, A Pearle, C Krettek.   

Abstract

BACKGROUND: Trauma navigation applications employ conventional mechanical surgical instruments for the simulation of drilling trajectories. Few complex mechanical targeting instruments, such as guides with fixed angles or with multiple parallel cannulas, have been adapted into trauma navigation systems. We have integrated a complex mechanical tool, the parallel drill guide (PDG) for the minimally invasive treatment of femoral neck fractures, into a trauma navigation module. The combined advantage of a complex yet commonly used mechanical tool with the benefits of fluoroscopic navigation was evaluated.
MATERIAL AND METHODS: To adapt the conventional PDG to a fluoroscopic navigation system, the instrument was fitted with a non-detachable reflective marker array. Navigation engineers developed custom software to enable visualization of the navigated PDG. A comparison of conventional versus navigated PDG techniques was performed on plastic bone models and cadavers. No software or mechanical failures occurred with the navigated PDG procedures.
RESULTS: While the total operative time was 30% more with navigation compared with conventional techniques, the total radiation time for the navigated group was reduced by more than 60%. This study demonstrates the successful integration of a cannulated parallel drill guide with a fluoroscopic navigation system.
CONCLUSION: The continuous display of the complex PDG mechanical instrument, with multiple parallel virtual trajectories, enables safe and accurate parallel screw placement. The integration of complex mechanical instrumentation with navigation for the accurate placement of hardware represents an attractive direction in multiple trauma applications.

Entities:  

Mesh:

Year:  2006        PMID: 17004045     DOI: 10.1007/s00113-006-1142-1

Source DB:  PubMed          Journal:  Unfallchirurg        ISSN: 0177-5537            Impact factor:   1.000


  18 in total

1.  [Precision in standardized Iso-C-Arm based navigated boring of the proximal femur].

Authors:  M Arand; M Schempf; L Kinzl; T Fleiter; D Pless; F Gebhard
Journal:  Unfallchirurg       Date:  2001-12       Impact factor: 1.000

2.  Intracapsular femoral neck fractures: the importance of early reduction and stable osteosynthesis.

Authors:  J Szita; P Cserháti; U Bosch; J Manninger; T Bodzay; K Fekete
Journal:  Injury       Date:  2002-12       Impact factor: 2.586

3.  Treatment of femoral neck fractures in Hungary with the Manninger screw.

Authors:  K Fekete; T Laczkó; I Flóris; P Cserháti; L Tasnádi
Journal:  Injury       Date:  2002-12       Impact factor: 2.586

Review 4.  Computer-assisted orthopaedic surgery: do we need CAOS?

Authors:  J M Sikorski; S Chauhan
Journal:  J Bone Joint Surg Br       Date:  2003-04

5.  [CT and fluoroscopy based navigation in pelvic surgery].

Authors:  U Stöckle; B König; K Schaser; I Melcher; N P Haas
Journal:  Unfallchirurg       Date:  2003-11       Impact factor: 1.000

6.  Significance of urgent (within 6h) internal fixation in the management of fractures of the neck of the femur.

Authors:  J Manninger; G Kazar; G Fekete; K Fekete; S Frenyo; F Gyarfas; T Salacz; A Varga
Journal:  Injury       Date:  1989-03       Impact factor: 2.586

7.  Standard multiplanar fluoroscopy versus a fluoroscopically based navigation system for the percutaneous insertion of iliosacral screws: a cadaver model.

Authors:  Cory Collinge; David Coons; Paul Tornetta; John Aschenbrenner
Journal:  J Orthop Trauma       Date:  2005-04       Impact factor: 2.512

8.  Accuracy study of computer-assisted drilling: the effect of bone density, drill bit characteristics, and use of a mechanical guide.

Authors:  T Hüfner; J Geerling; G Oldag; M Richter; M Kfuri; T Pohlemann; C Krettek
Journal:  J Orthop Trauma       Date:  2005 May-Jun       Impact factor: 2.512

9.  [Femoral neck fracture in the elderly--differential procedure].

Authors:  A Ekkernkamp; P A Ostermann; G Muhr
Journal:  Zentralbl Chir       Date:  1995       Impact factor: 0.942

10.  [Osteosynthesis of femoral neck fractures by double-cannulated screws (preliminary report)].

Authors:  K Fekete; T Salacz; J Manninger; G Kazár; T Laczkó; P Cserháti; P Füles; Z Vendégh; L Losonczy
Journal:  Magy Traumatol Orthop Helyreallito Seb       Date:  1992
View more
  7 in total

1.  Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study.

Authors:  Marcus Christian Müller; Peter Belei; Peter H Pennekamp; Koroush Kabir; Dieter C Wirtz; Christof Burger; Oliver Weber
Journal:  Int Orthop       Date:  2012-02-23       Impact factor: 3.075

2.  Computer-Assisted Orthopedic and Trauma Surgery.

Authors:  Timo Stübig; Henning Windhagen; Christian Krettek; Max Ettinger
Journal:  Dtsch Arztebl Int       Date:  2020-11-20       Impact factor: 5.594

3.  [Drilling with 3D fluoroscopic navigation in osteonecrosis of the femoral condyle].

Authors:  M Citak; D Kendoff; T Stübig; C Krettek; T Hüfner
Journal:  Unfallchirurg       Date:  2008-05       Impact factor: 1.000

4.  [Precision of guidewire placement--can it be improved by applying the new, isocentric aiming principle?].

Authors:  N Suhm; P Toggwiler; M Hänni; V Quarz; A Appelt
Journal:  Unfallchirurg       Date:  2007-05       Impact factor: 1.000

5.  [Evaluation of a 2D fluoroscopy-based navigation system for insertion of femoral neck screws. An experimental study].

Authors:  M C Müller; P Belei; M de la Fuente; M Strake; K Kabir; O Weber; C Burger; K Radermacher; D C Wirtz
Journal:  Unfallchirurg       Date:  2012-09       Impact factor: 1.000

6.  A new adjustable parallel drill guide for internal fixation of femoral neck fracture: a developmental and experimental study.

Authors:  Varah Yuenyongviwat; Pakjai Tuntarattanapong; Boonsin Tangtrakulwanich
Journal:  BMC Musculoskelet Disord       Date:  2016-01-11       Impact factor: 2.362

7.  2D-fluoroscopic based navigation for Gamma 3 nail insertion versus conventional procedure- a feasibility study.

Authors:  A Wilharm; I Marintschev; G O Hofmann; F Gras
Journal:  BMC Musculoskelet Disord       Date:  2013-02-28       Impact factor: 2.362

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.