| Literature DB >> 29618344 |
Qiuliang Zhu1, Bin Xu2, Jinzhu Lv2, Maohua Yan2.
Abstract
BACKGROUND: The design and application of assisted instrumentation for internal fixation of femoral neck fractures with multiple screws are still evolving. A novel guide based on a femoral neck section was designed to improve the accuracy of screw placement, and its efficacy was evaluated.Entities:
Keywords: Femoral neck fracture; Femoral neck section; Guide; Internal fixation; Multiple screws
Mesh:
Year: 2018 PMID: 29618344 PMCID: PMC5885363 DOI: 10.1186/s12891-018-2026-6
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Picture showed that the shape of femoral neck section (FNS) related with the medial aspect (A) and lateral surface (B) of proximal femur: The ellipse represented FNS; ∠α Femoral neck torsion angle; b Long axis of FNS; c Coronal axis of proximal femur
Fig. 2Overall view (A) and medial surface (B) of the guide of femoral neck section (GFNS) and the conventional guide (C) supplied by manufacturer for internal fixation of femoral neck fracture: a angle logo; b mounting hole; c parallel guide hole; d centre hole; e adjust hole
The general data of 20 femurs
| Group | Number | Femoral neck section (mm) | Femoral neck torsion angle (deg.) | |
|---|---|---|---|---|
| Long axis | Short axis | |||
| DFNS | 10 | 36.67 ± 3.62* | 25.57 ± 2.44* | 19.30 ± 3.01* |
| Conventional | 10 | 37.08 ± 2.74 | 24.98 ± 2.95 | 20.12 ± 2.36 |
*P > 0.05 Comparing with Conventional Group
Fig. 3The operative process with GFNS: Inserted the first guide-wire into appropriate place with freehand under AP and lateral views (A). Fixed and draw FNS on GFNS (red ellipse) by related data, identify the excellent hole (red point), good hole (blue point) and bad hole (white hole). After making sure the guide wires were in ideal positions by lateral and AP image, fixed with appropriate length screws
Comparative analysis of screw placement accuracy in 2 groups
| group | Screw parallelism (deg.) | Screw area ratio | DS (mm) | DC (mm) | DSC (mm) | |
|---|---|---|---|---|---|---|
| AP | Lateral | |||||
| conventional | 1.73 ± 0.62 | 1.52 ± 0.60 | 0.16 ± 0.02 | 10.38 ± 3.91 | 8.02 ± 1.51 | 4.21 ± 1.31 |
| GFNS | 1.68 ± 0.83# | 1.61 ± 0.57# | 0.29 ± 0.02* | 16.87 ± 4.12* | 11.05 ± 1.94* | 2.40 ± 1.24* |
GFNS guide of femoral neck section, DS distance between screws, DC distance between the centre of FNS, DSC distance between screw to femoral neck cortex
*P < 0.05; #p > 0.05 Comparing with Conventional group
comparative analysis of parameters during surgical procedure in 2 groups
| Group | Operation time (min) | Fluoroscopic time (no.) | Drilling attempt (no.) |
|---|---|---|---|
| conventional | 24.80 ± 9.12 | 8.89 ± 2.17 | 7.97 ± 3.60 |
| GFNS | 27.33 ± 7.61# | 9.51 ± 1.72# | 4.81 ± 1.27* |
*P < 0.05; #p > 0.05 Comparing with Conventional group