Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee.

David H Birnie1, Ratika Parkash, Derek V Exner, Vidal Essebag, Jeffrey S Healey, Atul Verma, Benoit Coutu, Teresa Kus, Iqwal Mangat, Felix Ayala-Paredes, Pablo Nery, George Wells, Andrew D Krahn.   

Abstract

BACKGROUND: Approximately 268,000 Fidelis leads were implanted worldwide until distribution was suspended because of a high rate of early failure. Careful analyses of predictors of increased lead failure hazard are required to help direct future lead design and also to inform decision making on lead replacement. We sought to perform a comprehensive analysis of all potential predictors in a multicenter study. METHODS AND
RESULTS: A total of 3169 Sprint Fidelis leads were implanted in 11 centers with a total of 251 failures. Lead failure rates at 3, 4, and 5 years were 5.3%, 10.6%, and 16.8%, respectively. The rate of lead failure continues to accelerate (P<0.001). There were 4 independent predictors of failure: center, sex, access vein, and previous lead failure. Women had a higher hazard of failure (hazard ratio 1.51; 95% confidence interval, 1.14-2.04; P=0.005). Both axillary and subclavian access increased the hazard of failure (P=0.007); hazard ratio for axillary was 1.94, (95% confidence interval, 1.23-3.04) and for subclavian 1.63 (95% confidence interval, 1.08-2.46). Previous lead failure increased the hazard of a subsequent Fidelis failure with a hazard ratio of 3.12 (95% confidence interval, 1.80-5.41; P<0.001).
CONCLUSIONS: The rate of Fidelis failure continues to increase over time, with failures approaching 17% at 5 years. Women, patients with leads inserted via the subclavian or axillary vein, and those with a previous lead fracture were at greatest risk of Fidelis failure. Our data suggest that Fidelis replacement should be strongly considered at the time of generator replacement.

Entities:  

Mesh:

Year:  2012        PMID: 22311781     DOI: 10.1161/CIRCULATIONAHA.111.053744

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  22 in total

1.  [Annual report 2013 of the German Cardiac Pacemaker And Defibrillator Register--Part 2: implantable cardioverter-defibrillators. Pacemaker and AQUA Institute for Applied Quality Improvement and Research in Health Care GmbH workgroup].

Authors:  A Markewitz
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-12

2.  [Annual Report 2011 of the German pacemaker and defibrillator register: Section pacemakers and AQUA-Institute for Applied Quality Improvement and Research in Health Care].

Authors:  A Markewitz
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-12

Review 3.  [What must intensive care physicians know about implantable cardioverter defibrillator?].

Authors:  E Gatterer
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-09-20       Impact factor: 0.840

Review 4.  [ICD leads].

Authors:  Carsten W Israel; Mohamed Karim Sheta
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06

5.  [2010 annual report of the German pacemaker and defibrillator register: section pacemakers and AQUA-Institute for Applied Quality Improvement and Research in Health Care Ltd].

Authors:  A Markewitz
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-12

6.  [Annual Report 2012 of the German Heart pacemaker defibrillator round register: Section on pacemaker and AQUA Institute for Applied Quality Improvement and Research in Health Care GmbH].

Authors:  A Markewitz
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2014-12

7.  Prophylactic implantable defibrillators in dilated cardiomyopathy.

Authors:  W Grimm
Journal:  Herz       Date:  2012-12       Impact factor: 1.443

8.  Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads.

Authors:  Liza P Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; John D Ferguson; J Michael Mangrum; Sandeep Kamath; Pamela K Mason
Journal:  J Interv Card Electrophysiol       Date:  2012-12-20       Impact factor: 1.900

9.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

Review 10.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.