Literature DB >> 27624809

Sensing and detection in Medtronic implantable cardioverter defibrillators.

Mark L Brown1, Charles D Swerdlow2.   

Abstract

Ensuring sensing and detection of ventricular tachycardia (VT) and ventricular fibrillation (VF) was a prerequisite for the clinical trials that established the survival benefit of implantable cardioverter defibrillators (ICDs). However, for decades, a high incidence of unnecessary shocks limited patients' and physicians' acceptance of ICD therapy. Oversensing, misclassification of supraventricular tachycardia (SVT) as VT, and self-terminating VT accounted for the vast majority of unnecessary shocks. Medtronic ICDs utilize sensitive baseline settings with minimal blanking periods to ensure accurate sensing of VF, VT, and SVT electrograms. Programmable algorithms reject oversensing caused by far-field R waves, T waves, and non-physiologic signals caused by lead failure. A robust hierarchy of SVT-VT discriminators minimize misclassification of SVT as VT. These features, combined with evidence-based programming, have reduced the 1‑year inappropriate shock rate to 1.5 % for dual-/triple-chamber ICDs and to 2.5 % for single-chamber ICDs.

Entities:  

Keywords:  Electrogram; Pacemaker; Tachycardia, supraventricular; Tachycardia, ventricular; Ventricular fibrillation

Mesh:

Year:  2016        PMID: 27624809     DOI: 10.1007/s00399-016-0450-6

Source DB:  PubMed          Journal:  Herzschrittmacherther Elektrophysiol        ISSN: 0938-7412


  32 in total

1.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesús Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  Heart Rhythm       Date:  2015-12-01       Impact factor: 6.343

2.  Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee.

Authors:  David H Birnie; Ratika Parkash; Derek V Exner; Vidal Essebag; Jeffrey S Healey; Atul Verma; Benoit Coutu; Teresa Kus; Iqwal Mangat; Felix Ayala-Paredes; Pablo Nery; George Wells; Andrew D Krahn
Journal:  Circulation       Date:  2012-02-06       Impact factor: 29.690

3.  ICD-unresponsive ventricular arrhythmias.

Authors:  Josée Michaud; Irina Horduna; Marc Dubuc; Paul Khairy
Journal:  Heart Rhythm       Date:  2009-06-06       Impact factor: 6.343

4.  Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results.

Authors:  Angelo Auricchio; Edward J Schloss; Takashi Kurita; Albert Meijer; Bart Gerritse; Steven Zweibel; Faisal M AlSmadi; Charles T Leng; Laurence D Sterns
Journal:  Heart Rhythm       Date:  2015-01-28       Impact factor: 6.343

5.  Inappropriate shocks in patients with Fidelis® lead fractures: impact of remote monitoring and the lead integrity algorithm.

Authors:  Zalmen Blanck; Kathi Axtell; Kathy Brodhagen; Laura O'Hearn; Tammy Albelo; Cheryl Ceretto; Anwer Dhala; Jasbir Sra; Masood Akhtar
Journal:  J Cardiovasc Electrophysiol       Date:  2011-05-03

6.  Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implications for device programming.

Authors:  M H Anderson; F D Murgatroyd; K Hnatkova; B Xie; S Jones; E Rowland; D E Ward; A J Camm; M Malik
Journal:  Pacing Clin Electrophysiol       Date:  1997-12       Impact factor: 1.976

7.  Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure.

Authors:  Bruce D Gunderson; Jeffrey M Gillberg; Mark A Wood; Pugazhendhi Vijayaraman; Richard K Shepard; Kenneth A Ellenbogen
Journal:  Heart Rhythm       Date:  2006-02       Impact factor: 6.343

8.  Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads.

Authors:  Charles D Swerdlow; Bruce D Gunderson; Kevin T Ousdigian; Athula Abeyratne; Robert W Stadler; Jeffrey M Gillberg; Amisha S Patel; Kenneth A Ellenbogen
Journal:  Circulation       Date:  2008-11-03       Impact factor: 29.690

9.  Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator.

Authors:  C D Swerdlow; P S Chen; R M Kass; J R Allard; C T Peter
Journal:  J Am Coll Cardiol       Date:  1994-05       Impact factor: 24.094

10.  Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study.

Authors:  Bruce L Wilkoff; Brian D Williamson; Richard S Stern; Stephen L Moore; Fei Lu; Sung W Lee; Ulrika M Birgersdotter-Green; Mark S Wathen; Isabelle C Van Gelder; Brooke M Heubner; Mark L Brown; Keith K Holloman
Journal:  J Am Coll Cardiol       Date:  2008-08-12       Impact factor: 24.094

View more
  2 in total

1.  Failure to Treat Life-Threatening Ventricular Tachyarrhythmias in Contemporary Implantable Cardioverter-Defibrillators: Implications for Strategic Programming.

Authors:  Anna Margrethe Thøgersen; Jacob Moesgaard Larsen; Jens Brock Johansen; Moeen Abedin; Charles D Swerdlow
Journal:  Circ Arrhythm Electrophysiol       Date:  2017-09

2.  A case of withheld defibrillator treatment for 9 minutes of ventricular tachycardia.

Authors:  Evan Grove; Vincent Y See; Timm M Dickfeld
Journal:  HeartRhythm Case Rep       Date:  2020-11-20
  2 in total